HeartFlow Analysis FFRct version 3 (UDI 00853341006060)Recalls & reported problems

Heartflow, Inc. · The HeartFlow Analysis is AI-based medical device software developed for the clinical quantitative and qualitative analysis of CT DICOM data. It is a tool for the analysis of CT DICOM-compliant cardiac images and data, to assess the anatomy and function of the coronary arteries in the risk stratification and evaluation of coronary artery disease. The software displays coronary anatomy and functional information using graphics and text, including computed and derived quantities of percent stenosis, plaque volumes, blood flow, pressure and velocity, to aid the clinician in the assessment and treatment planning of coronary artery disease. The HeartFlow Analysis is performed on previously physician-acquired image data and are unrelated to acquisition equipment and clinical workstations. · Class 2

Quick summary

This HeartFlow Analysis software reads CT scans of the heart to help doctors assess coronary artery disease. Most of the 35 reports describe analysis results that were wrong or incomplete, such as blocked areas being missed or readings released for the wrong artery. Three reports involve injuries and there are no deaths. There are no recalls name this device. Reports have dropped sharply in the past 12 months, from 16 to none.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
35
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 00853341006060), received between Mar 24, 2023 and Sep 11, 2025 — classified as 3 injuries, 32 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 4 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 3 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 4 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings19
  • False Negative Result15
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions19
  • Insufficient Information16
  • Chest Pain2
  • Angina1

Recent adverse event reports

The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A HEARTFLOW ANALYSIS WAS INCORRECTLY RELEASED WITH FFRCT VALUES PROVIDED FOR THE LEFT ANTERIOR DESCENDING (LAD) SYSTEM WITH A STENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23031417 ↗
  • MalfunctionSep 2, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A COMPUTED TOMOGRAPHY (CT) DATASET WAS POTENTIALLY INCORRECTLY ACCEPTED FOR PROCESSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22938920 ↗
  • MalfunctionAug 29, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A COMPUTED TOMOGRAPHY (CT) DATASET WAS POTENTIALLY INCORRECTLY ACCEPTED FOR PROCESSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22928373 ↗
  • MalfunctionAug 27, 2025False Negative Result; False Negative Result

    HEARTFLOW IDENTIFIED A POTENTIAL FALSE NEGATIVE RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22909424 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A POTENTIAL OCCLUSION WAS MISSED IN THE LCX ON THE HEARTFLOW ANALYSIS THAT WAS RELEASED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22743246 ↗
  • MalfunctionAug 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A POTENTIAL COMPUTED TOMOGRAPHY (CT) DATASET WAS INCORRECTLY ACCEPTED FOR PROCESSING AN ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22743239 ↗
  • MalfunctionJun 13, 2025False Negative Result; False Negative Result

    HEARTFLOW IDENTIFIED A POTENTIAL FALSE NEGATIVE RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22206338 ↗
  • MalfunctionMay 20, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A HEARTFLOW ANALYSIS WAS INCORRECTLY RELEASED WITH FFRCT VALUES PROVIDED FOR THE RIGHT CORONARY ARTERY (RCA) SYSTEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22056775 ↗
  • MalfunctionMay 20, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A HEARTFLOW ANALYSIS WAS INCORRECTLY RELEASED WITH FFRCT VALUES PROVIDED FOR THE RIGHT CORONARY ARTERY (RCA) SYSTEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22055964 ↗
  • MalfunctionMay 6, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED THAT A COMPUTED TOMOGRAPHY (CT) DATASET WAS POTENTIALLY INCORRECTLY ACCEPTED FOR PROCESSING AN ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21973201 ↗
  • MalfunctionApr 24, 2025False Negative Result; False Negative Result

    HEARTFLOW IDENTIFIED A POTENTIAL FALSE NEGATIVE RESULT.

    Patient problems: Insufficient Information

    MAUDE report 21903027 ↗
  • MalfunctionApr 24, 2025False Negative Result; False Negative Result

    HEARTFLOW IDENTIFIED A POTENTIAL FALSE NEGATIVE RESULT.

    Patient problems: Insufficient Information

    MAUDE report 21902732 ↗
  • MalfunctionApr 17, 2025False Negative Result; False Negative Result

    HEARTFLOW IDENTIFIED A POTENTIAL FALSE NEGATIVE RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21859483 ↗
  • MalfunctionJan 10, 2025False Negative Result; False Negative Result

    HEARTFLOW IDENTIFIED A POTENTIAL FALSE NEGATIVE RESULT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21130101 ↗
  • MalfunctionDec 10, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A POTENTIAL COMPUTED TOMOGRAPHY (CT) DATASET WAS INCORRECTLY ACCEPTED FOR PROCESSING AN ANALYSIS. A HEALTHCARE PROFESSIONAL (HCP) ADDITIONALLY REPORTED "THIS IS A CASE OF CHRONIC CORONARY SYNDROME WITHOUT SYMPTOMS. IN (B)(6) 2023 THE PATIENT ALSO UNDERWENT A STRESS ECG TEST WITH NEGATIVE RESULTS. THEREFORE, NO FURTHER SCANS ARE BEING CONSIDERED." THE HCP NOTED THAT THE PATIENT SEEMS TO PROPERLY FOLLOW THE INDICATIONS FOR WHEN TO TAKE THEIR MEDICATIONS. THE PATIENT IS SCHEDULED FOR CONSULTATION EVERY 6 MONTHS AND HAD A RECENT VISIT ON (B)(6) 2024. THEIR NEXT VISIT WILL BE IN (B)(6) OF 2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20898383 ↗
  • MalfunctionDec 5, 2024False Negative Result; False Negative Result

    HEARTFLOW IDENTIFIED A POTENTIAL FALSE NEGATIVE RESULT.

    Patient problems: Insufficient Information

    MAUDE report 20859559 ↗
  • MalfunctionOct 18, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A POTENTIAL COMPUTED TOMOGRAPHY (CT) DATASET WAS INCORRECTLY ACCEPTED FOR PROCESSING AN ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 20487141 ↗
  • MalfunctionSep 20, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A POTENTIAL COMPUTED TOMOGRAPHY (CT) DATASET WAS INCORRECTLY ACCEPTED FOR PROCESSING AN ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 20273801 ↗
  • MalfunctionSep 17, 2024False Negative Result; False Negative Result

    HEARTFLOW IDENTIFIED POTENTIAL OVERSIZED MODELING OF THE LEFT MAIN ARTERY IN THE HEARTFLOW ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 20242151 ↗
  • MalfunctionSep 3, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A POTENTIAL INCORRECTLY PROVIDED ANALYSIS FOR THE LAD VESSEL AND RCA VESSEL MODELED TOO LARGE IN THE HEARTFLOW ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 20131629 ↗
  • MalfunctionSep 3, 2024False Negative Result; False Negative Result

    HEARTFLOW IDENTIFIED POTENTIAL OVERSIZED MODELING OF THE LEFT MAIN ARTERY IN THE HEARTFLOW ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 20131357 ↗
  • MalfunctionAug 22, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A POTENTIAL COMPUTED TOMOGRAPHY (CT) DATASET WAS INCORRECTLY ACCEPTED FOR PROCESSING AN ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20043701 ↗
  • MalfunctionJul 3, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A HEARTFLOW ANALYSIS WAS INCORRECTLY RELEASED WITH FFRCT VALUES PROVIDED FOR THE RIGHT CORONARY ARTERY (RCA) SYSTEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19672341 ↗
  • MalfunctionJun 14, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A POTENTIAL COMPUTED TOMOGRAPHY (CT) DATASET WAS INCORRECTLY ACCEPTED FOR PROCESSING AN ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19541004 ↗
  • MalfunctionMay 16, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    HEARTFLOW IDENTIFIED A POTENTIAL COMPUTED TOMOGRAPHY (CT) DATASET WAS INCORRECTLY ACCEPTED FOR PROCESSING AN ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 19337706 ↗

Other System, Image Processing, Radiological devices