NovoCut CTX-C001Recalls & reported problems

Smith & Nephew, Inc. · NOVOCUT SUTURE MANAGER · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00853541006129), received between Oct 4, 2022 and Apr 13, 2026 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Break1
  • Failure to Cut1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Pain2
  • Laceration(s)1
  • Loss of Range of Motion1
  • Swelling/ Edema1
  • Unspecified Musculoskeletal problem1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 13, 2026Break

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR, AFTER THE IMPLANTATION OF THE FIRST FAST-FIX FLEX, THE SUTURE WAS SNAPPED, WHEN ONE NOVOCUT WAS INSERTED INTO THE JOINT WITH THE SUTURE LOADED OVER THE TIP. THE PROCEDURE WAS RESUMED, AFTER A 5-MINUTE DELAY, USING AN EQUIVALENT S+N BACK-UP. ADDITIONAL ANESTHESIA WAS NOT REQUIRED AS CONSEQUENCE OF THE SURGICAL PROLONGATION. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24859686 ↗
  • MalfunctionMar 18, 2025Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT DURING A MENISCUS REPAIR SURGERY, THE NOVOCUT SUTURE MANAGER WAS UNABLE TO CLEANLY CUT THE SUTURE TAIL OF THE FAST-FIX FLEX, LEADING TO STRAGGLES EDGES AND THE FAST-FIX FLEX IMPLANTS HAD TO BE REMOVED. BOTH IMPLANTS WERE REMOVED FROM INSIDE THE PATIENT USING ARTHROSCOPIC GRASPER TO PULL IT BACK THROUGH THE MENISCUS. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE S+N KNOT PUSHER/SUTURE CUTTER WAS USED FOR FURTHER FAST-FIX ANCHOR PLACEMENT. THERE WAS A SURGICAL DELAY OF LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21631882 ↗
  • InjuryMay 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, APPROXIMATELY SIX (6) MONTHS AFTER AN ACL + MENISCAL ROOT REPAIR PROCEDURE PERFORMED ON (B)(6) 2022, WHERE A NOVOCUT SUTURE MANAGER WAS USED; A PATIENT EXPERIENCED SEVERAL ADVERSE EVENTS ON DIFFERENT DATES: ON (B)(6) 2024 THE PATIENT HAD SIGNIFICANT PAIN, REDUCED RANGE OF MOTION AND INABILITY TO REACH FULL FLEXION DUE TO CYCLOPS LESION, THESE AFFECTIONS WERE TREATED WITH PHYSICAL THERAPY; HOWEVER, THERE WERE NOT RESOLVED. FURTHERMORE, THE PATIENT PRESENTED SIGNIFICANT SWELLING DUE TO CYCLOPS LESION, FOR WHICH AN MRI WAS PERFORMED. ASIDE FROM THE ISSUES MENTIONED, ON (B)(6) 2024, THE PATIENT HAD FIXED FLEXION DEFORMITY WITH INABILITY TO REHAB THE DEFICIT, THEREFORE, AN ARTHROSCOPY AND DEBRIDEMENT OF CYCLOPS LESION AND NOTCHPLASTY WERE PERFORMED ON (B)(6) 2024. THE PATIENT OUTCOME IS RESOLVED/RECOVERED.

    Patient problems: Loss of Range of Motion; Pain; Unspecified Musculoskeletal problem

    MAUDE report 19332608 ↗
  • InjuryApr 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER A KNEE ARTHROSCOPY, USING ONE (1) NOVOCUT SUTURE, THE PATIENT EXPERIENCED PAIN AND SWELLING IN THE RIGHT KNEE, AND THE PATIENT REPORTED FEELING A LUMP INSIDE THE RIGHT KNEE. THE PATIENT WAS TREATED ON AN ADDITIONAL SURGERY AND IS NOT RECOVERED.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 16748741 ↗
  • InjuryOct 4, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER AN ARTHROSCOPY USING A NOVOCUT SUTURE, THE PATIENT EXPERIENCED A TWISTING INJURY TO THE RIGHT KNEE. A LARGE DISPLACED FRAGMENT OF LATERAL MENISCUS TEAR WAS NOTED ON MRI. THE PATIENT WAS TREATED BY CONSERVATIVE MANAGEMENT. THE ISSUE IS RESOLVING.

    Patient problems: Laceration(s)

    MAUDE report 15540073 ↗

Other Accessories, Arthroscopic devices

All Accessories, Arthroscopic devices →