NEUWAVE PR15XTRecalls & reported problems
Neuwave Medical, Inc. · NEUWAVE PR XT Probe 15CM 15ga · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This NeuWave PR XT microwave ablation probe appears in 89 adverse event reports, with no deaths and 63 injuries. Many reports describe probe tips breaking off during procedures, with fragments lodged in patients, and gas leaking from the shaft. Listed patient problems include burns and blood pooling. There were no reports in the past 12 months, down from 7 in the year before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 89
- Deaths reported
- 0
- Injuries reported
- 63
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 89 adverse event reports that list this exact device (DI 00853719006401), received between May 2, 2017 and Aug 20, 2025 — classified as 63 injuries, 26 malfunctions.
The most frequently reported specific device problem is “Break” (35 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break35
- Appropriate Device Problem Term/Code Not Available33
- Adverse Event Without Identified Device or Use Problem19
- Gas/Air Leak8
- Excessive Heating4
- Delivered as Unsterile Product2
- Detachment of Device or Device Component2
- Improper or Incorrect Procedure or Method2
Patient problems
- Foreign Body In Patient28
- No Clinical Signs, Symptoms or Conditions26
- Insufficient Information7
- Burn(s)6
- Obstruction/Occlusion6
- Hematoma5
- Pneumothorax4
- Abdominal Pain3
Recent adverse event reports
The 25 most recent of 89 reports, in the reporter's own words as published by the FDA.
InjuryAug 20, 2025Break; Break
IT WAS REPORTED DURING AN ABLATION PROCEDURE THE TIP OF THE PROBE BROKE OFF. HE WAS NOT OPERATING UNDER A BONE OR ANYTHING SIMILAR. THERE WERE NO PATIENT ADVERSE EVENT.
Patient problems: Foreign Body In Patient
MAUDE report 22852317 ↗InjuryJun 12, 2025Break; Break
IT WAS REPORTED THAT DURING A CT GUIDED LIVER ABLATION THAT THE PROBE WAS BEING GUIDED BETWEEN THE RIBS AND THE PROBE TIP BROKE OFF AND IS LODGED IN THE CARTILAGE BETWEEN THE RIBS. A NEW PROBE WAS USED TO CONTINUE WITH THE PROCEDURE.
Patient problems: Foreign Body In Patient
MAUDE report 22197434 ↗InjuryMar 21, 2025Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, WHEN USING AN ABLATION PROBE, IT DOES NOT STORE THE "BUBBLES" TEST IN ITS MEMORY AND REQUESTED IT WHILE THE PROCEDURE WAS IN PROGRESS ON THE PATIENT, FORCING ITS REMOVAL AND RE-TESTING. THIS NECESSITATED THREE PASSAGES INSTEAD OF THE USUAL ONE. THE BUBBLE TEST WAS PERFORMED TWICE MORE, THEN THE PROBE WAS CHANGED FOR ANOTHER MODEL OF THE SAME REFERENCE AND LOT, THIS TIME SUCCESSFULLY. A WEEK LATER, THE PATIENT COMPLAINED OF ABDOMINAL PAIN AND CONSULTED US. A CT SCAN WAS PERFORMED AND AIR WAS VISUALIZED IN THE LIVER LESIONS. THE CT SCAN IS USUALLY DONE 1 MONTH AFTER THE OPERATION, SO NO COMPARISON IS POSSIBLE AT THIS STAGE, BUT THERE IS A POTENTIAL EFFECT DUE TO THE PROBE PROBLEM. AT PRESENT, THE PATIENT IS DOING WELL, BUT IS EXPERIENCING ABDOMINAL PAIN, WHICH IS NOT FREQUENT FOLLOWING THIS TYPE OF OPERATION.
Patient problems: Abdominal Pain; Abscess; Unspecified Infection
MAUDE report 21663447 ↗InjuryFeb 21, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT DURING AN ABLATION, THEY HAD A THREE PROBE ABLATION GOING. CHANNEL THREE PROBE, UNDER ULTRASOUND THE GAS WAS EMITTING BACK WAY UP THE SHAFT NOT JUST ON THE EMITTING ZONE. THE GAS WAS SEEN OUTSIDE THE EMITTING ZONE, WITHIN THE FIRST THIRTY SECONDS AND THEN CONTINUOUS. THEY STOPPED ABLATION WITH CHANNEL THREE. THEY CONTINUED ABLATION WITH CHANNEL ONE AND TWO. THEY SHUT OFF CHANNEL THREE. AND PRECEDED WITH CHANNEL ONE AND TWO FOR AN ADDITIONAL THIRTY SECONDS. AND THEN PAUSED ALL CHANNELS AND DID A CT SCAN. CASE WAS COMPLETED WITH TANK A. WHILE RUNNING ON TANK A THERE WAS AN ABNORMAL CLICKING SOUND. NO PATIENT HARM WAS REPORTED.
Patient problems: Abscess; Hematoma; No Clinical Signs, Symptoms or Conditions; Unspecified Infection
MAUDE report 21433356 ↗InjuryDec 20, 2024Break; Break
IT WAS REPORTED THAT WHEN THEY STARTED ABLATING, ONE OF THE TWO PROBES BEGAN TO FAIL AFTER THE FIRST MINUTE. SURGEON CONTINUED THE CASE WITH JUST ONE PROBE. AFTER COMPLETING THE CASE, THEY WENT TO REMOVE THE FAILED PROBE BUT THE TIP OF IT STARTED TO DANGLE. A SMALL FRAGMENT WAS SEEN IN THE KIDNEY ON X-RAY. NO PLANS MADE TO REMOVE IT DUE TO IT NOT BEING SET TO CAUSE ANY FURTHER HARM TO THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 20988155 ↗InjuryDec 11, 2024Break; Break
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE THAT THE PROBE BROKE IN THE PATIENT DUE TO THE CORD BEING PULLED OUT WHILE THE PATIENT WAS BEING SCANNED. 1 CM OF THE PROBE WAS LEFT IN THE PROBE. IT¿S UNKNOWN IF A FUTURE PROCEDURE WILL BE MADE TO REMOVE THE REMAINDER.
Patient problems: Foreign Body In Patient
MAUDE report 20907071 ↗MalfunctionDec 11, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING A LIVER MICROWAVE ABLATION THAT THE ABLATION TEMPERATURE IS CONSTANTLY AROUND 92 DEGREE C (S7, 2.5 CM LESION) FOR THE FIRST CYCLE 7 MINUTES. THE ABLATION ZONE IS NOT LARGE ENOUGH AS EXPECTED WHICH BELOW 100 DEGREE C, THUS HE HAS TO ADD ANOTHER CYCLE 2:50MINS. NO PARTS OF THE INSTRUMENT LEFT INSIDE PATIENT'S CAVITY. THE PROCEDURE WAS DELAYED 2:50MINS BUT COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20905892 ↗InjuryOct 11, 2024Break; Excessive Heating; Break; Excessive Heating
IT WAS REPORTED THAT DURING A LIVER ABLATION PROCEDURE THAT WHILE USING TWO PROBES THE TEMPERATURE WAS GETTING PRETTY HIGH AND WOULD SHUT OFF AND ON. WHEN PROBES WERE REMOVED IT WAS FOUND THAT THE TIPS HAD "BURNT" OFF AND WERE LEFT IN THE PATIENT. A THIRD PROBE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES OTHER THAN THE TIPS OF THE FIRST TWO PROBES BEING LEFT IN THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 20426467 ↗InjuryOct 11, 2024Break; Excessive Heating; Break; Excessive Heating
IT WAS REPORTED THAT DURING A LIVER ABLATION PROCEDURE THAT WHILE USING TWO PROBES THE TEMPERATURE WAS GETTING PRETTY HIGH AND WOULD SHUT OFF AND ON. WHEN PROBES WERE REMOVED IT WAS FOUND THAT THE TIPS HAD "BURNT" OFF AND WERE LEFT IN THE PATIENT. A THIRD PROBE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES OTHER THAN THE TIPS OF THE FIRST TWO PROBES BEING LEFT IN THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 20426455 ↗MalfunctionAug 22, 2024Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT DURING A LIVER ABLATION PROCEDURE THAT PRIOR TO INTRODUCTION INTO THE PATIENT THE TIP OF THE PROBE BROKE BUT DID NOT DID NOT DETACH. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20029563 ↗MalfunctionAug 5, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING AN ABLATION THE DEVICE'S ABLATION ZONES ARE A LOT SMALLER THAN WHAT THEY ARE SUPPOSED TO BE ACCORDING TO THE PHYSICIAN THAT HAS HAD AN ISSUE WITH IT FOR A GOOD BIT. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19914376 ↗MalfunctionJun 11, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT THE CUSTOMER RECEIVED ONE MANUFACTURERS BOX THAT HAD NO PRODUCT CODE, LOT NUMBER, OR EXPIRATION DATE REFERENCED AND THERE WAS A LABEL THAT SAYS, ¿NOT FOR HUMAN USE¿. THERE WAS NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19509871 ↗MalfunctionApr 16, 2024Break; Break
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE THE PROBE TIP WAS BROKEN WHILE TRYING TO REACH DEEP LESION. NO PROBE COMPONENTS WERE LEFT INSIDE PATIENT. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19115449 ↗InjuryApr 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED VIA AN ABLATION CLINICAL TRIAL (B)(6) THE PATIENT EXPERIENCES SMALL RIGHT PNEUMOTHORAX AND SUBCAPSULAR HEMATOMA. THE RELATIONSHIP TO THE STUDY DEVICE FOR BOTH EVENTS IS POSSIBLE AND THE RELATIONSHIP TO THE STUDY PROCEDURE FOR BOTH EVENTS IS CAUSAL. BOTH EVENTS REQUIRED IN-PATIENT OR PROLONGATION OF EXISTING HOSPITALIZATION. THE SMALL RIGHT PNEUMOTHORAX WAS TREATED WITH NEEDLE EXUFFLATION. THERE WAS NO TREATMENT FOR THE SUBCAPSULAR HEMATOMA.
Patient problems: Hematoma; Pneumothorax
MAUDE report 19070079 ↗MalfunctionMar 6, 2024Break; Break
IT WAS REPORTED THAT DURING A LIVER ABLATION PROCEDURE THAT THE TIP OF THE PROBE BROKE OFF IN THE LIVER. FRAGMENT WAS FOUND AND REMOVED. UNKNOWN AS TO HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18847027 ↗MalfunctionFeb 23, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE CUSTOMER RECEIVED DAMAGED PRODUCT. THE PACKAGING HAD HOLES AND THE STERILE PACKAGING HAS A HOLE IN IT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18769304 ↗MalfunctionDec 15, 2023Adverse Event Without Identified Device or Use Problem; Excessive Heating; Adverse Event Without Identified Device or Use Problem; Excessive Heating
IT WAS REPORTED THAT DURING THE ABLATION, 2 PROBES WERE 1 CM APART, BUT AFTER PROCEDURE THEY NOTICED HIGHER TEMPERATURE ON ONE OF THEM AROUND 165 DEGREES CELSIUS. WHEN THEY PULLED OUT PROBE, THEY NOTICED SKIN BURN/FROST BITE ON THE PATIENT. NO ERROR CODES FOR THE DEVICES WERE SHOWN. ONCE PULLED OUT, AN ENGINEER TESTED THE PROBE WITH HIGH TEMPERATURE IN WATER AND IT SEEMED TO WORK FINE.
Patient problems: Burn(s)
MAUDE report 18340166 ↗InjuryNov 20, 2023Break; Break
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, PR15XT PROBE TIP BROKE OFF IN PATIENT AND HAD TO BE RETRIEVED VIA INCISION OF SKIN AND STITCHED BACK UP. THE PHYSICIAN BARELY ATTEMPTED TO INSERT THE PROBE AND IT BROKE OFF. TIP PARTS HAD TO BE EXTRACTED VIA EXTRA SCANS AND CUTTING INTO SKIN SITE.
Patient problems: Insufficient Information; Pain
MAUDE report 18169385 ↗InjuryNov 20, 2023Break; Break
IT WAS REPORTED THAT DURING AN INTERVENTIONAL IMAGING, IT WAS OBSERVED THAT AFTER 7 MINUTES OF TREATMENT TO ABLATE A LIVER TUMOR, THE DEVICE WAS FOUND TO BE IN ERROR ON ONE OF THE TWO NEEDLES, WITH AN ERROR MESSAGE BEING DISPLAYED. AFTER SEVERAL ATTEMPTS TO RESTART THE ABLATION, AND AN ATTEMPT TO CHANGE THE CHANNEL, THE NEEDLE WAS WITHDRAWN. IT WAS FOUND THAT THE TIP OF THE NEEDLE HAD BEEN DAMAGED AND THAT A PIECE HAD REMAINED IN THE PATIENT'S LIVER. IMAGING WAS PERFORMED AND CONFIRMED THE PRESENCE OF THE FRAGMENT. NO REMOVAL OF THE FRAGMENT IS PLANNED. THIS INCIDENT ALSO EXPOSED THE PATIENT TO THE RISK OF PARTIAL REMOVAL OF THE LIVER TUMOUR.
Patient problems: Foreign Body In Patient
MAUDE report 18169311 ↗MalfunctionNov 3, 2023Failure to Read Input Signal; Failure to Read Input Signal
IT WAS REPORTED THAT DURING THE LEFT ADRENAL ABLATIONS PROCEDURE THAT WHILE USING UP TO THREE PROBES. ONE PROBE STOPPED WORKING AND WOULD NOT ALLOW THEM TO RETEST. THE PROBLEM WITH THE PROBE DID NOT CAUSE ANY DAMAGE TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18064420 ↗InjuryOct 5, 2023Break; Break
IT WAS REPORTED THAT DURING AN ABLATION THE PROBE STOPPED WORKING. IT STOPPED AFTER 7 MINUTES OF TREATMENT. THE RADIOLOGIST RESTARTED THE ABLATION MODE BUT THE PROBE WAS NOT WORKING. HE PLUGGED THE PROBE BACK INTO ANOTHER CHANNEL BUT IT STILL DIDN'T WORK. THEY USED A SECOND PROBE AND THE TREATMENT WENT WELL AFTERWARDS. THEY WERE ABLE TO COMPLETE THE PROCEDURE WITH NO CONSEQUENCES FOR THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 17878358 ↗MalfunctionSep 25, 2023Break; Break
IT WAS REPORTED THAT DURING ABLATION THE PROBE TIP MELTED OFF IN THE PATIENT WHILE ON THE TABLE. IT WAS RETRIEVED. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17807028 ↗MalfunctionSep 22, 2023Break; Material Integrity Problem; Excessive Heating; Break; Material Integrity Problem; Excessive Heating
IT WAS REPORTED THAT DURING AN ABLATION THEY HAD A PATIENT INCIDENT WITH A PROBE TIP BREAKING OFF INSIDE THE PATIENT. BURNED FOR 4 TO 4.5 MINUTES. TEMPERATURE JUMPED UP SIGNIFICANTLY AND THEN STOPPED, REFLECTED POWER ERROR. THEY TRIED TO TURN IT ON AGAIN AND IT BURNED FOR A SECOND BEFORE THEY REALIZED THE TIP BURNED OFF. NO PATIENT CONSEQUENCE SINCE THE PROBE TIP WAS RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17803128 ↗InjurySep 22, 2023Break; Appropriate Device Problem Term/Code Not Available; Break; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT TWO MINUTES IN THE SURGERY THEY RECEIVED AN ERROR "REFLECTIVE POWER." THEY ATTEMPTED TO TROUBLE SHOOT THEY DECIDED THEY NEEDED TO CHANGE PROBE AND UPON REMOVING PROBE, THEY NOTICED THE PROBE WAS NOT THERE AND THAT IT HAD BROKEN OFF INSIDE OF PATIENT. THEY FINISHED THE PROCEDURE AND CALLED REP. THE TIP IS CURRENTLY IN THE BACK MUSCLE OF THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 17801043 ↗InjuryJul 27, 2023Break; Appropriate Device Problem Term/Code Not Available; Break; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED WHO WAS IN THE CASE THAT DURING A THREE PROBE LIVER ABLATION PROCEDURE THAT TWO OF THE TIPS BROKE OFF IN THE PATIENT AND WERE NOT RECOVERED. NEW PROBES WERE USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 17406994 ↗
FDA clearance and classification
The FDA regulates product code NEY (system, ablation, microwave and accessories) under 21 CFR 878.4400 as a Class II device. 21 CFR 878.4400 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NEUWAVE PR15XT
Has the NEUWAVE PR15XT been recalled?
No FDA recall record in our data names this exact device identifier (DI 00853719006401), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NEUWAVE PR15XT?
The FDA's MAUDE database lists 89 adverse event reports for this device, received between May 2, 2017 and Aug 20, 2025. In those reports, the most common specific device problems are “Break” (35), “Gas/Air Leak” (8) and “Excessive Heating” (4); the most common patient problems are “Foreign Body In Patient” (28), “Burn(s)” (6) and “Obstruction/Occlusion” (6). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NEUWAVE PR15XT?
None of the reports are classified as deaths, 63 as injuries and 26 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NEUWAVE PR15XT MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the NEUWAVE PR15XT, and is it still sold?
The FDA's device database lists Neuwave Medical, Inc. as the company responsible for it (brand NEUWAVE, model PR15XT). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES devices
- Emprint CA20L2Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 recalls1 death43 reports
- Emprint CA15L2Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 recalls1 death41 reports
- Emprint CA30L2Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 recalls2 reports
- Emprint CA15L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall49 reports
- Solero 700106001-USAngiodynamics, Inc. · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES92 reports
- Emprint CA20L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall18 reports
- Emprint CA30L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall1 death9 reports
- Emprint CAGEN1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 deaths53 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page