Emprint CA30L2 (UDI 10884521706606)Recalls & reported problems
Covidien LP · Long Percutaneous Antenna with Thermosphere (TM) Technology Large Spherical Ablation Zone - Reinforced Antenna · Class 2
All models of Emprint (19)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521706606), received between Mar 15, 2021 and Nov 1, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
2 other recalls cover devices cleared under the same FDA submission (K163105); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Nov 15, 2018Class IIOpen, Classified
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
Z-0625-2019 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Aug 7, 2017Class IIOpen, Classified
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
Z-0561-2018 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Detachment of Device or Device Component1
- Device Alarm System1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 1, 2021Device Alarm System; Adverse Event Without Identified Device or Use Problem; Device Alarm System; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE DEVICE HAD AN ALARM ISSUE. THE MACHINE SHOWED "TEMPERATURE ERROR" WITH COOLING SYSTEM TURNED ON AND SALINE FLOW WAS NOT OBSTRUCTED BUT CANNOT FIND THE REASON. THEY CLEARED THE ALARM BY PRESSING RESET, BUT THE ALERT MARK SHOWED AGAIN. THEY TRIED A DEMO DEVICE AND IT WORKED FINE. PROCEDURE WAS CANCELLED AND RESCHEDULED WHILE THE PATIENT WAS UNDER ANESTHESIA. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12726520 ↗MalfunctionMar 15, 2021Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING AN ABLATION PROCEDURE, THE TOP OF THE ANTENNA BROKE. THERE WAS NO ARTIFACT LEFT IN PATIENT. THERE WAS NO PATIENT INJURY. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND THAT THE UNIT'S TIP WAS DETACHED, MISSING AND WAS NOT RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11488648 ↗
FDA clearance and classification
Cleared through Special 510(k) K163105 on Dec 6, 2016 (decision: Substantially Equivalent). Submitted by Covidien, LLC as “Emprint Ablation System”. The FDA review took 29 days. FDA record ↗
The FDA regulates product code NEY (system, ablation, microwave and accessories) under 21 CFR 878.4400 as a Class II device. 21 CFR 878.4400 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K163105. One submission often covers many models, so check the product and lot details in each record.
- Z-0626-2019 · Nov 15, 2018 · Open, ClassifiedCovidien Emprint Short Percutaneous Antenna with Thermosphere Technology, Material CA15L2 intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation)
- Z-0627-2019 · Nov 15, 2018 · Open, ClassifiedCovidien Emprint Standard Percutaneous Antenna with Thermosphere Technology, CA20L2 intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of sof
Common questions about the Emprint CA30L2
Has the Emprint CA30L2 been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-0625-2019 (Class II), began on Nov 15, 2018: The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Emprint CA30L2?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Mar 15, 2021 and Nov 1, 2021. In those reports, the most common specific device problems are “Break” (1), “Detachment of Device or Device Component” (1) and “Device Alarm System” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Emprint CA30L2?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Emprint CA30L2 FDA approved?
It was cleared through Special 510(k) K163105 on Dec 6, 2016. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Emprint CA30L2 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Emprint CA30L2, and is it still sold?
The FDA's device database lists Covidien LP as the company responsible for it (brand Emprint, model CA30L2). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES devices
- Emprint CA20L2Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 recalls1 death43 reports
- Emprint CA15L2Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 recalls1 death41 reports
- Emprint CA15L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall49 reports
- Solero 700106001-USAngiodynamics, Inc. · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES92 reports
- NEUWAVE PR15XTNeuwave Medical, Inc. · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES89 reports
- Emprint CA20L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall18 reports
- Emprint CA30L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall1 death9 reports
- Emprint CAGEN1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 deaths53 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page