Quick summary

This LifeVest wearable defibrillator electrode belt is a wearable defibrillator component. Reports mostly describe circuit failures, structural problems, and false signal readings, including inappropriate shocks. There are 93 reports naming it, with 1 death and 9 injuries, and no recalls name this device. Reports rose sharply, from 17 in the previous 12 months to 73 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
114
Deaths reported
1
Injuries reported
15
Recalls naming it
0

Summary

The FDA's MAUDE database contains 114 adverse event reports that list this exact device (DI 00855778005388), received between Oct 7, 2024 and Sep 30, 2026 — classified as 1 death, 15 injuries, 98 malfunctions.

The most frequently reported specific device problem is “Circuit Failure” (83 reports).

6 other recalls cover devices cleared under the same FDA submission (P010030); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 4 reportsDec 2024: 0 reportsJan 2025: 7 reportsFeb 2025: 3 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 5 reportsApr 2026: 8 reportsMay 2026: 12 reportsJun 2026: 8 reportsJul 2026: 14 reportsAug 2026: 19 reportsSep 2026: 26 reportsOct 2026: 0 reports20242025202626
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Circuit Failure83
  • Structural Problem14
  • Incorrect Interpretation of Signal12
  • Over-Sensing3
  • Communication or Transmission Problem2
  • Key or Button Unresponsive/not Working2
  • Adverse Event Without Identified Device or Use Problem1
  • Battery Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions98
  • Arrhythmia16

Recent adverse event reports

The 25 most recent of 114 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Circuit Failure

    A US distributor reported that a patient's electrode belt could not run detect and treat testing.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944781 ↗
  • MalfunctionSep 29, 2026Circuit Failure

    A US distributor reported that a patient's electrode belt could not run detect and treat testing.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936698 ↗
  • MalfunctionSep 29, 2026Circuit Failure

    A US distributor reported that a patient's electrode belt could not run detect and treat testing.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936601 ↗
  • InjurySep 29, 2026Incorrect Interpretation of Signal

    Patient problems: Arrhythmia

    MAUDE report 50933249 ↗
  • MalfunctionSep 29, 2026Circuit Failure

    A US distributor reported that an electrode belt was damaged and displaying adjust belt and check pad messages.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50931287 ↗
  • MalfunctionSep 29, 2026Circuit Failure

    A US distributor reported that a patient's electrode belt could not run detect and treat testing.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50931073 ↗
  • InjurySep 29, 2026Incorrect Interpretation of Signal

    A patient received three inappropriate shocks. Double counting contributed to the false detection. The response buttons were pressed after treatment. The patient was at facility at time of treatment. The patient did continue use of lifevest. No injury was reported.

    Patient problems: Arrhythmia

    MAUDE report 50930460 ↗
  • InjurySep 28, 2026Incorrect Interpretation of Signal

    A patient received an inappropriate shock. Double counting contributed to the false detection. The response buttons were pressed during the event.The patient did not seek medical attention and did continue use of lifevest. No injury was reported.

    Patient problems: Arrhythmia

    MAUDE report 50920131 ↗
  • MalfunctionSep 28, 2026Circuit Failure

    A US Distributor returned an electrode belt was unable to pass essential performance testing.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916712 ↗
  • MalfunctionSep 28, 2026Circuit Failure

    A US distributor reported that an electrode belt had tactile / vibration alarms or lack thereof.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916415 ↗
  • InjurySep 28, 2026Incorrect Interpretation of Signal

    A patient received an inappropriate shock. Double counting contributed to the false detection. The response buttons were not pressed during the event. The patient went to the hospital for further evaluation and continues to wear the LifeVest. No injury was reported.

    Patient problems: Arrhythmia

    MAUDE report 50913571 ↗
  • MalfunctionSep 24, 2026Structural Problem

    A US distributor reported that an electrode belt had bent pins or damaged E-Belt connector.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884741 ↗
  • MalfunctionSep 23, 2026Circuit Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872811 ↗
  • MalfunctionSep 23, 2026Circuit Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872733 ↗
  • MalfunctionSep 23, 2026Structural Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872453 ↗
  • MalfunctionSep 15, 2026Circuit Failure

    A US distributor reported that an electrode belt was unable to complete incoming functionality testing.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50771845 ↗
  • MalfunctionSep 15, 2026Structural Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50771398 ↗
  • MalfunctionSep 15, 2026Circuit Failure

    A US distributor reported that an electrode belt was unable to complete incoming functionality testing.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50771169 ↗
  • MalfunctionSep 14, 2026Structural Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50761880 ↗
  • InjurySep 10, 2026Incorrect Interpretation of Signal

    Patient problems: Arrhythmia

    MAUDE report 50722493 ↗
  • MalfunctionSep 8, 2026Circuit Failure

    A US distributor returned a belt and reported being unable to complete essential performance testing.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50665887 ↗
  • MalfunctionSep 5, 2026Circuit Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50630506 ↗
  • MalfunctionSep 4, 2026Circuit Failure

    A US distributor returned a belt and reported being unable to complete essential performance testing.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50610961 ↗
  • InjurySep 3, 2026Over-Sensing

    A patient received an inappropriate shock. Motion artifact contributed to the false detection. The response buttons were pressed after the event. The patient went to the hospital for further evaluation and was discharged home. Patient continues to wear the LifeVest.No injury was reported from the treatment. No deficiencies were alleged against the device.

    Patient problems: Arrhythmia

    MAUDE report 50592010 ↗
  • MalfunctionSep 3, 2026Structural Problem; Key or Button Unresponsive/not Working

    A US distributor returned a monitor and reported that the response buttons were non-functional.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50591216 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P010030. One submission often covers many models, so check the product and lot details in each record.

  • Z-0353-2018 · Sep 12, 2017 · Terminated
    LifeVest Wearable Defibrillator Model 4000, Product Number 10A0988-A01. The LifeVest system is indicated for patients 18 years of age and older who are at risk for sudden cardiac
  • Z-0625-2011 · Sep 24, 2010 · Terminated
    Model 4000 Monitor component of the LifeVest Wearable Defibrillator
  • Z-0883-2009 · Oct 21, 2008 · Terminated
    Monitor portion of the LifeVest device - specifically the Defibrillator printed circuit assembly (PCA) within the monitor. The LifeVest is indicated for adult patients who are at
  • Z-0061-2008 · Sep 14, 2007 · Terminated
    Battery Pack portion of the LifeVest device. The LifeVest is a wearable cardioverter defibrillator worn by a patient at risk for sudden cardiac arrest (SCA).
  • Z-0936-2007 · Apr 27, 2007 · Terminated
    Electrode Belt, component of the LifeVest Wearable Defibrillator, Product number 10A0889-A01. Specifically the 3 Therapy Electrode Pads attached to the Electrode Belt.
  • Z-0578-03 · Jan 10, 2003 · Terminated
    Life Vest (WCD 3000) Electrode Belt. Wearable Cardioverter Defibrillator. Product label 20A0051-FI.