GalaSHAPE 3D 100042-06Recalls & reported problems
Tepha, Inc. · GalaSHAPE 3D scaffold is a bioresorbable surgical mesh manufactured from poly-4-hydroxybutyrate (P4HB). · Class 2 · Implant
All models of GalaSHAPE 3D (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 18
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00855920006195), received between Nov 13, 2017 and Nov 2, 2022 — classified as 18 injuries.
The most frequently reported specific device problem is “Patient-Device Incompatibility” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Patient-Device Incompatibility14
- Device Contamination With Biological Material2
- Microbial Contamination of Device2
- Contamination1
- Device Appears to Trigger Rejection1
Patient problems
- Bacterial Infection5
- Necrosis3
- Seroma3
- Hematoma2
- Wound Dehiscence2
- Capsular Contracture1
- Cellulitis1
- Erythema1
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
InjuryNov 2, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2019, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A VERTICAL INCISION. TWO PIECES OF GALASHAPE (GALAFLEX 3D) WERE USED. ON (B)(6) 2019, RIGHT BREAST SEROMA WAS REPORTED. ON (B)(6) 2019, PATIENT HAD DEBRIDEMENT OF RIGHT BREAST WOUND AS WELL AS REMOVAL OF EXPOSED RIGHT BREAST IMPLANT AND EXPOSED GALASHAPE (GALAFLEX 3D) MESH. REPLACEMENT OF RIGHT BREAST IMPLANT AND PRIMARY CLOSURE OF RIGHT BREAST OPEN WOUND. PATIENT WAS TREATED WITH MEDICATION (KEFLEX 500MG, CIPRO 500MG). ON (B)(6) 2019, RIGHT BREAST INFECTION WAS REPORTED,(SERRATIA MARCESCENS AND PSEUDOMONAS AERUGINOSA.) PATIENT WAS TREATED WITH CONTINUED MEDICATION (KEFLEX 500MG, CIPRO 500MG). RESOLVED WITH SEQUELAE ON (B)(6) 2019. ON (B)(6) 2019, PATIENT HAD DEBRIDEMENT OF SMALL OPENING OF RIGHT BREAST INCISION. RESOLVED WITHOUT SEQUELAE ON (B)(6) 2019. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED SEROMA AND INFECTION ARE LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Bacterial Infection; Seroma
MAUDE report 15713414 ↗InjuryNov 2, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2017, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A VERTICAL INCISION. TWO PIECES OF GALAFLEX 3D (DEVICE #1 AND DEVICE #2) WERE USED. ON (B)(6) 2017, NECROSIS; LEFT AREOLA WAS REPORTED. PATIENT WAS TREATED WITH MEDICATION (NITROBID, BETADINE WIPES). RESOLVED WITH SEQUELAE ON (B)(6) 2018. PATIENT IS REPORTED TO HAVE A LEFT AREOLAR SCAR. AS REPORTED THE PATIENT¿S AE WAS ASSESSED BY THE CLINICIAN AS SEVERE AND A SERIOUS ADVERSE EVENT. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED NECROSIS IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Necrosis; Scar Tissue
MAUDE report 15713375 ↗InjuryJul 28, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2019, PATIENT UNDERWENT BILATERAL MASTOPEXY/REDUCTION; INFERIOR PEDICLE, USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALASHAPE (GALAFLEX 3D) WERE USED. NO CONCOMITANT PROCEDURES WERE DONE. ON (B)(6) 2019, WOUND DEHISCENCE; BILATERAL BREASTS WAS REPORTED. PATIENT TREATED WITH MEDICATION (TOPICAL AQUAPHOR). RESOLVED WITHOUT SEQUELAE ON (B)(6) 2019. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE WOUND DEHISCENCE IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Wound Dehiscence
MAUDE report 15123728 ↗InjuryJul 28, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2019, PATIENT UNDERWENT BILATERAL MASTOPEXY/REDUCTION; INFERIOR PEDICLE, USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALASHAPE (GALAFLEX 3D) WERE USED. NO CONCOMITANT PROCEDURES WERE DONE. ON (B)(6) 2019, EDEMA; BILATERAL BREASTS WAS REPORTED. PATIENT TREATED WITH MEDICATION (TOPICAL AQUAPHOR). RESOLVED WITHOUT SEQUELAE ON (B)(6) 2020. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE EDEMA IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Swelling/ Edema
MAUDE report 15123717 ↗InjuryJul 28, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2019, PATIENT UNDERWENT BILATERAL MASTOPEXY/REDUCTION; INFERIOR PEDICLE, USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALASHAPE (GALAFLEX 3D) WERE USED. NO CONCOMITANT PROCEDURES WERE DONE. ON (B)(6) 2019, ERYTHEMA/REDNESS; BILATERAL BREASTS WAS REPORTED. PATIENT WAS TREATED WITH MEDICATION (TOPICAL AQUAPHOR). RESOLVED WITHOUT SEQUELAE ON (B)(6) 2020. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE ERYTHEMA/REDNESS IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Erythema
MAUDE report 15123711 ↗InjuryJul 26, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2019, PATIENT UNDERWENT BILATERAL MASTOPEXY; INFERIOR PEDICLE, USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALASHAPE (GALAFLEX 3D) WERE USED. CONCOMITANT PROCEDURE INCLUDED VASER ASSISTED LIPOSUCTION TO POSTERIOR HIPS AND ABDOMEN. ON (B)(6) 2019, BILATERAL NIPPLE NECROSIS WAS REPORTED. PATIENT HAD DEBRIDEMENT PROCEDURE ON (B)(6) 2019. RESOLVED WITHOUT SEQUELAE ON (B)(6) 2019. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED NECROSIS IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Necrosis
MAUDE report 15101770 ↗InjuryJul 12, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(6): ON (B)6) 2017, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A VERTICAL INCISION. 2 PIECES OF GALASHAPE (GALAFLEX 3D) WERE USED. CONCOMITANT PROCEDURE WAS LIPOSUCTION. ON (B)(6) 2017, LEFT BREAST HEMATOMA WAS REPORTED. HEMATOMA WAS MILKED THROUGH SMALL OPENING OF INCISION. RESOLVED WITHOUT SEQUELAE ON (B)(6) 2017. RESOLVED WITH SEQUELAE ON AN UNKNOWN DATE. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED HEMATOMA IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Hematoma
MAUDE report 14988488 ↗InjuryJul 12, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2017, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A VERTICAL INCISION. 2 PIECES OF GALASHAPE (GALAFLEX 3D) WERE USED. CONCOMITANT PROCEDURE WAS LIPOSUCTION. ON (B)(6) 2017, LEFT BREAST SEROMA WAS REPORTED. PATIENT TREATED WITH PLACEMENT OF PENROSE DRAIN, WET TO DRY DRESSING, WOUND VAC. AND WITH MEDICATION (CIPRO). RESOLVED WITH SEQUELAE ON (B)(6) 2017. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED SEROMA IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Seroma
MAUDE report 14988342 ↗InjuryJul 12, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(6): ON (B)(6) 2017, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A VERTICAL INCISION. 2 PIECES OF GALASHAPE (GALAFLEX 3D) WERE USED. CONCOMITANT PROCEDURE WAS LIPOSUCTION. ON (B)(6) 2017, BILATERAL WOUND DEHISCENCE WAS REPORTED. PATIENT UNDERWENT DELAYED PRIMARY CLOSURE OF BREAST WOUNDS ON (B)(6) 2017 AND WAS TREATED WITH MEDICATION (KEFLEX, CIPRO). RESOLVED WITH SEQUELAE ON AN UNKNOWN DATE. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED WOUND DEHISCENCE IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Wound Dehiscence
MAUDE report 14988298 ↗InjuryJun 21, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2017, PATIENT UNDERWENT BILATERAL REVISION AUGMENTATION MASTOPEXY USING A VERTICAL INCISION. 2 PIECES OF GALASHAPE (GALAFLEX 3D) WERE USED. CONCOMITANT PROCEDURES INCLUDED CAPSULORRAPHIES AND PIXIE EAR CORRECTION. ON (B)(6) 2018, BILATERAL CAPSULAR CONTRACTURE WAS REPORTED. EVIDENCE OF GRADE II/III CAP CON BILATERALLY. PATIENT WAS TREATED WITH SINGULAIR. PATIENT WAS TO FOLLOW UP IN 5 MONTHS HOWEVER, LAST DOCUMENTED FOLLOW UP ON CASE REPORT FORMS IS ON (B)(6) 2018. UNKNOWN END DATE. REVIEW BY MEDICAL AFFAIRS IDENTIFIES CAPSULAR CONTRACTURE AS LIKELY RELATED TO THE DEVICE AND NOT SERIOUS.
Patient problems: Capsular Contracture
MAUDE report 14756148 ↗InjuryJun 15, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2017, PATIENT UNDERWENT BILATERAL REVISION AUGMENTATION; PTOSIS, USING A VERTICAL INCISION. 2 PIECES OF GALASHAPE WERE USED. CONCOMITANT PROCEDURE WAS MASTOPEXY. ON (B)(6) 2017, DELAYED WOUND HEALING; LEFT MID VERTICAL BREAST WOUND, WAS REPORTED. PATIENT WAS TREATED WITH MEDICATION (SILVADENE, SILVER NITRATE). RESOLVED WITHOUT SEQUELAE ON (B)(6) 2017. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED DELAYED WOUND HEALING IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Impaired Healing
MAUDE report 14693606 ↗InjuryJun 15, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2017, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A VERTICAL INCISION. 2 PIECES OF GALASHAPE (GALAFLEX 3D) WERE USED. THERE WERE NO CONCOMITANT PROCEDURES DONE. ON (B)(6) 2017, NECROSIS; BILATERAL AREOLAR COMPLEXES, WAS REPORTED. REPORTED AS SAE. PATIENT HAD SKIN GRAFTING PROCEDURE ON (B)(6) 2017 AND WAS ALSO TREATED WITH MEDICATION (SILVADENE, KEFLEX 500MG, NITRO-BID). RESOLVED WITHOUT SEQUELAE ON (B)(6) 2017. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED NECROSIS IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Necrosis
MAUDE report 14693571 ↗InjuryJun 7, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2017, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A VERTICAL INCISION. 2 PIECES OF GALASHAPE (GALAFLEX 3D) WERE USED. CONCOMITANT PROCEDURES WERE BRACHIOPLASTY AND LIPOSUCTION TO ARMS, AXILLA AND CHEST. ON 5/15/17, CELLULITIS OF LEFT BREAST WAS REPORTED. SAE. WOUND CULTURE REVEALED ACINETOBACTER. PATIENT HAD INCISION AND DRAINAGE IN THE OPERATING ROOM ON (B)(6) 2017. AREAS OF UNINCORPORATED SCAFFOLD WERE REMOVED. THERE WERE AREAS OF WELL INCORPORATED MESH THAT WERE LEFT IN PLACE. PATIENT WAS ALSO TREATED WITH MEDICATION (SILVADENE, IV VANCOMYCIN, IV CEFEPIME, CIPRO). INFECTIOUS DISEASE REFERRAL AND TREATMENT. RESOLVED WITH SEQUELAE ON (B)(6) 2017. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED CELLULITIS IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Bacterial Infection; Cellulitis
MAUDE report 14621861 ↗InjuryApr 1, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2017, PATIENT UNDERWENT BILATERAL REVISION AUGMENTATION; CAPSULAR CONTRACTURE USING A VERTICAL INCISION. 2 PIECES OF GALASHAPE (GALAFLEX 3D) WERE USED. CONCOMITANT PROCEDURES INCLUDED CAPSULORRAPHY, CAPSULECTOMIES AND MASTOPEXY. ON 9/5/17, CULTURES FROM THE VERTICAL INCISION SHOWED STAPHYLOCOCCUS AUREUS PLUS A GRAM NEGATIVE BACTERIA. PATIENT HAD ADDITIONAL SURGERY ON (B)(6) 2017 CONSISTING OF INCISION AND DRAINAGE OF LEFT BREAST. LEFT BREAST IMPLANT AND LEFT BREAST MESH WERE REMOVED. MEDICATION (LOTRISONE CREAM, KEFLEX, DIFLUCAN, AUGMENTIN) WAS ALSO GIVEN AS TREATMENT. ON (B)(6) 2018, PATIENT HAD A SUBSEQUENT NON P4HB SURGERY - LEFT BREAST IMPLANT WAS REPLACED AND RIGHT PERIAREOLAR SCAR WAS REVISED. NO MENTION OF LEFT BREAST MESH BEING REPLACED, JUST LEFT BREAST IMPLANT. COMPLETED WITHOUT COMPLICATIONS. AE RESOLVED WITHOUT SEQUELAE ON (B)(6) 2018. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE AE OF INFECTION IS LIKELY RELATED AND SERIOUS.
Patient problems: Bacterial Infection
MAUDE report 13977880 ↗InjuryApr 1, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(6): ON (B)(6) 2017, PATIENT UNDERWENT BILATERAL REVISION AUGMENTATION; CAPSULAR CONTRACTURE USING A VERTICAL INCISION. 2 PIECES OF GALASHAPE (GALAFLEX 3D) WERE USED. CONCOMITANT PROCEDURES INCLUDED CAPSULORRAPHY, CAPSULECTOMIES AND MASTOPEXY. ON (B)(6) 2017, LEFT BREAST SEROMA/HEMATOMA WAS REPORTED. AE RESOLVED WITHOUT SEQUELAE ON (B)(6) 2018. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE AE OF SEROMA HEMATOMA IS LIKELY RELATED AND SERIOUS.
Patient problems: Hematoma; Seroma
MAUDE report 13977762 ↗InjuryOct 24, 2018Contamination; Contamination
PHYSICIAN REPORTED THAT 9 DAYS AFTER THE SURGERY; RE-AUGMENTATION OF LEFT BREAST WITH IMPLANT AND MESH LOCATED AT THE INFERIOR POLE (LEFT BREAST ONLY), THE PATIENT PRESENTED AN ELEVATED WHITE BLOOD CELL COUNT AND SHOWED SIGNS AND SYMPTOMS OF AN INFECTION - REDNESS, FEVER, MALAISE AND ACHES. THE PATIENT WAS SENT TO THE HOSPITAL, THE BREAST IMPLANT AND MESH WERE REMOVED, THE AREA FLUSHED AND A NEW IMPLANT WAS PLACED. NO MESH WAS USED IN THE SECOND SURGERY.
Patient problems: Unspecified Infection
MAUDE report 8000102 ↗InjuryNov 13, 2017Microbial Contamination of Device; Device Contamination With Biological Material; Microbial Contamination of Device; Device Contamination With Biological Material
INFECTED LEFT BREAST SEROMA / HEMATOMA INVOLVING MESH.
Patient problems: Bacterial Infection
MAUDE report 7025447 ↗InjuryNov 13, 2017Microbial Contamination of Device; Device Contamination With Biological Material; Microbial Contamination of Device; Device Contamination With Biological Material
FDA clearance and classification
The FDA regulates product code OOD (surgical film) under 21 CFR 878.3300 as a Class II device. 21 CFR 878.3300 ↗
Devices of this type normally need 510(k) clearance before sale.
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: To be implanted to reinforce soft tissue or bone where weakness exists. Intended to be used wherever temporary wound support is required, to reinforce soft tissues where weakness exists in the urological, gynecological, or gastrointestinal anatomy, or for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result. The absorbable protective film also may help minimize the potential for tissue attachment to the device in case of direct contact with the viscera.
Common questions about the GalaSHAPE 3D 100042-06
Has the GalaSHAPE 3D 100042-06 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00855920006195), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the GalaSHAPE 3D 100042-06?
The FDA's MAUDE database lists 18 adverse event reports for this device, received between Nov 13, 2017 and Nov 2, 2022. In those reports, the most common specific device problems are “Patient-Device Incompatibility” (14), “Device Contamination With Biological Material” (2) and “Microbial Contamination of Device” (2); the most common patient problems are “Bacterial Infection” (5), “Necrosis” (3) and “Seroma” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the GalaSHAPE 3D 100042-06?
None of the reports are classified as deaths, 18 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the GalaSHAPE 3D 100042-06 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the GalaSHAPE 3D 100042-06, and is it still sold?
The FDA's device database lists Tepha, Inc. as the company responsible for it (brand GalaSHAPE 3D, model 100042-06). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Surgical Film devices
- GalaFLEX Scaffold 100035-09Tepha, Inc. · Surgical Film2 deaths90 reports
- GalaFLEX Scaffold 100035-15Tepha, Inc. · Surgical Film65 reports
- GalaFORM 3D 100043-02Tepha, Inc. · Surgical Film41 reports
- Phasix 1190500Davol Inc. · surgical film39 reports
- GalaSHAPE 3D 100042-04Tepha, Inc. · Surgical Film37 reports
- GalaFORM 3D 100043-04Tepha, Inc. · Surgical Film29 reports
- Phasix 1190400Davol Inc. · surgical film26 reports
- Phasix 1190200Davol Inc. · surgical film22 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page