QL Locking Syringe Device QL38Recalls & reported problems
Atrion Medical Products, Inc. · Locking Syringe Device Atrion Part Number 96406: 38mL · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00856877002032), received between Sep 27, 2022 and Jan 26, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Defective Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Device1
- Fracture1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Unspecified Heart Problem1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 26, 2026Defective Device
A COMPLAINT WAS RECEIVED ON 9/19/2025 REPORTING: "AS REPORTED BY THE FIELD CLINICAL SPECIALIST, DURING A TRANSFEMORAL TRANSCATHETER AORTIC VALVE REPLACEMENT PROCEDURE WITH A 29MM SAPIEN 3 VALVE, THE INDEFLATOR LOCKED UP AFTER INFLATION. THE BALLOON COULD NOT BE DEFLATED USING THE INDEFLATOR AND HAD TO BE DEFLATED MANUALLY.". A PCR WAS OPENED AND AN INVESTIGATION LAUNCHED TO DETERMINE THE POTENTIAL ROOT CAUSE AND WHETHER CORRECTIVE ACTION WAS NECESSARY. A SECOND COMPLAINT FROM THE SAME LOT WAS RECEIVED ON 11/07/2025 STATING: AS REPORTED BY THE FIELD CLINICAL SPECIALIST (FCS), DURING A TRANSFEMORAL TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVR) PROCEDURE WITH A 29 MM SAPIEN 3 ULTRA RESILIA VALVE, WHILE DEPLOYING THE VALVE, THE INDEFLATOR WOULD NOT PULL NEGATIVE TO COME DOWN ON THE BALLOON. THE LOCKING MECHANISM WOULD NOT RELEASE THE THREADED PORTION OF THE INDEFLATOR PLUNGER. IT WAS CONFIRMED THAT THE ATRION QL38 WAS UNLOCKED WHEN TRYING TO PULL NEGATIVE TO DEFLATE THE BALLOON, BUT IT WOULD NOT RELEASE. THE INDEFLATOR WAS QUICKLY DISCONNECTED FROM THE DELIVERY SYSTEM, ALLOWING THE BALLOON TO PASSIVELY EMPTY. A 50 CC SYRINGE WAS QUICKLY ATTACHED TO THE DELIVERY SYSTEM STOPCOCK AND PULL
Patient problems: Unspecified Heart Problem
MAUDE report 24179841 ↗MalfunctionSep 27, 2022Fracture; Leak/Splash; Fracture; Leak/Splash
A COMPLAINT WAS RECEIVED REPORTING: "AS REPORTED BY A FIELD CLINICAL SPECIALIST (FCS), DURING VALVE DEPLOYMENT OF A 29 MM SAPIEN 3 VALVE IN THE NATIVE AORTIC VALVE VIA TRANSFEMORAL APPROACH, THE ATRION QL38 INFLATION DEVICE CRACKED WITH APPROXIMATELY 3CC STILL IN THE INFLATION DEVICE. A 20CC SYRINGE WAS CONNECTED TO THE DELIVERY SYSTEM TO COMPLETE THE DEPLOYMENT. A SECOND ATRION QL38 WAS OPEN AND USED TO INSURE THE VALVE WAS FULLY INFLATED. CURRENTLY THE DEVICE IS WITH R&D FOR CT SCANNING, AND WE PLAN TO CONDUCT SIMILAR DIAGNOSTICS AS THE PREVIOUS INVESTIGATION."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15495497 ↗
Other Syringe, Balloon Inflation devices
- Indeflator 1000184Abbott Vascular Inc. · Syringe, balloon inflation2 recalls155 reports
- Indeflator 1000186-115Abbott Vascular Inc. · Syringe, balloon inflation2 recalls124 reports
- Alliance II M00550600Boston Scientific Corporation · Syringe, balloon inflation191 reports
- Indeflator 1000185Abbott Vascular Inc. · Syringe, balloon inflation1 recall131 reports
- Indeflator 1000186Abbott Vascular Inc. · Syringe, balloon inflation2 recalls56 reports
- Encore™ 26 M001151050Boston Scientific Corporation · Syringe, balloon inflation1 recall54 reports
- PRESTO ID4030Bard Peripheral Vascular, Inc. · Syringe, balloon inflation1 death98 reports
- Encore™ 26 Advantage Kit H74904527011Boston Scientific Corporation · Syringe, balloon inflation1 recall28 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.