Skip to content

QL Locking Syringe Device QL38Recalls & reported problems

Atrion Medical Products, Inc. · Locking Syringe Device Atrion Part Number 96406: 38mL · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00856877002032), received between Sep 27, 2022 and Jan 26, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Defective Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Device1
  • Fracture1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Unspecified Heart Problem1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 26, 2026Defective Device

    A COMPLAINT WAS RECEIVED ON 9/19/2025 REPORTING: "AS REPORTED BY THE FIELD CLINICAL SPECIALIST, DURING A TRANSFEMORAL TRANSCATHETER AORTIC VALVE REPLACEMENT PROCEDURE WITH A 29MM SAPIEN 3 VALVE, THE INDEFLATOR LOCKED UP AFTER INFLATION. THE BALLOON COULD NOT BE DEFLATED USING THE INDEFLATOR AND HAD TO BE DEFLATED MANUALLY.". A PCR WAS OPENED AND AN INVESTIGATION LAUNCHED TO DETERMINE THE POTENTIAL ROOT CAUSE AND WHETHER CORRECTIVE ACTION WAS NECESSARY. A SECOND COMPLAINT FROM THE SAME LOT WAS RECEIVED ON 11/07/2025 STATING: AS REPORTED BY THE FIELD CLINICAL SPECIALIST (FCS), DURING A TRANSFEMORAL TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVR) PROCEDURE WITH A 29 MM SAPIEN 3 ULTRA RESILIA VALVE, WHILE DEPLOYING THE VALVE, THE INDEFLATOR WOULD NOT PULL NEGATIVE TO COME DOWN ON THE BALLOON. THE LOCKING MECHANISM WOULD NOT RELEASE THE THREADED PORTION OF THE INDEFLATOR PLUNGER. IT WAS CONFIRMED THAT THE ATRION QL38 WAS UNLOCKED WHEN TRYING TO PULL NEGATIVE TO DEFLATE THE BALLOON, BUT IT WOULD NOT RELEASE. THE INDEFLATOR WAS QUICKLY DISCONNECTED FROM THE DELIVERY SYSTEM, ALLOWING THE BALLOON TO PASSIVELY EMPTY. A 50 CC SYRINGE WAS QUICKLY ATTACHED TO THE DELIVERY SYSTEM STOPCOCK AND PULL

    Patient problems: Unspecified Heart Problem

    MAUDE report 24179841 ↗
  • MalfunctionSep 27, 2022Fracture; Leak/Splash; Fracture; Leak/Splash

    A COMPLAINT WAS RECEIVED REPORTING: "AS REPORTED BY A FIELD CLINICAL SPECIALIST (FCS), DURING VALVE DEPLOYMENT OF A 29 MM SAPIEN 3 VALVE IN THE NATIVE AORTIC VALVE VIA TRANSFEMORAL APPROACH, THE ATRION QL38 INFLATION DEVICE CRACKED WITH APPROXIMATELY 3CC STILL IN THE INFLATION DEVICE. A 20CC SYRINGE WAS CONNECTED TO THE DELIVERY SYSTEM TO COMPLETE THE DEPLOYMENT. A SECOND ATRION QL38 WAS OPEN AND USED TO INSURE THE VALVE WAS FULLY INFLATED. CURRENTLY THE DEVICE IS WITH R&D FOR CT SCANNING, AND WE PLAN TO CONDUCT SIMILAR DIAGNOSTICS AS THE PREVIOUS INVESTIGATION."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15495497 ↗

FDA clearance and classification

The FDA regulates product code MAV (syringe, balloon inflation) under 21 CFR 870.1650 as a Class II device. 21 CFR 870.1650 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the QL Locking Syringe Device QL38

Has the QL Locking Syringe Device QL38 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00856877002032), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the QL Locking Syringe Device QL38?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Sep 27, 2022 and Jan 26, 2026. In those reports, the most common specific device problems are “Defective Device” (1), “Fracture” (1) and “Leak/Splash” (1); the most common patient problem is “Unspecified Heart Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the QL Locking Syringe Device QL38?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the QL Locking Syringe Device QL38, and is it still sold?

The FDA's device database lists Atrion Medical Products, Inc. as the company responsible for it (brand QL Locking Syringe Device, model QL38). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Syringe, Balloon Inflation devices

All Syringe, Balloon Inflation devices →

Sources and updates