Surefire Precision Infusion System SPS-21120-35Recalls & reported problems
Surefire Medical, Inc. · 120cm, 0.021" Microcatheter with 3.5mm Expandable Tip · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00857136004088), received between Jul 19, 2016 and Jun 15, 2017 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Operates Differently Than Expected1
- Improper or Incorrect Procedure or Method1
Patient problems
- Ulcer1
- Vascular Dissection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJun 15, 2017Device Operates Differently Than Expected
THE PHYSICIAN WAS PLANNING TO USE A SUREFIRE PRECISION INFUSION SYSTEM TO PERFORM A LOWER LOBE KIDNEY EMBOLIZATION THROUGH THE RENAL ARTERY. THE PHYSICIAN TRACKED OUT TO THE DESIRED LOCATION WITH A BOSTON SCIENTIFIC RENEGADE STC MICROCATHETER AND FATHOM 016 GUIDE WIRE. THE RENEGADE MICROCATHETER WAS REMOVED AND EXCHANGED FOR THE SUREFIRE PRECISION INFUSION SYSTEM WHICH WAS TRACKED OVER THE GUIDEWIRE. WHEN THE SUREFIRE PRECISION INFUSION SYSTEM EXITED THE GUIDE CATHETER, ARTERIAL VASOSPASM WAS ENCOUNTERED. THE PHYSICIAN ATTEMPTED TO ADVANCE THE SUREFIRE PRECISION INFUSION SYSTEM THROUGH THE AREA OF VASOSPASM. THIS RESULTED IN DISSECTION OF THE VESSEL. PHYSICIAN INTERVENTION WAS REQUIRED TO STOP THE BLEEDING AND REMOVE THROMBOSIS FROM THE ARTERY. THE PATIENT WAS ADMITTED TO THE HOSPITAL OVERNIGHT FOR MONITORING AND RELEASED THE FOLLOWING DAY.
Patient problems: Vascular Dissection
MAUDE report 6645256 ↗InjuryJul 19, 2016Improper or Incorrect Procedure or Method
THE PHYSICIAN WAS USING THE SUREFIRE PRECISION INFUSION SYSTEM TO DELIVER Y90 RADIOEMBOLIZATION SPHERES TO A TUMOR IN THE LIVER. PRIOR TO INFUSION OF THE RADIOEMBOLIZATION SPHERES, THE PHYSICIAN COILED THE RIGHT GASTRIC ARTERY TO PREVENT FLOW OF EMBOLICS TO THE STOMACH. THE SUREFIRE PRECISION WAS PLACED DISTAL TO THE COILED VESSEL. DURING INFUSION NO SLOW FLOW OR REFLUX OF THE RADIOEMBOLIZATION SPHERES WAS OBSERVED. FOLLOWING THE RADIOEMBOLIZATION PROCEDURE A PET CT IMAGE SHOWED RADIATION ACTIVITY IN THE ANTRUM OF THE STOMACH. PATIENT WAS ADMITTED TO THE HOSPITAL AND DISCHARGED AFTER 5 DAYS. ENDOSCOPY SHOWED 2 SMALL ULCERS IN THE STOMACH. NO OTHER MEDICAL OR SURGICAL INTERVENTIONS WERE REPORTED. BASED UPON DISCUSSION WITH THE PHYSICIAN, USER ERROR IN DEVICE POSITIONING AND APPROPRIATE DEVICE SELECTION FOR THE VESSEL SIZE MAY HAVE CONTRIBUTED TO THE EVENT. BASED UPON NO OBSERVATION OF SLOW FLOW OR REFLUX DURING THE PROCEDURE, THE EVENT IS POTENTIALLY DUE TO A SECONDARY UNIDENTIFIED VESSEL LEADING TO STOMACH, BUT NO IMAGES OF THE PROCEDURE WERE PROVIDED TO CONFIRM.
Patient problems: Ulcer
MAUDE report 5807624 ↗
Other Catheter, Intravascular, Diagnostic devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H749085263001Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page