Operating room table - Product Usage: intended for use… recallZ-2584-2021 · Deerfield Imaging, Inc. · initiated Aug 25, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Deerfield Imaging, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

ORT200/300 rotational lock may fail to function such that it cannot change states between locked and unlocked, due to a mechanical hardware issue.

FDA root cause category: Device Design

What the company is doing

A Recall notification letter titled, "URGENT FIELD CORRECTION NOTICE" was sent to consignees on 08/25/2021 via Fed-Ex. E-mail communications were also sent to the customers on 8/26/2021. The letter instructs the consignee to fill out the second page of the notice and return it to the firm. The letter states that the IMRIS Customer Service will contact the consignee to schedule implementation of this corrective action with the facility. The letter mentions that, "IMRIS recommends these preventive actions: Per the OR Table Operator Manuals, the table should be tested daily in all standard positions and inspected for anomalies including the fit of the covers, tabletops, headrest, pendant connection and rotation. During usage, verify the table position is locked by manually pushing the table to ensure it does not exhibit rotational movement. If the rotational lock is observed to not function properly, please contact IMRIS Customer Service immediately."

Recall details

Recall number
Z-2584-2021
Status
Terminated
Initiated by the company
Aug 25, 2021
Recalling firm
Deerfield Imaging, Inc.
Quantity
13 units
Product code
KXJ
Distribution
Worldwide distribution - US Nationwide distribution FL, NH, TX, OH, CA, OR, UT, TN, NC and the countries of Canada, Australia, Sweden, Japan.

Product description

Operating room table - Product Usage: intended for use during diagnostic examinations, radiologic procedures or surgical procedures to support and position a patient.

Code information (models, lots, serial numbers)

UDI: 00857534006738 and 00857534006349 Model Number: 113821-000 (Serial Number: 10002220); Model Number: 113821-600 (Serial Numbers: 10002692, 10006152, 10006991, 10004513, 10006997, and 10003754); Model Number: 114148-000 (Serial Numbers: 10004463, 10006891, 10005189, 10005190, 10004327, and 10004997)

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 88661).