ORT300 Table (UDI 00857534006462)Recalls & reported problems
IMRIS Imaging, Inc. · ORT300 Modular Table Main Unit Assy, 114093-600 · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00857534006462), received between Feb 7, 2022 and Dec 9, 2022 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Aug 6, 2026Open, Classified
The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
Z-3034-2026 · Root cause: Nonconforming Material/Component · FDA record ↗
- Aug 25, 2021Terminated
ORT200/300 rotational lock may fail to function such that it cannot change states between locked and unlocked, due to a mechanical hardware issue.
Z-2586-2021 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak5
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 9, 2022Fluid/Blood Leak; Fluid/Blood Leak
DURING ROUTINE PREVENTIVE MAINTENANCE, AN IMRIS SERVICE ENGINEER OBSERVED LEAKAGE OF HYDRAULIC FLUID FROM A PTFE HOSE UNION WITHIN THE OPERATING ROOM TABLE. NO TABLE FUNCTIONAL ISSUES WERE OBSERVED AND NO REPORTS HAVE BEEN RECEIVED FROM THE CUSTOMER RELATING TO THE FLUID LEAK. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15955521 ↗MalfunctionNov 30, 2022Unintended Movement; Unintended Movement
THE CUSTOMER REPORTED THAT DURING THEIR MORNING QUALITY ASSURANCE / FUNCTIONAL CHECK, THE ORT300 ROTATIONAL LOCKING MECHANISM WOULD NOT LOCK THE TABLE POSITION INTO PLACE. NO PATIENT WAS INVOLVED AT THE TIME; THE CUSTOMER REPORTED AN UPCOMING CASE PLANNED FOR THE OPERATING ROOM IN WHICH THE ISSUE OCCURRED WAS MOVED INTO THE ADJACENT OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15890744 ↗MalfunctionFeb 10, 2022Fluid/Blood Leak; Fluid/Blood Leak
ON (B)(6) 2019, THE CUSTOMER REPORTED THAT HYDRAULIC FLUID LEAKS OCCURED DAILY AND DURING SURGERY FROM THE ORT300 TABLE. APPROXIMATELY 4CM IN DIAMETER OF PUDDLE FLUID LEAK OCCURRED HOWEVER; THE LEAK DID NOT AFFECT THE FUNCTION OF THE TABLE. THERE WAS NO PATIENT HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13507186 ↗MalfunctionFeb 7, 2022Fluid/Blood Leak; Fluid/Blood Leak
ON (B)(6) 2018, THE CUSTOMER REPORTED THAT THE TABLE WAS EXHIBITING A SIGNIFICANT HYDRAULIC FLUID LEAK. THERE WAS NO PATIENT HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13476411 ↗MalfunctionFeb 7, 2022Fluid/Blood Leak; Fluid/Blood Leak
DURING AN OPEN BRAIN TUMOR PROCEDURE ON (B)(6) 2019, THE ORT300 OPERATING ROOM TABLE DRIFTED IN THE LEFT ROLL POSITION AT APPROXIMATELY 8-10 DEGREES. THERE WAS A 20 MINUTE DELAY IN THE CASE WHILE THE OPERATING ROOM (OR) STAFF ATTEMPTED TO FIT THE PATIENT INTO THE SCANNER WITHOUT COMPROMISING THE FIELD OR PATIENT POSITION. THE OPERATING ROOM STAFF WAS ABLE TO COUNTERBALANCE THE ROLL MOVEMENT WITH TWO PEOPLE LEANING ON THE RIGHT SIDE OF THE OR TABLE TO KEEP IT LEVELED WHILE THE MAGNET MOVED OVER THE PATIENT. THERE WAS NO PATIENT HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13476371 ↗MalfunctionFeb 7, 2022Fluid/Blood Leak; Fluid/Blood Leak
ON (B)(6) 2018, THE CUSTOMER REPORTED THAT THERE WAS A HYDRAULIC OIL LEAK IN THE ORT300 TABLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13475587 ↗
Other Table, Radiologic devices
- NA 6265300GE Medical Systems, L.L.C. · Table, radiologic2 recalls0 reports
- PST500 4080300Baxter Medical Systems GmbH + Co. KG · Table, radiologic1 recall4 reports
- TS 7500 1730720Baxter Medical Systems GmbH + Co. KG · Table, radiologic1 recall1 report
- TS 7500 1730731Baxter Medical Systems GmbH + Co. KG · Table, radiologic1 recall0 reports
- TS 7500 1854086Baxter Medical Systems GmbH + Co. KG · Table, radiologic1 recall0 reports
- NA 6265300-20GE Medical Systems, L.L.C. · Table, radiologic1 recall0 reports
- TS 7000 1604786Baxter Medical Systems GmbH + Co. KG · Table, radiologic1 recall0 reports
- TS 7500 1854088Baxter Medical Systems GmbH + Co. KG · Table, radiologic1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.