Skip to content

ORT300 TableRecalls & reported problems across all models

IMRIS Imaging, Inc. · 2 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one ORT300 Table model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
27
Deaths reported
0
Injuries reported
0
Recalls
3

We found 2 models of ORT300 Table by IMRIS Imaging, Inc. in the FDA's device registry (GUDID). Together they are named in 27 adverse event reports in the FDA's MAUDE database, received between Feb 2, 2022 and Jan 29, 2026 — classified as 27 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (15 reports).

3 FDA recall records name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

ORT300 Table recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Aug 6, 2026Open, Classified

    The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.

    Z-3034-2026 · Root cause: Nonconforming Material/Component · names 2 of these models · Recall details · FDA record ↗

  2. Aug 25, 2021Terminated

    ORT200/300 rotational lock may fail to function such that it cannot change states between locked and unlocked, due to a mechanical hardware issue.

    Z-2586-2021 · Root cause: Device Design · Recall details · FDA record ↗

  3. Aug 25, 2021Terminated

    ORT200/300 rotational lock may fail to function such that it cannot change states between locked and unlocked, due to a mechanical hardware issue.

    Z-2585-2021 · Root cause: Device Design · Recall details · FDA record ↗

Most reported ORT300 Table problems

Device problems

  • Fluid/Blood Leak15
  • Unintended Movement11
  • Loose or Intermittent Connection4
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions27

ORT300 Table models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent ORT300 Table adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other IMRIS Imaging, Inc. products

All IMRIS Imaging, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page