072 Aspiration System FG 00806Recalls & reported problems

Q'Apel Medical LLC · As part of the 072 Aspiration System, the 072 Aspiration Catheter with a compatible suction pump is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment. · Class 2

All models of 072 Aspiration System (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00857545008097), received between Jun 21, 2024 and Jan 31, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Use of Device Problem” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 26, 2025Class IOpen, Classified

    Aspiration catheter distal tip features and characteristics may not be in the scope of FDA clearance; tip detachment, vessel rupture, and vasospasm could occur.

    Z-1512-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Use of Device Problem3
  • Compatibility Problem1

Patient problems

  • Extravasation1
  • Hemorrhage/Blood Loss/Bleeding1
  • No Clinical Signs, Symptoms or Conditions1
  • Vasoconstriction1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 31, 2025Use of Device Problem; Use of Device Problem

    RETROSPECTIVE FILING FOR AN EVENT THAT OCURED 5/2/2024. THE CASE WAS REPORTED BY Q'APEL MEDICAL'S CHIEF MARKETING OFFICER WHO WAS PRESENT FOR THE CASE AND REPORTED THE FINDINGS. THE INITIAL REVIEW BY Q'APEL MEDICAL'S MEDICAL PHYSICAN AND "QUALIFIED PERSON" DETERMINED THAT THE Q'APEL DEVICE DID NOT CONTRIBUTE TO THE INJURY AND THEREFORE, THE EVENT WAS NOT REPORTABLE. NEW INFORMATION WAS OBTAINED BY Q'APEL MEDICAL FROM THE OPERATOR OF THE DEVICE ON 1/15/2025 WHICH STATED: "THERE IS NO FLUORO LOOP OF THE 2ND PASS HIPPO NAVIGATING UP TO THE CLOT. THERE IS ONE OF THE FIRST PASS AND IT NAVIGATED SMOOTHLY WITH NO WIRE ISSUES. I CAN NOT IMAGINE HOW THAT AMOUNT OF EXTRAVASATION OCCURRED FROM A MICROWIRE PERFORATION. " THE NEW INFORMATION CONCLUDED THAT THE Q'APEL MEDICAL DEVICE LIKELY CONTRINUTED TO THE INJURY THAT REQUIRED MEDICAL INTERVENTION. CASE DATE:(B)(6) 2024. 78 YO [YEAR OLD] FEMALE PATIENT. STROKE CASE R M1 OCCLUSION. 3.5 SINCE LKW. NO TNK OR TPA ADMINISTERED DUE TO A FALL. WALRUS BGC USED AS GUIDE. 1 PASS WITH HIPPO CHEEATH WITH AN ARISTOTLE 18 GW. RESULT WAS TICI 0. 2ND PASS WITH HIPPO CHEETAH AND ARISTOTLE 18 GW. ON RUN POST ASPIRATION, VESSEL RUPTURE WITH CONTRAST EXTRAVASATIO

    Patient problems: Extravasation; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 21284855 ↗
  • InjuryJan 31, 2025Use of Device Problem; Use of Device Problem

    RETROSPECTIVE FILING FOR AN EVENT THAT OCCURRED 12/31/2023. INFORMATION WAS OBTAINED WHILE REVIEWING DATA FROM Q'APEL' S RETROSPECTIVE COHORT STUDY WITH HIPPO. FOLLOW UP INFORMATION WAS OBTAINED BY Q'APEL MEDICAL FROM THE OPERATOR OF THE DEVICE ON 1/16/2025 WHICH STATED: "HIPPO ADVANCED TO M1 FOR THROMBECTOMY AND FOLLOW UP RUN SHOWS VASOSPASM. I REVIEWED IMAGING. AFTER 1ST PASS THERE IS MINIMAL VASOSPASM I WOULD HAVE TO SAY LIKELY CAUSED BY HIPPO (VASOSPASM I DONT THINK IT WAS CLINICALLY SIGNIFICANT IN REVIEW HOWEVER THE OPERATOR ADMINISTERED 8 MG VERAPAMIL TO TREAT THE VASOSPASM. FINAL TICI SCORE WAS 2C". VESSULE OCCLUDED-LMCA M1, FIRST PASS TICI SCORE-0, FINAL PASS TICI SCORE-2C. REPORTED DEVICES INVOLVED IN THE PROCEDURE: ZOOM 88, HIPPO, ZOOM 71, ZOOM 35, ZOOM POD.

    Patient problems: Vasoconstriction

    MAUDE report 21284834 ↗
  • MalfunctionJun 21, 2024Use of Device Problem; Compatibility Problem; Use of Device Problem; Compatibility Problem

    ASPIRATION SYSTEM (HIPPO) WAS BEING REMOVED/BACKED OUT OF THE ZOOM RDL FOLLOWING ASPIRATION. DR (B)(6) ATTEMPTED TO PASS A ZOOM 71 THROUGH THE ZOOM RDL ON A SECOND PASS AND IT WOULD NOT ADVANCE THROUGH. THE ZOOM RDL AND ZOOM 71 WERE THEN COMPLETELY REMOVED FROM THE BODY IN WHICH IT WAS DISCOVERED THAT THERE WAS A PIECE OF THE HIPPO INSIDE THE RDL TOWARDS THE DISTAL TIP. NO PART OF THE HIPPO TIP REMAINED IN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19583959 ↗

FDA clearance and classification

Cleared through 510(k) K222786 on Aug 25, 2023 (decision: Substantially Equivalent). Submitted by Q'Apel Medical, Inc. as “072 Aspiration System”. The FDA review took 344 days. FDA record ↗

The FDA regulates product code NRY (catheter, thrombus retriever) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗

FDA definition of this device type: The device is intended to restore blood flow by removing thrombus/clots in patients experiencing ischemic stroke. The prior clearances in this category have been limited to simply identifying catheter placed in the peripheral, coronary, and neurovasculature. It is felt that this is a unique claim and should not be combined with previously cleared catheters under a general procode.

Common questions about the 072 Aspiration System FG 00806

Has the 072 Aspiration System FG 00806 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1512-2025 (Class I), began on Feb 26, 2025: Aspiration catheter distal tip features and characteristics may not be in the scope of FDA clearance; tip detachment, vessel rupture, and vasospasm could occur. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the 072 Aspiration System FG 00806?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Jun 21, 2024 and Jan 31, 2025. In those reports, the most common specific device problems are “Use of Device Problem” (3) and “Compatibility Problem” (1); the most common patient problems are “Extravasation” (1), “Hemorrhage/Blood Loss/Bleeding” (1) and “Vasoconstriction” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the 072 Aspiration System FG 00806?

None of the reports are classified as deaths, 2 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the 072 Aspiration System FG 00806 FDA approved?

It was cleared through 510(k) K222786 on Aug 25, 2023. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the 072 Aspiration System FG 00806, and is it still sold?

The FDA's device database lists Q'Apel Medical LLC as the company responsible for it (brand 072 Aspiration System, model FG 00806). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, Thrombus Retriever devices

All Catheter, Thrombus Retriever devices →