072 Aspiration System recallZ-1512-2025 · Qapel Medical Inc. · initiated Feb 26, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Qapel Medical Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Aspiration catheter distal tip features and characteristics may not be in the scope of FDA clearance; tip detachment, vessel rupture, and vasospasm could occur.

FDA root cause category: Device Design

What the company is doing

On 2/26/2025, removal and discontinuation notices were mailed to customers were asked to do the following: 1) Quarantine affected devices and return to the firm. 2) If you have provided this product to others, let the firm know so they can notify them of this action and retrieve any remaining product. 3) Complete and return the Acknowledgement and Response Card via email to [email protected] If you have any questions regarding this letter, please contact customer service for further information at [email protected], or by phone at 510-738-6255 between the hours of 8am - 5pm PST.

Recall details

Recall number
Z-1512-2025
Status
Open, Classified
Initiated by the company
Feb 26, 2025
Posted by the FDA
Apr 7, 2025
Recalling firm
Qapel Medical Inc.
Quantity
1617
Product code
NRY
FDA submissions
K222786
Distribution
US: CA, IL, NE, TN, TX, NJ, NC, SD, FL, MI, PA, GA, AL, IN, NY, WI, AZ, MN, LA, OH, OK, SC, NV, ND, ID, MO, CT, WA, MA. OUS: QATAR, United Arab Emirates, Republic of Kazakhstan

Product description

Hippo 072 Aspiration System including Cheetah Delivery Tool and Aspiration Tubing, REF: APT6072-132. Hippo 072 Aspiration System including Cheetah Delivery Tool, REF: AP6072-132. 072 Aspiration Tubing, REF: APT-95

Code information (models, lots, serial numbers)

REF/UDI-DI/Lot(Expiration): APT6072-132/00857545008127/FG241008C-03(04/07/2025), FG240916C-04(03/17/2025), FG240905C-04(03/06/2025). AP6072-132/00857545008097/FG241206A-03(06/08/2025), FG240917A-01(03/17/2025). APT-95/00857545008103/FG241206A-04(06/08/25)