SynCardia Freedom® Onboard Battery 295025-001Recalls & reported problems

Syncardia Systems, LLC · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This SynCardia Freedom onboard battery is the portable power unit for the SynCardia artificial heart driver. Most of the 61 reports describe batteries that will not charge, will not hold a charge, or show a full charge but fail to power the driver. Two injuries were reported and no deaths. There are no recalls naming this device. Reports rose from none in the prior 12 months to 4 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
62
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 62 adverse event reports that list this exact device (DI 00858000003084), received between Dec 8, 2017 and Sep 16, 2026 — classified as 2 injuries, 60 malfunctions.

The most frequently reported specific device problem is “Battery Problem” (50 reports).

5 other recalls cover devices cleared under the same FDA submission (P030011); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 4 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Battery Problem50
  • Charging Problem9
  • Failure to Charge6
  • Appropriate Device Problem Term/Code Not Available4
  • Mechanical Problem3
  • Audible Prompt/Feedback Problem2
  • No Apparent Adverse Event2
  • Output Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions40
  • No Consequences Or Impact To Patient19
  • No Patient Involvement9
  • Loss of consciousness2
  • No Known Impact Or Consequence To Patient2
  • Patient Problem/Medical Problem1

Recent adverse event reports

The 25 most recent of 62 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Charging Problem

    Per hospital staff, Freedom Onboard Battery S/N (b)(6) will not charge. Battery has been placed in multiple charging ports on charger and reportedly will not charge. No adverse impact to patient reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50790148 ↗
  • MalfunctionMar 3, 2026Charging Problem; Output Problem

    PER HOSPITAL STAFF, ONBOARD BATTERY WILL NOT HOLD A CHARGE OR FUNCTION IN A FREEDOM DRIVER. DEVICE WAS NOT IN PATIENT USE AT TIME OF COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24500016 ↗
  • MalfunctionFeb 18, 2026Output Problem

    PER HOSPITAL STAFF, ONE OF THE ONBOARD FREEDOM DRIVER BATTERIES THAT CAME IN THE DISCHARGE KIT HAS THE ISSUE WHERE IT SHOWS AS BEING FULLY CHARGED BUT WILL NOT TURN ON A DRIVER. THE ISSUE IS WITH BATTERY SN (B)(6). BATTERY NOT IN PATIENT USE AT TIME OF COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24379352 ↗
  • MalfunctionFeb 12, 2026Charging Problem

    PER HOSPITAL STAFF, PATIENT WAS SEEN IN THE CLINIC AND BROUGHT TWO FREEDOM DRIVER BATTERIES AND STATED THAT THEY WILL NOT CHARGE. NO ADVERSE IMPACT TO PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24326453 ↗
  • MalfunctionFeb 11, 2026Charging Problem

    PER HOSPITAL STAFF, PATIENT WAS SEEN IN THE CLINIC AND BROUGHT TWO FREEDOM DRIVER BATTERIES AND STATED THAT THEY WILL NOT CHARGE. NO ADVERSE IMPACT TO PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24317740 ↗
  • MalfunctionSep 23, 2024Battery Problem; Battery Problem

    PER DISTRIBUTOR, BATTERIES ARE DEFECTIVE. NO REPORTED ADVERSE IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20288078 ↗
  • MalfunctionSep 18, 2024Battery Problem; Battery Problem

    PER DISTRIBUTOR, BATTERIES ARE DEFECTIVE. NO REPORTED ADVERSE IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20249086 ↗
  • MalfunctionSep 18, 2024Battery Problem; Battery Problem

    PER DISTRIBUTOR, BATTERIES ARE DEFECTIVE. NO ADVERSE IMPACT TO PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20248919 ↗
  • MalfunctionSep 18, 2024Battery Problem; Battery Problem

    PER DISTRIBUTOR, BATTERIES ARE DEFECTIVE. NO REPORTED ADVERSE IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20248588 ↗
  • MalfunctionOct 21, 2022Battery Problem; Battery Problem

    THE REPORTED ISSUES INVOLVE THE FOLLOWING SYNCARDIA TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T) SYSTEM COMPONENTS AND ARE REPORTED UNDER THREE SEPARATE MEDICAL DEVICE REPORTS: (1) FREEDOM ONBOARD BATTERY S/N (B)(4) (MFR REPORT # 3003761017-2022-00109); (2) FREEDOM ONBOARD BATTERY S/N (B)(4) (MFR REPORT # 3003761017-2022-00110); (3) FREEDOM BATTERY CHARGER S/N (B)(4)). THE FREEDOM ONBOARD BATTERIES AND FREEDOM BATTERY CHARGER WERE NOT SUPPORTING A PATIENT. THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THE FREEDOM BATTERY CHARGER DID NOT CHARGE ANY FREEDOM ONBOARD BATTERIES. ADDITIONALLY, TWO FREEDOM ONBOARD BATTERIES WERE UNABLE TO CHARGE. THE CUSTOMER IS UNCLEAR IF THESE ISSUES ARE RELATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15652020 ↗
  • MalfunctionOct 21, 2022Battery Problem; Battery Problem

    DURING ROUTINE EVALUATION, A SYNCARDIA TECHNICIAN REPORTED THAT THE FREEDOM ONBOARD BATTERY SYSTEM MANAGEMENT (SMBUS) DATA WAS NOT ACTIVE WITHOUT POWER, THE CHARGE INDICATOR LEDS DID NOT ILLUMINATE WITHOUT POWER AND THE BATTERY DID NOT POWER A FREEDOM DRIVER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15651930 ↗
  • MalfunctionOct 19, 2022Battery Problem; Battery Problem

    THE REPORTED ISSUES INVOLVE THE FOLLOWING SYNCARDIA TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T) SYSTEM COMPONENTS AND ARE REPORTED UNDER THREE SEPARATE MEDICAL DEVICE REPORTS: (1) FREEDOM ONBOARD BATTERY S/N (B)(4) (MFR REPORT # 3003761017-2022-00109); (2) FREEDOM ONBOARD BATTERY S/N (B)(4) (MFR REPORT # 3003761017-2022-00110); (3) FREEDOM BATTERY CHARGER S/N (B)(4)). THE FREEDOM ONBOARD BATTERIES AND FREEDOM BATTERY CHARGER WERE NOT SUPPORTING A PATIENT. THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THE FREEDOM BATTERY CHARGER DID NOT CHARGE ANY FREEDOM ONBOARD BATTERIES. ADDITIONALLY, TWO FREEDOM ONBOARD BATTERIES WERE UNABLE TO CHARGE. THE CUSTOMER IS UNCLEAR IF THESE ISSUES ARE RELATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15631917 ↗
  • MalfunctionOct 13, 2022Battery Problem; Battery Problem

    THE FREEDOM ONBOARD BATTERY WAS NOT SUPPORTING A PATIENT AT THE TIME OF THE EVENT. THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THAT DURING A TRAINING EVENT THE FREEDOM ONBOARD BATTERY DID NOT HAVE A CHARGE. THE HOSPITAL ATTEMPTED TO CHARGE THE BATTERY WITHOUT SUCCESS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15598825 ↗
  • MalfunctionOct 13, 2022Battery Problem; Battery Problem

    THE FREEDOM ONBOARD BATTERY WAS NOT SUPPORTING A PATIENT AT THE TIME OF THE EVENT. THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THAT DURING A TRAINING EVENT THE FREEDOM ONBOARD BATTERY DID NOT HAVE A CHARGE. THE HOSPITAL ATTEMPTED TO CHARGE THE BATTERY WITHOUT SUCCESS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15598762 ↗
  • MalfunctionSep 7, 2022Battery Problem; Battery Problem

    THE FREEDOM ONBOARD BATTERY WAS NOT SUPPORTING A PATIENT AT THE TIME OF THE EVENT. THE CUSTOMER, A (B)(6) HOSPITAL, REPORT THAT THE FREEDOM ONBOARD BATTERY DID NOT HOLD A CHARGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15375092 ↗
  • MalfunctionAug 31, 2022Battery Problem; Battery Problem

    THE FREEDOM ONBOARD BATTERY WAS NOT SUPPORTING A PATIENT AT THE TIME OF THE EVENT. THE CUSTOMER, A SYNCARDIA AUTHORIZED DISTRIBUTOR, REPORTED THAT THE FREEDOM ONBOARD BATTERY DID NOT CHARGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15332324 ↗
  • MalfunctionJun 30, 2022Battery Problem; Battery Problem

    THE FREEDOM ONBOARD BATTERY WAS NOT SUPPORTING A PATIENT. THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THAT THE FREEDOM ONBOARD BATTERY DID NOT WORK/CHARGE TO ITS FULLEST CAPACITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14877779 ↗
  • MalfunctionJan 10, 2022Battery Problem; Appropriate Device Problem Term/Code Not Available; Battery Problem; Appropriate Device Problem Term/Code Not Available

    THE REPORTED ISSUES INVOLVE THE FOLLOWING SYNCARDIA TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T) SYSTEM COMPONENTS AND ARE REPORTED UNDER TWO SEPARATE MEDICAL DEVICE REPORTS: (1) FREEDOM ONBOARD BATTERY S/N (B)(4) (MFR REPORT # 3003761017-2022-00001 AND (2) FREEDOM ONBOARD BATTERY S/N (B)(4) (MFR REPORT # 3003761017-2022-00002). THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THAT THE PATIENT WAS AT HOME WHEN THE FREEDOM DRIVER STARTED OVERHEAT AND EXHIBITED AN ALARM. THE PATIENT ALSO REPORTED THAT HE EXCHANGED THE FREEDOM DRIVER ONBOARD BATTERIES AND THE DRIVER STOPPED ALARMING. THE PATIENT PUT THE FREEDOM ONBOARD BATTERIES, THAT WERE IN THE DRIVER, INTO THE BATTERY CHARGER AND REPORTED THEY WOULD NOT FULLY CHARGE. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13214910 ↗
  • MalfunctionJan 10, 2022Battery Problem; Appropriate Device Problem Term/Code Not Available; Battery Problem; Appropriate Device Problem Term/Code Not Available

    THE REPORTED ISSUES INVOLVE THE FOLLOWING SYNCARDIA TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T) SYSTEM COMPONENTS AND ARE REPORTED UNDER TWO SEPARATE MEDICAL DEVICE REPORTS: (1) FREEDOM ONBOARD BATTERY S/N (B)(4) (MFR REPORT # 3003761017-2022-00001 AND (2) FREEDOM ONBOARD BATTERY S/N (B)(4) (MFR REPORT # 3003761017-2022-00002). THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THAT THE PATIENT WAS AT HOME WHEN THE FREEDOM DRIVER STARTED OVERHEAT AND EXHIBITED AN ALARM. THE PATIENT ALSO REPORTED THAT HE EXCHANGED THE FREEDOM DRIVER ONBOARD BATTERIES AND THE DRIVER STOPPED ALARMING. THE PATIENT PUT THE FREEDOM ONBOARD BATTERIES, THAT WERE IN THE DRIVER, INTO THE BATTERY CHARGER AND REPORTED THEY WOULD NOT FULLY CHARGE. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13214757 ↗
  • MalfunctionJan 10, 2022Battery Problem; Battery Problem

    THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THAT THE PATIENT'S FREEDOM ONBOARD BATTERY WAS NOT HOLDING A CHARGE FOR AS LONG AS THE PATIENT EXPECTED. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13204995 ↗
  • MalfunctionJan 10, 2022Battery Problem; Battery Problem

    THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THAT THE PATIENT'S FREEDOM ONBOARD BATTERY WAS NOT HOLDING A CHARGE FOR AS LONG AS THE PATIENT EXPECTED. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13204988 ↗
  • MalfunctionNov 8, 2021Battery Problem; Appropriate Device Problem Term/Code Not Available; Battery Problem; Appropriate Device Problem Term/Code Not Available

    THE FREEDOM ONBOARD BATTERY WAS NOT SUPPORTING A PATIENT. THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THAT ONCE THE FREEDOM ONBOARD BATTERY WAS INSERTED INTO A FREEDOM DRIVER, THE DRIVER ANNUNCIATED AN ALARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12775521 ↗
  • MalfunctionSep 3, 2021Battery Problem; No Apparent Adverse Event; Battery Problem; No Apparent Adverse Event

    THE CUSTOMER, A SYNCARDIA AUTHORIZED DISTRIBUTOR, REPORTED THAT PATIENT'S FREEDOM ONBOARD BATTERY DISCHARGED IMMEDIATELY AS SOON AS IT WAS INSERTED INTO THE FREEDOM DRIVER AFTER A COMPLETE CHARGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12422262 ↗
  • MalfunctionSep 3, 2021Battery Problem; Battery Problem

    THE CUSTOMER, A SYNCARDIA AUTHORIZED DISTRIBUTOR, REPORTED THAT PATIENT'S FREEDOM ONBOARD BATTERY DISCHARGED IMMEDIATELY AS SOON AS IT WAS INSERTED INTO THE FREEDOM DRIVER AFTER A COMPLETE CHARGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12422255 ↗
  • MalfunctionAug 16, 2021Battery Problem; No Apparent Adverse Event; Battery Problem; No Apparent Adverse Event

    THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THAT THE PATIENT DROVE TO THE HOSPITAL FOR CLINIC VISIT AND WHILE ON WALL POWER THE FREEDOM DRIVER GAVE A LOW BATTERY ALARM. THE FREEDOM ONBOARD BATTERY WAS EXCHANGED AND THE ALARM RESOLVED. WHEN THE BATTERY WAS TESTED OUTSIDE OF THE DRIVER IT SHOWED NO CHARGE. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12329109 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030011. One submission often covers many models, so check the product and lot details in each record.

  • Z-0126-2024 · Aug 1, 2023 · Completed
    SynCardia 50cc TAH-t, Clinical, REF 570500, Sterile EO
  • Z-1022-2023 · Nov 2, 2022 · Open, Classified
    SynCardia Total Artificial Heart System (70CC): 500101-001 SynCardia Total Artificial Heart System (50CC): 570500-000
  • Z-0659-2017 · Oct 21, 2016 · Terminated
    5000 Series Freedom Drivers. Freedom Driver System for Temporary Total Artificial Heart (TAH-t). Part number 595000-001
  • Z-2731-2015 · Aug 6, 2015 · Terminated
    Temporary Total Artificial Heart (TAH-t) with Freedom Driver System REF 595000-001 Product Usage: Freedom Drivers are components of the SynCardia temporary Total Artificial Hear
  • Z-0620-2013 · Dec 5, 2012 · Terminated
    Companion External Battery, Part Number: 293001-001 The Companion External Battery is a component of the Companion 2 Driver System and the Companion Driver System. Both Driver Sys