Skip to content

SynCardia Companion Hospital Cart 397003-001Recalls & reported problems

Syncardia Systems, LLC · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00858000003138), received between Jun 21, 2022 and Aug 7, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Display or Visual Feedback Problem” (7 reports).

5 other recalls cover devices cleared under the same FDA submission (P030011); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Display or Visual Feedback Problem7
  • Appropriate Device Problem Term/Code Not Available1
  • Electrical /Electronic Property Problem1
  • No Display/Image1
  • Power Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    PER HOSPITAL STAFF, THE HOSPITAL CART SCREEN TURNED PINK, WHEN BOTTOM LEFT CORNER OF THE SCREEN WAS BUMPED IT WOULD FLASH AND GLITCH. TOUCH SCREEN WAS NOT WORKING. NO ADVERSE IMPACT TO PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22743913 ↗
  • MalfunctionJun 18, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    THE CUSTOMER, A SYNCARDIA AUTHORIZED DISTRIBUTOR, REPORTED THE COMPANION HOSPITAL CART DISPLAY SCREEN HAD A PIXILATED MAGENTA HUE, AND THAT THE FUNCTIONALITY OF THE DRIVER WAS NOT OTHERWISE AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19564040 ↗
  • MalfunctionMay 3, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    PER HOSPITAL STAFF, TOUCH SCREEN ON THE HOSPITAL CART DISPLAY DOES NOT RESPOND APPROPRIATELY. MULTIPLE ATTEMPTS AT CALIBRATING THE SCREEN WERE UNSUCCESSFUL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19236900 ↗
  • MalfunctionFeb 29, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    PER HOSPITAL STAFF, THE HOSPITAL CART SCREEN WAS BLACK WHEN THE COMPANION 2 DRIVER WAS CONNECTED. THE TOUCH SCREEN STILL FUNCTIONED BUT THE SCREEN DID NOT LIGHT UP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18812823 ↗
  • MalfunctionOct 9, 2023No Display/Image; Appropriate Device Problem Term/Code Not Available; No Display/Image; Appropriate Device Problem Term/Code Not Available

    MONITOR/SCREEN OF HOSPITAL CART IS BLANK. C2 DRIVER SCREEN DOES NOT PROJECT TO HOSPITAL CART SCREEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17901725 ↗
  • MalfunctionMay 4, 2023Power Problem; Power Problem

    THE CUSTOMER, A SYNCARDIA AUTHORIZED DISTRIBUTOR, REPORTED THE HOSPITAL CART SWITCHED TO BATTERY POWER WHILE CONNECTED TO AC POWER. THE POWER CABLE WAS REPLACED BUT THE ISSUE PERSISTED WITH THE CART.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16872953 ↗
  • MalfunctionJan 31, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem

    THE CUSTOMER, A CERTIFIED SYNCARDIA HOSPITAL, REPORTED THAT THERE WAS DIFFICULTY IN GETTING TOUCH SCREEN TO RECOGNIZE TOUCH, THAT THE SCREEN WOULD NOT CALIBRATE, AND WHEN DRIVER WAS REMOVED FROM THE HOSPITAL CART, THE DRIVER'S SCREEN FUNCTIONED AS INTENDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16273443 ↗
  • MalfunctionOct 13, 2022Display or Visual Feedback Problem; Display or Visual Feedback Problem

    THE COMPANION HOSPITAL CART WAS NOT SUPPORTING A PATIENT AT THE TIME OF THE EVENT. THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THAT THE DISPLAY ON THE COMPANION HOSPITAL CART WAS "FROZEN", I.E., NOT RECOGNIZING TOUCH. THE HOSPITAL STAFF COULD NOT RECALIBRATE THE SCREEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15601296 ↗
  • MalfunctionJun 21, 2022Display or Visual Feedback Problem; Electrical /Electronic Property Problem; Display or Visual Feedback Problem; Electrical /Electronic Property Problem

    DURING ROUTINE EVALUATION, A SYNCARDIA TECHNICIAN REPORTED THAT THE COMPANION HOSPITAL CART POWER CORD ICON DID NOT APPEAR ON THE SCREEN AS DESIGNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14757063 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030011. One submission often covers many models, so check the product and lot details in each record.

  • Z-0126-2024 · Aug 1, 2023 · Completed
    SynCardia 50cc TAH-t, Clinical, REF 570500, Sterile EO
  • Z-1022-2023 · Nov 2, 2022 · Open, Classified
    SynCardia Total Artificial Heart System (70CC): 500101-001 SynCardia Total Artificial Heart System (50CC): 570500-000
  • Z-0659-2017 · Oct 21, 2016 · Terminated
    5000 Series Freedom Drivers. Freedom Driver System for Temporary Total Artificial Heart (TAH-t). Part number 595000-001
  • Z-2731-2015 · Aug 6, 2015 · Terminated
    Temporary Total Artificial Heart (TAH-t) with Freedom Driver System REF 595000-001 Product Usage: Freedom Drivers are components of the SynCardia temporary Total Artificial Hear
  • Z-0620-2013 · Dec 5, 2012 · Terminated
    Companion External Battery, Part Number: 293001-001 The Companion External Battery is a component of the Companion 2 Driver System and the Companion Driver System. Both Driver Sys

Other Artificial Heart devices

All Artificial Heart devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page