KneeAlign 2 133631Recalls & reported problems
Orthalign, Inc. · KneeAlign 2 Unit · Class 2
All models of KneeAlign 2 (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00858704006008), received between Jun 30, 2017 and May 5, 2022 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings3
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Joint Disorder1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE POD WAS NOT CORRESPONDENCE WITH THE SENSOR OF THE KNEEALIGN. THE READING ON THE SCREEN WAS SHOWN AS 13 DEGREE FLEXION, WITH SURGEON CHECKED ON THE RESECTION GUIDE AND IT'S NOT. THE ISSUE WAS COMPLETED WITH OPEN A NEW KNEEALIGN2 CONSUMABLE AND CHANGE OF THE KNEEALIGN SENSOR. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14286500 ↗MalfunctionApr 16, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT ALL CALIBRATION WAS DONE AND PASSED. KNEEALIGN POD HAS REGISTERED FEMUR OFFSET AND CENTRE OF HIP ROTATION PERFORMED BY SURGEON. SURGEON ADJUSTED FINAL ANGLE AS VARUS 1 DEGREE FOR FEMURCUTTING BLOCK TO SIT HOWEVER UPON REMOVING THE KNEEALIGN POD AND SENSOR, AN OBVIOUS VALGUS CUT WAS SHOWN WHEN THE SURGEON DOUBLE CHECK WITH A RESECTION GUIDE PRIOR SAWING THE FEMUR. SURGEON ABORT THE USAGE OF KNEEALIGN IN ASSISTING FOR FEMUR CUTS AND PROCEED ON USING CONVENTIONAL JIG INSTEAD. SURGEON THEN ATTEMPT TO USE KNEE-ALIGN AGAIN FOR TIBIAL CUTS, SAME OUTCOME AS ABOVE MENTIONED. SURGEON DECIDED TO PROCEED WITH CONVENTIONAL SAWING WITHOUT THE AID OF NAVIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11681557 ↗InjuryJun 30, 2017Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE SURGEON DID MAKE HIS CUTS WITH THE UNIT. HE DID NOT AGREE WITH THE FEMORAL RESECTION (IT WAS A EXCESSIVE VALGUS ANGLE) AND HAD TO REVISE HIS CUT WITH THE TRADITIONAL RESECTION TOOLS.
Patient problems: Joint Disorder
MAUDE report 6682593 ↗
Other Orthopedic Stereotaxic Instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026