Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00858704006534), received between Nov 8, 2017 and Oct 12, 2023 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Communication or Transmission Problem1
- Device Sensing Problem1
- Incorrect Measurement1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- Bone Fracture(s)2
- No Clinical Signs, Symptoms or Conditions2
- Joint Laxity1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryOct 12, 2023Incorrect Measurement; Incorrect Measurement
EVERYTHING CALIBRATED FINE, BUT THE CUT ON THE FEMUR WAS SIGNIFICANTLY OFF. THE SURGERY WAS COMPLETED WITH STANDARD INSTRUMENTS.
Patient problems: Joint Laxity
MAUDE report 17920461 ↗MalfunctionOct 29, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE SUMMARY SCREEN FLEXION EXTENSION NUMBERS DID NOT MATCH THE PREVIOUS ORIGINALLY REGISTERED FLEXION EXTENSION NUMBERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12719478 ↗MalfunctionAug 5, 2021Communication or Transmission Problem; Communication or Transmission Problem
AN ORTHALIGN PLUS NAVIGATION UNIT SET THE DISTAL FEMORAL VALGUS ANGLE SO FAR OFF THAT IT'S BEEN VISIBLY INCORRECT FROM A DISTANCE (15 DEGREES OR MORE). REP CHECKED AND ASSISTED NURSE WITH SET UP AND CALIBRATION, ALL STEPS FOLLOWED AS PER TECHNIQUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12285342 ↗MalfunctionMay 16, 2018Device Sensing Problem; Adverse Event Without Identified Device or Use Problem; Device Sensing Problem; Adverse Event Without Identified Device or Use Problem
ORTHALIGN NAVIGATION UNIT SHOWED A SYSTEM FAULT 6 DURING CALIBRATION, PRIOR TO PATIENT USE. WHEN OBTAINING THE CUP ANGLE THE NAVIGATION UNIT SCREEN DISPLAYED "SYSTEM UNSTABLE" AND WOULD NOT DISPLAY ANGLES. WHEN OBTAINING LEG LENGTH THE NAVIGATION UNIT DISPLAYED A MEASUREMENT OF 141 MM SHORTER AT THE END OF THE CASE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7520582 ↗InjuryNov 8, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SURGEON REPORTED THAT A FRACTURE IN THE TIBIA OCCURRED DURING PLACEMENT OF THE TIBIAL CUTTING BLOCK DURING AN UNICOMPARTMENTAL KNEE ARTHROPLASTY PROCEDURE NAVIGATED BY THE ORTHALIGN PLUS SYSTEM. THE FRACTURE REQUIRED USE OF A SMALL FRAGMENT PLATE AND SCREWS TO STABILIZE THE PROXIMAL TIBIA BELOW THE RESECTION SITE.
Patient problems: Bone Fracture(s)
MAUDE report 7012102 ↗InjuryNov 8, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SURGEON REPORTED THAT A PATIENT WHO HAD UNDERGONE AN UNICOMPARTMENTAL KNEE ARTHROPLASTY PROCEDURE NAVIGATED BY THE ORTHALIGN PLUS SYSTEM RETURNED THREE AND A HALF MONTHS LATER WITH A FRACTURE IN THE TIBIA BELOW THE TIBIAL RESECTION. THE FRACTURE REQUIRED A REVISION PROCEDURE WITH A LONG TIBIAL STEM TO CONSTRAIN THE ARTHROPLASTY COMPONENTS AND BYPASS THE FRACTURE SITE.
Patient problems: Bone Fracture(s)
MAUDE report 7011820 ↗
Other Orthopedic Stereotaxic Instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026