ProxiCor for Cardiac Tissue Repair CMCV-064-401Recalls & reported problems
Elutia Med LLC · ProxiCor for Cardiac Tissue Repair is indicated for use as an intracardiac patch or pledget for tissue repair and suture-line buttressing. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00859389005041), received between Jan 13, 2017 and Sep 30, 2020 — classified as 1 death, 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Delamination” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Delamination1
- Patient-Device Incompatibility1
Patient problems
- Aortic Insufficiency1
- Cardiac Arrest1
- Drug Resistant Bacterial Infection1
- Foreign Body Reaction1
- Inflammation1
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
DeathSep 30, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON 6/18/2020, AZIYO BIOLOGICS CONTRACT MANUFACTURER NOTIFIED AZIYO OF A RECENTLY ACCEPTED FOR PRINT ARTICLE IN PACING AND CLINICAL ELECTROPHYSIOLOGY (PACE) [REFERENCE: PACING CLIN ELECTROPHYSIOL. 2020 JUN 16. DOI: 10.1111/PACE.13988. ONLINE AHEAD OF PRINT]. THE ARTICLE TITLED "SURGICAL AND ENDOSCOPIC MANAGEMENT OF A PERICARDIOESOPHAGEAL FISTULA AFTER RADIOFREQUENCY PULMONARY VEIN ISOLATION MANAGEMENT OF A PERICARDIOESOPHAGEAL FISTULA" WAS EPUBLISHED ON JUNE 16, 2020. THIS ARTICLE WAS DISCOVERED DURING THE CONTRACT MANUFACTURER'S LITERATURE REVIEW CYCLE AND REVIEWED IN SUPPORT OF AZIYO BIOLOGICS POST MARKET SURVEILLANCE PROCESS. THE CASE STUDY REPORTS THE EVENTS RELATIVE TO A (B)(6) YEAR OLD MALE PATIENT HAVING PREVIOUSLY UNDERGONE AN RF PULMONARY VEIN ISOLATION PROCEDURE 3 WEEKS PRIOR, PRESENTED WITH CHEST PAIN. CTA AND ECHOCARDIOGRAM SHOWED PNEUMOPERICARDIUM AND BARIUM ESOPHAGRAM SHOWED EXTRAVASATION FROM ESOPHOGUS INTO THE PERICARDIUM WITHOUT CONNECTION TO THE LEFT ATRIUM. PERICARDIAL DEFECT WAS IDENTIFIED AND A CORMATRIX (NOW AZIYO BIOLOGICS INC.) PROXICOR ECM PATCH WAS PLACED (MODEL #: UNKNOWN, LOT #: UNKNOWN - LIKELY PROXICOR FOR CARDIAC TISSUE REPAIR). THE FISTULA RESULTING F
Patient problems: Cardiac Arrest; Drug Resistant Bacterial Infection
MAUDE report 10608322 ↗InjuryAug 7, 2019Patient-Device Incompatibility; Patient-Device Incompatibility
IT WAS REPORTED THAT A MALE PATIENT UNDERWENT A COMPLEX AORTIC VALVE REPAIR ON (B)(6) 2016. THE CORMATRIX ECM FOR CARDIAC TISSUE REPAIR WAS SOAKED IN A SALINE SOLUTION FOR 1 MINUTE. ON (B)(6) 2019 (PATIENT REPORTED AS (B)(6) YEARS OLD AT TIME OF EVENT) DUE TO RECURRENT CENTRAL AORTIC INSUFFICIENCY HE NEEDED TO BE RE-OPERATED FOR A ROSS PROCEDURE. A REPLACEMENT OF THE AORTIC VALVE WITH PULMONARY AUTOGRAFT WAS COMPLETED. IT CAN BE NOTED THAT THE PATIENT HAS DYSPLASTIC AORTIC VALVE AND COULD HAVE CONTRIBUTED TO THE EVENT. PROBABLE CAUSE OF THE EVENT WAS LISTED AS DELAMINATION OF THE ECM CAUSING CALCIFICATION AND LEAFLET IMMOBILITY. THE SITE STATES THAT THE EVENT IS PROBABLY RELATED TO THE PROCEDURE AND PROBABLY RELATED TO THE ECM DEVICE. THE DELAMINATION OF THE ECM MATERIAL WAS THOUGHT TO BE IN RELATION TO THE PATIENT COMORBIDITIES AND BOUGHT THE PATIENT MORE TIME BEFORE THE NEXT PROCEDURE (ROSS). THE PHYSICIANS WERE SATISFIED WITH THE OUTCOME.
Patient problems: Aortic Insufficiency; Foreign Body Reaction; Inflammation
MAUDE report 8870610 ↗MalfunctionJan 13, 2017Delamination; Delamination
IT WAS REPORTED THAT ON (B)(6) 2016, A 7CM X 10CM CORMATRIX ECM FOR CARDIAC TISSUE REPAIR (CMCV-064-401 LOT# M16H1192) WAS BEING PREPARED FOR A PULMONIC VALVE APPLICATION (OFF-LABEL USE) ON THE "DRY TABLE'. THE 7CM X 10CM PIECE OF TISSUE WAS OBSERVED TO BE UNUSUALLY THICK, AND THEN THE TISSUE SEPARATED INTO TWO LAYERS. THE ECM TISSUE HAD NOT BEEN HYDRATED AT THE TIME OF OBSERVING THE DELAMINATION. THE SURGEON ABANDONED CREATING THE VALVE WITH THIS PIECE OF TISSUE, AND PROCEEDED TO TAKE ANOTHER 7CM X 10CM PACKAGE OF CORMATRIX ECM FOR CARDIAC TISSUE REPAIR FROM THE SHELF AND COMPLETE THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient; No Patient Involvement
MAUDE report 6251921 ↗
Other Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene devices
- CryoPatch® SG Pulmonary Human Cardiac Branch Patch SGP020Artivion, Inc. · Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene2 recalls1 report
- Hemashield M002000195140Intervascular Sas · Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene1 recall2 reports
- Peri-Guard PC0404NSynovis Life Technologies, Inc. · PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE1 recall0 reports
- Supple Peri-Guard PC0814SNSynovis Life Technologies, Inc. · PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE1 recall0 reports
- SJM™ C0914St. Jude Medical, Inc. · PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE1 recall0 reports
- SJM™ C0510St. Jude Medical, Inc. · PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE1 recall0 reports
- CryoPatch® SG Pulmonary Human Cardiac Hemi-Artery Patch SGPH00Artivion, Inc. · Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene1 recall0 reports
- Bard® PTFE Felt Pledgets 007963Bard Peripheral Vascular, Inc. · PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE37 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page