CryoPatch® SG Pulmonary Human Cardiac Branch Patch SGP020Recalls & reported problems
Artivion, Inc. · CryoPatch SG Pulmonary Branch The CryoPatch SG Pulmonary Human Cardiac Patch is a human heart valve aseptically recovered from donated human tissue. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00877234000485), received between Sep 18, 2025 and Sep 18, 2025 — classified as 1 injury.
2 FDA recall records name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K101866); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 28, 2024Class IIOpen, Classified
Grafts were found to have been approved, released, shipped and implanted with an invalid test results.
Z-2532-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Nov 10, 2020Class IITerminated
On November 10, 2020, it was discovered that there are post- processing microbial culture test failures for other allografts, processed using a TRIS saline solution lot that was also used in processing allografts subjected to the recall.
Z-0653-2021 · Root cause: Material/Component Contamination · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Aneurysm1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE INITIAL REPORT RECEIVED VIA EMAIL ON (B)(6) 2025, THE RETURNED IMPLANT SUMMARY CARD IS MARKED TO INDICATE THAT AN ARTIVION GRAFT WAS EXPLANTED ON (B)(6) 2025. A QUERY OF THE IMPLANT DATABSE PROVIDED SGP020 SERIAL NUMBER: (B)(6) THAT WAS IMPLANTED ON (B)(6) 2023. ADDITIONAL INFORMATION FROM THE HOSPITAL RELAYED A RV TO PA CONDUIT REPLACEMENT WAS PERFORMED. AN ANEURYSM BELOW THE CONDUIT WAS THE INDICATION. THE ANEURYSM DID NOT INVOLVE THE CONDUIT. GRAFT WAS NOT SAVED FOR EVALUATION. THE PATIENT IS DOING WELL. NO ADDITIONAL INFORMATION FORTHCOMING. THIS INVESTIGATION IS RELEGATED TO SGP020 SERIAL NUMBER: (B)(6) WITH THE ALLEGATION OF EXPLANT.
Patient problems: Aneurysm
MAUDE report 23085962 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K101866. One submission often covers many models, so check the product and lot details in each record.
- Z-2272-2018 · May 14, 2018 · TerminatedCryopatch SG Pulmonary Human Cardiac Branch, Catalog Number SGP020
- Z-0809-2014 · Nov 22, 2013 · TerminatedPulmonary Hemi-Artery SG Used in heart surgery
- Z-1561-2011 · Feb 17, 2011 · TerminatedPulmonary Hemi-Artery SG Used in heart surgery
Other Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene devices
- Hemashield M002000195140Intervascular Sas · Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene1 recall2 reports
- Peri-Guard PC0404NSynovis Life Technologies, Inc. · PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE1 recall0 reports
- Supple Peri-Guard PC0814SNSynovis Life Technologies, Inc. · PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE1 recall0 reports
- SJM™ C0914St. Jude Medical, Inc. · PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE1 recall0 reports
- SJM™ C0510St. Jude Medical, Inc. · PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE1 recall0 reports
- CryoPatch® SG Pulmonary Human Cardiac Hemi-Artery Patch SGPH00Artivion, Inc. · Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene1 recall0 reports
- Bard® PTFE Felt Pledgets 007963Bard Peripheral Vascular, Inc. · PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE37 reports
- Bard® PTFE Felt Pledgets 007970Bard Peripheral Vascular, Inc. · PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE33 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page