The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00860011670902), received between Jun 3, 2024 and Jul 15, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Application Program Version or Upgrade Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Application Program Version or Upgrade Problem1
  • Difficult to Open or Close1
  • Labelling, Instructions for Use or Training Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 15, 2024Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT IN BETWEEN CASES A THIRD PARTY ORGANIZATION TRAY WAS PLACED ON TOP OF A LOOSE SYRINGE IN DRAWER ONE, WHICH CAUSED THE TRAY TO ACT AS AN ELEVATED OBSTRUCTION CATCHING ON THE INTERIOR AFTER CLOSURE, PREVENTING NORMAL FUNCTION. THE DRAWER OPENED ENOUGH TO CREATE A GAP WHERE A FILE FOLDER WAS USED TO DISLODGE THE SYRINGE AND THE TRAY WAS ALLOWED TO SETTLE BACK DOWN, REMOVING THE OBSTRUCTION AND RESTORING NORMAL FUNCTION OF THE DRAWER. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19742861 ↗
  • MalfunctionJun 3, 2024Labelling, Instructions for Use or Training Problem; Application Program Version or Upgrade Problem; Labelling, Instructions for Use or Training Problem; Application Program Version or Upgrade Problem

    IT WAS REPORTED THAT PRIOR TO A CESAREAN SECTION, THE CABINET DRAWERS WERE ELECTRONICALLY INACCESSIBLE AND AN ATTEMPT TO USE THE MANUAL OVERRIDE FAILED BECAUSE THE USER WAS UNAWARE THAT THE KEY WAS REQUIRED FOR ACCESS. THE DEVICE WAS IMMEDIATELY POWER CYCLED AND WORKED CORRECTLY PRIOR TO THE PROCEDURE. NO INJURY OR ADVERSE IMPACT TO THE PATIENT WAS REPORTED. THE INSTRUCTIONS FOR USE AND TRAINING PROCEDURE WERE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions