Skip to content

SUREcore Plus Biopsy Instrument SP-LA18-25Recalls & reported problems

Uro-1, Inc. · The SUREcore Plus Biopsy Instrument, with a length of 25 cm, facilitates biopsy of soft tissue. It consists of a biopsy needle and an instrument ("gun"). The needle is advanced under imaging to the target tissue and the instrument fires the needle into the tissue to obtain a core of tissue. The tissue core is than sent to pathology for analysis. It is a single-use device. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00860014338663), received between Sep 29, 2026 and Sep 29, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Material Twisted/Bent

    Disposable biopsy core needle was noted to be bent after use on the patient.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927400 ↗

Other Instrument, Biopsy devices

All Instrument, Biopsy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page