Skip to content

Trackmaster TMX428 220VRecalls & reported problems

Full Vision, Inc. · TMX428 220V Trackmaster Treadmill · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. May 12, 2023Open, Classified

    Treadmill malfunction: drive PCB can misread the signals and stop or reverse and then accelerate forward.

    Z-2053-2023 · Root cause: Other · FDA record ↗

  2. Aug 31, 2022Open, Classified

    This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.

    Z-0365-2023 · Root cause: Other · FDA record ↗

Other Treadmill, Powered devices

All Treadmill, Powered devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page