Trackmaster TMX428 110VRecalls & reported problems
Full Vision, Inc. · TMX428 110V Trackmaster Treadmill · Class 1
All models of Trackmaster (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- May 12, 2023Open, Classified
Treadmill malfunction: drive PCB can misread the signals and stop or reverse and then accelerate forward.
Z-2052-2023 · Root cause: Other · Recall details · FDA record ↗
- Aug 31, 2022Open, Classified
This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.
Z-0364-2023 · Root cause: Other · Recall details · FDA record ↗
Other Treadmill, Powered devices
- T2100-ST T2100-ST2Full Vision, Inc. · Treadmill, Powered4 recalls0 reports
- T2100-ST T2100-ST1Full Vision, Inc. · Treadmill, Powered2 recalls0 reports
- Trackmaster TMX428CP 110VFull Vision, Inc. · Treadmill, Powered2 recalls0 reports
- Trackmaster TMX428CP 220VFull Vision, Inc. · Treadmill, Powered2 recalls0 reports
- Trackmaster TMX428 220VFull Vision, Inc. · Treadmill, Powered2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page