vRad PACS 2Recalls & reported problems
Virtual Radiologic Corporation · vRad PACS 2 · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00862234000313), received between Apr 3, 2026 and Apr 3, 2026 — classified as 1 death.
The most frequently reported specific device problem is “Intermittent Program or Algorithm Execution” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 9, 2023Open, Classified
The error resulted in intermittent failure of current (primary) radiology study images not displaying to the interpreting teleradiologist.
Z-2235-2023 · Root cause: Software Manufacturing/Software Deployment · FDA record ↗
Most reported problems
Device problems
- Intermittent Program or Algorithm Execution1
Patient problems
- Intracranial Hemorrhage1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
DeathApr 3, 2026Intermittent Program or Algorithm Execution
ON (B)(6) 2023, A RADIOLOGIST USING VRAD PACS WITH MAMMOGRAPHY EXPERIENCED AN INTERMITTENT SOFTWARE ANOMALY. THE SOFTWARE ANOMALY RESULTED IN PATIENT'S CURRENT RADIOLOGICAL IMAGE TO NOT LAUNCH, INDICATED BY A PERSISTENT "LOADING" MESSAGE DISPLAYED TO THE RADIOLOGIST USER. THE RADIOLOGIST THEN LAUNCHED A PRIOR STUDY (INDICATED BY A "COMPARISON" LABEL ON THE IMAGE), BYPASSING SEVERAL VISUAL CUES AND CONVENTIONS. THE RADIOLOGIST RENDERED AN INTERPRETATION BASED ON HISTORIC IMAGES THAT DID NOT CAPTURE THE PATIENT'S CURRENT, CRITICAL CONDITION, AND THE INTERPRETATION WAS SUBMITTED TO THE LOCAL TREATING FACILITY. ON 20-MAY-2023, VRAD LEARNED THE PATIENT DIED EARLIER IN THE MONTH.
Patient problems: Intracranial Hemorrhage
MAUDE report 24776400 ↗
Other System, Image Processing, Radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports