SecurAcath 5FRecalls & reported problems
Interrad Medical, Inc · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00865382000228), received between Jul 31, 2017 and Jul 16, 2020 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Device Issue” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- No Apparent Adverse Event2
- Device Issue1
- Material Integrity Problem1
- Physical Property Issue1
- Torn Material1
Patient problems
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 16, 2020No Apparent Adverse Event; No Apparent Adverse Event
A TRIPLE LUMEN BARD POWERPICC WAS PLACED ON (B)(6) AND SECURED WITH A 5F SECURACATH IN THE RIGHT BRACHIAL VEIN FOR IV FLUIDS ON A COVID POSITIVE PATIENT. ON (B)(6) THE GREY LUMEN WAS FOUND TO BE LEAKING NEAR OR AT THE SECURACATH. THE CATHETER WAS REPLACED BY OVERWIRE WITHOUT DIFFICULTY. DRESSING CHANGES OCCURRED ON (B)(6) (DUE TO OLD BLOOD) AND ON (B)(6) (DUE TO IT BEING MOIST). FLOW WAS NOTED TO BE SLUGGISH ON BOTH LUMENS FROM (B)(6) THROUGH (B)(6). THE USER FACILITY FDA REPORT NUMBER PROVIDED WAS MW5095271.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10284180 ↗MalfunctionJun 12, 2019No Apparent Adverse Event; No Apparent Adverse Event
(B)(6) MEDICAL CLINICAL EDUCATOR RECEIVED A CALL ABOUT A NURSE WHO WAS PLACING A 5F SECURACATH DEVICE ON AN ARROW 5.5F MIDLINE CATHETER AND WAS UNABLE TO SNAP THE SECURACATH BASE AND COVER TOGETHER. SHE OPENED A NEW ONE THAT WORKED WITHOUT ANY PROBLEMS AND SAVED THAT DEVICE THAT DID NOT SNAP TOGETHER. THE DEVICE WAS RETURNED TO (B)(6) MEDICAL AND WAS ABLE TO BE SNAPPED ON TO A 5.5F ARROW MIDLINE CATHETER WITHOUT ISSUE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8693483 ↗MalfunctionJul 31, 2017Device Issue; Material Integrity Problem; Physical Property Issue; Torn Material; Device Issue; Material Integrity Problem; Physical Property Issue; Torn Material
(B)(6) 2017: (B)(4) DISTRIBUTOR MEDICAL SPECIALTIES (B)(4) WAS NOTIFIED BY A CLINICAL NURSE CONSULTANT OF AN INCIDENT INVOLVING SECURACATH. A SECURACATH HAD BEEN PLACED WITH A PICC ON (B)(6) 2017. ON (B)(6) 2017, IT WAS OBSERVED THAT THE SECURACATH HAD COME OUT WHILE STILL ATTACHED TO THE PICC WHICH HAD MIGRATED 5CM. ON INSPECTION BY THE PICC TEAM, THEY OBSERVED THE NITINOL "FOOT" ON ONE SIDE OF THE SECURACATH HAD SEPARATED FROM THE REST OF THE SECURACATH AND WAS STILL UNDER THE SKIN, THE "FOOT" COULD BE FELT JUST BENEATH THE SKIN, AND WAS CAUSING THE PATIENT NO DISCOMFORT. THE SIZE OF THE "FOOT" IS .194 INCHES (4.9 MM). ON (B)(6) 2017, THE PATIENT RETURNED TO THE HOSPITAL SO THAT A NEW PICC LINE COULD BE INSERTED, WHICH WENT WELL. THE SITE WHERE THE "FOOT" OF THE SECURACATH HAS SEPARATED WAS ASSESSED AND HAD HEALED WITHOUT PROBLEM. IT WAS DECIDED TO NOT ATTEMPT REMOVAL AT THIS TIME AS THE PATIENT WAS STARTING A CHEMOTHERAPY CYCLE THE SAME DAY. THE FOOT OF THE SECURACATH WAS CAUSING THE PATIENT NO DISCOMFORT. THE PICC TEAM WAS OF THE OPINION THAT COMPLETE REMOVAL WOULD NOT BE DIFFICULT SHOULD THE PATIENT WANT TO PROCEED.
Patient problems: Patient Problem/Medical Problem
MAUDE report 6754559 ↗
Other Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page