Micro Plug Set 90002Recalls & reported problems

Ka Medical LLC · Micro Plug Set - 5mm · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00867966000129), received between Jan 19, 2022 and May 8, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Off-Label Use” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Off-Label Use2
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 8, 2026Premature Separation

    THE ACCOUNT ALLEGES THAT WHILE USING A 5 MM MICRO PLUG SET TO CLOSE A PDA IN A 2 KG BABY VIA FEMORAL VENOUS APPROACH, THE DEVICE PREMATURELY DISCONNECTED DURING DEPLOYMENT FROM THE DISTAL END OF THE DELIVERY CATHETER AND EMBOLIZED INTO THE DESCENDING AORTA. AN INTRODUCER SHEATH WAS PLACED IN THE FEMORAL ARTERY AND THE DEVICE WAS RETRIEVED WITH A SNARE. A NEW 5 MM MICRO PLUG SET WAS USED TO CLOSE THE PDA. NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25126299 ↗
  • InjuryJan 19, 2022Off-Label Use; Off-Label Use

    THE ACCOUNT ALLEGES THAT DURING AN OFF LABEL MICRO PLUG IMPLANT PROCEDURE ON A (B)(6) CHILD DIAGNOSED WITH PATIENT DUCTUS ARTERIOSUS (PDA), A 5MM MICRO PLUG WAS FIRST SUCCESSFULLY DEPLOYED, MIGRATED, AND EMBOLIZED WITHIN THE PATIENT'S PULMONARY ARTERY. THE PHYSICIAN THEN SUCCESSFULLY EXTERNALIZED THE DEVICE USING A SNARE. A DIFFERENT MANUFACTURER'S DEVICE WAS THEN SUCCESSFULLY DEPLOYED IN THIS PATIENT. THE PATIENT TOLERATED THE PROCEDURE WELL WITH NO ADDITIONAL CONSEQUENCES TO REPORT. PATIENT WAS DISCHARGED AS EXPECTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 13290623 ↗
  • InjuryJan 19, 2022Off-Label Use; Off-Label Use

    THE ACCOUNT ALLEGES THAT DURING AN OFF LABEL MICRO PLUG IMPLANT PROCEDURE ON A (B)(6) CHILD DIAGNOSED WITH PATIENT DUCTUS ARTERIOSUS (PDA), A 5MM MICRO PLUG WAS FIRST SUCCESSFULLY DEPLOYED, MIGRATED, AND EMBOLIZED WITHIN THE PATIENT'S PULMONARY ARTERY. THE PHYSICIAN THEN SUCCESSFULLY EXTERNALIZED THE DEVICE USING A SNARE. A 6 MM MICRO PLUG DEVICE WAS THEN DEPLOYED AND APPEARED TO BE STABLE. APPROXIMATELY THREE HOURS POST PROCEDURE, THE MICRO PLUG MIGRATED AND EMBOLIZED INTO THE PATIENT'S PULMONARY ARTERY. THE PHYSICIAN ATTEMPTED TO EXTERNALIZE THE DEVICE FROM THE PATIENT USING A SNARE UNSUCCESSFULLY. THE DECISION WAS MADE TO LEAVE THE DEVICE WITHIN THE PATIENT DUE TO THE FACT IT WAS NOT OBSTRUCTIVE TO BLOOD FLOW. THE PATIENT WAS THEN REFERRED TO SURGERY FOR SUCCESSFUL PDA LIGATION. THE PATIENT TOLERATED ALL PROCEDURES VERY WELL WITH NO ADDITIONAL CONSEQUENCES TO REPORT. PATIENT WAS DISCHARGED AS EXPECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13290622 ↗

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →