Skip to content

Micro Plug Set 90003Recalls & reported problems

Ka Medical LLC · Micro Plug Set - 6mm · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00867966000136), received between Jan 19, 2022 and Jan 19, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Off-Label Use” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Off-Label Use1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 19, 2022Off-Label Use; Off-Label Use

    THE ACCOUNT ALLEGES THAT DURING AN OFF LABEL MICRO PLUG IMPLANT PROCEDURE ON A (B)(6) CHILD DIAGNOSED WITH PATIENT DUCTUS ARTERIOSUS (PDA), A 5MM MICRO PLUG WAS FIRST SUCCESSFULLY DEPLOYED, MIGRATED, AND EMBOLIZED WITHIN THE PATIENT'S PULMONARY ARTERY. THE PHYSICIAN THEN SUCCESSFULLY EXTERNALIZED THE DEVICE USING A SNARE. A 6 MM MICRO PLUG DEVICE WAS THEN DEPLOYED AND APPEARED TO BE STABLE. APPROXIMATELY THREE HOURS POST PROCEDURE, THE MICRO PLUG MIGRATED AND EMBOLIZED INTO THE PATIENT'S PULMONARY ARTERY. THE PHYSICIAN ATTEMPTED TO EXTERNALIZE THE DEVICE FROM THE PATIENT USING A SNARE UNSUCCESSFULLY. THE DECISION WAS MADE TO LEAVE THE DEVICE WITHIN THE PATIENT DUE TO THE FACT IT WAS NOT OBSTRUCTIVE TO BLOOD FLOW. THE PATIENT WAS THEN REFERRED TO SURGERY FOR SUCCESSFUL PDA LIGATION. THE PATIENT TOLERATED ALL PROCEDURES VERY WELL WITH NO ADDITIONAL CONSEQUENCES TO REPORT. PATIENT WAS DISCHARGED AS EXPECTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 13290627 ↗

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →

Sources and updates