Sonablate 202-17000-0012Recalls & reported problems
Focus Surgery Inc · Sonablate Probe HF 30/40 3G is an image-guided device that allows physicians to ablate tissue using ultrasound energy. · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00868108000229), received between Sep 26, 2019 and Sep 26, 2019 — classified as 1 injury.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Fistula1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 26, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DR. (B)(6) REPORTED THAT ONE OF HIS PREVIOUS HIFU PATIENTS HAD DEVELOPED A RECTAL FISTULA. PATIENT WAS STATUS POST EXTERNAL BEAM RADIATION THERAPY (EBRT) AND BRACHYTHERAPY. PATIENT HAD A SUPRAPUBIC CATHETER PLACED THE DAY OF THE CASE. AFTER THE SUPRAPUBIC CATHETER WAS DISCONTINUED, PER THE PHYSICIAN, THE PATIENT DEVELOPED DIARRHEA WHICH WAS LATER DIAGNOSED AS A RECTAL FISTULA. DR. (B)(6) PLACED URETHRAL CATHETER WHICH WAS IN PLACE FOR 3 WEEKS AS OF (B)(6) 2019. PER DR. (B)(6) AND VERIFIED VIA A REVIEW OF THE CASE IMAGES, THE PROSTATE AND SVS WERE ABLATED. MANUFACTURER REQUESTED ACCESS TO THE PATIENT'S TREATMENT IMAGES TO REVIEW THE CASE, WHICH DR. (B)(6) WELCOMED. A REVIEW OF THE CASE IMAGES BY MANUFACTURER OCCURRED ON AUGUST 30, 2019 AND RESULTED IN THE FOLLOWING OBSERVATIONS: THE PATIENT HAD A THICKER THAN NORMAL RECTAL WALL, LIKELY FROM PREVIOUS PROSTATE EBRT AND BRACHYTHERAPY. THE PHYSICIAN FOLLOWED POWER RECOMMENDATIONS. THE PHYSICIAN UTILIZED THE 3/6 DUTY CYCLE THROUGHOUT THE CASE. THE EQUIPMENT WORKED AS INTENDED. THE PHYSICIAN DID NOT IGNORE ANY ALARMS OR BLATANTLY MISUSE THE DEVICE.
Patient problems: Fistula
MAUDE report 9124115 ↗
Other High Intensity Ultrasound System For Prostate Tissue Ablation devices
- Sonasource 102-17000-0009Focus Surgery Inc · High Intensity Ultrasound System For Prostate Tissue Ablation2 reports
- TULSA-PRO 109171Profound Medical Inc · High Intensity Ultrasound System For Prostate Tissue Ablation1 report
- TULSA-PRO PAD-105Profound Medical Inc · High Intensity Ultrasound System For Prostate Tissue Ablation1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.