Skip to content

TULSA-PRO 109171Recalls & reported problems

Profound Medical Inc · TULSA-PRO ENDORECTAL COOLING DEVICE · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07540281001370), received between Jun 9, 2023 and Jun 9, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Difficult to Insert1

Patient problems

  • Perforation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 9, 2023Difficult to Insert; Difficult to Insert

    ON (B)(6) 2023, PATIENT UNDERWENT A PROSTATE ABLATION PROCEDURE WITH THE TULSA-PRO TRANSURETHRAL ULTRASOUND ABLATION SYSTEM. THE PROCEDURE INVOLVES INSERTION OF AN ENDORECTAT COOLING DEVICE (ECD) FOR PASSIVE COOLING OF THE RECTUM DURING THE ABLATION PROCEDURE. THE TREATING UROLOGIST REPORTED THAT DURING DEVICE INSERTION, INITIAL IMAGING REVEALED THAT THE ECD WAS NOT PROPERLY INSERTED AND AN ADJUSTMENT WAS REQUIRED TO INSERT THE DEVICE FURTHER INWARDS. THE UROLOGIST REPORTED THAT SHE FELT RESISTANCE DURING THE INSERTION BUT EVENTUALLY WAS CAPABLE TO PUSH THE DEVICE FURTHER INSIDE THE BODY. SUBSEQUENT MR IMAGES, SHOWED THE PROPER PLACEMENT OF THE ECD AND THE TEAM PROCEEDED WITH THE ULTRASOUND TREATMENT OF THE PROSTATE WITH NO UNUSUAL ISSUES RELATED TO RECTAL COOLING. THE FOLLOWING DAY ((B)(6) 2023), PATIENT REPORTED PAIN IN HIS STOMACH. FOLLOW-UP CT IMAGES REVEALED RECTAL PERFORATION BEYOND THE TREATMENT AREA WHERE THE ECD TIP ENDS. THE RADIOLOGIST SUSPECTS THAT THIS PERFORATION HAS HAPPENED DUE TO THE MECHANICAL DAMAGE DONE BY THE EXCESSIVE INSERTION OF THE ECD. RADIOLOGIST MENTIONED THAT THE PATIENT HAD PREVIOUS PROSTATE RADIATION THERAPY AND THEREFORE HIS RECTAL TISSUE COULD BE VU

    Patient problems: Perforation

    MAUDE report 17101694 ↗

FDA clearance and classification

Cleared through 510(k) K211858 on Sep 6, 2022 (decision: Substantially Equivalent). Submitted by Profound Medical, Inc. as “TULSA-PRO System”. The FDA review took 447 days. FDA record ↗

The FDA regulates product code PLP (high intensity ultrasound system for prostate tissue ablation) under 21 CFR 876.4340 as a Class II device. 21 CFR 876.4340 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the TULSA-PRO 109171

Has the TULSA-PRO 109171 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07540281001370), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the TULSA-PRO 109171?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Jun 9, 2023. That report lists device problem “Difficult to Insert” (1); patient problem “Perforation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the TULSA-PRO 109171?

None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the TULSA-PRO 109171 FDA approved?

It was cleared through 510(k) K211858 on Sep 6, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the TULSA-PRO 109171 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.

Who makes the TULSA-PRO 109171, and is it still sold?

The FDA's device database lists Profound Medical Inc as the company responsible for it (brand TULSA-PRO, model 109171). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other High Intensity Ultrasound System For Prostate Tissue Ablation devices

All High Intensity Ultrasound System For Prostate Tissue Ablation devices →