CryoPatch SG Pulmonary Human Cardiac Patch. The patch is… recallZ-0653-2021 · CryoLife, Inc. · initiated Nov 10, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CryoLife, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
On November 10, 2020, it was discovered that there are post- processing microbial culture test failures for other allografts, processed using a TRIS saline solution lot that was also used in processing allografts subjected to the recall.
FDA root cause category: Material/Component Contamination
What the company is doing
A Recall notification letter titled "Urgent - Voluntary Human Tissue Recall" was sent to consignees via email or hand delivered by firm representatives. On 11-Nov-2020, 4 letters were delivered to consignees. On 12-Nov-2020, 2 letters were delivered to consignees. The letter instructs the consignee to complete and return the enclosed postcard affirming that they received the notification. The affected product should be quarantined and returned to CryoLife immediately, if it has not been implanted. CryoLife Field Representatives were to provide information necessary to arrange for the return of the affected product. If the product has been implanted and the consignee has questions or requires clarification, the consignee is to contact Dr. Gregory Ray, CryloLife Medical Director at 800-438-8285.
Recall details
- Recall number
- Z-0653-2021
- Status
- Terminated
- Initiated by the company
- Nov 10, 2020
- Recalling firm
- CryoLife, Inc.
- Quantity
- 2 units
- Product code
- DXZ
- FDA submissions
- K101866
- Distribution
- Distributed nationwide to GA, AZ, TX, MO, and OK.
Product description
CryoPatch SG Pulmonary Human Cardiac Patch. The patch is cryopreserved in a tissue culture medium containing cryoprotectants within the innermost pouch of a three-pouch packaging system. The packaging system not only withstands ultracold temperatures, but also allows for aseptic introduction of the patch into the operating environment. Supercooling by liquid nitrogen boost is begun prior to crystallization to minimize ice crystal damage to the patch matrix. Finally, the patch is transferred for long-term storage at or below 135 C. CryoPatch SG is distributed in three anatomic configurations: pulmonary hemi-artery, pulmonary trunk and pulmonary branch.
Code information (models, lots, serial numbers)
Model/Catalog Number: SGP020 Serial Numbers: 11677186, 11679112, and 11681763 UDI: 00877234000485
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CryoPatch® SG Pulmonary Human Cardiac Branch Patch SGP020Artivion, Inc. · Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene2 recalls1 report
Other devices under the same FDA submission
1 device cleared under K101866 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- CryoPatch® SG Pulmonary Human Cardiac Hemi-Artery Patch SGPH00Artivion, Inc. · Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 86890).
- Z-0652-2021 · TerminatedCryoValve SG Pulmonary Human Heart Valve. The valve is cryopreserved in a tissue culture medium containing cryoprotectants within the innermost pouch of a three-pouch packaging sy…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page