HELIPAQ CERECYTE CHE100930-30Recalls & reported problems
Medos International Sàrl · HELIPAQ microcoil HELIPAQ 10 CERECYTE Coil 9mm x 30cm 190cm Contents: 1 Detachable Coil System · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00878528000600), received between May 11, 2017 and May 11, 2017 — classified as 1 malfunction.
The most frequently reported specific device problem is “Premature Activation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Premature Activation1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 11, 2017Premature Activation; Premature Activation
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING AN ANEURYSM COIL EMBOLIZATION PROCEDURE, A HELIPAQ COIL (CHE10093030/ S10856) WAS NOT SEEN ON THE WIRE FROM THE BEGINNING OF THE PROCESS. IT WAS REPORTED BY THE CUSTOMER THAT THERE WAS NO COIL. THE EVENT WAS IDENTIFIED PRIOR TO USE ON THE PATIENT DURING DEVICE PREPARATION. THE COIL WAS ALSO NOT SEEN IN THE PACKAGING. UPON ANALYSIS OF THE RETURNED PRODUCT, IT WAS FOUND THAT THE MECHANICALLY DETACHED COIL WAS IN THE INTRODUCER SHEATH AND THE DPU AND INTRODUCER SHEATH WERE SEPARATED. ANOTHER PRODUCT OF THE SAME BRAND WAS USED, BUT OF SMALLER LENGTH. THE PRODUCT WAS STORED IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE (IFU). THE PACKAGING DID NOT APPEAR DAMAGED AND THE DEVICE WAS PACKAGED AS EXPECTED. THE EVENT DID NOT RESULT IN A PROCEDURE DELAY AND THE DEVICE WAS NO USED IN THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6560708 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page