MICRUSPHERE CERECYTE CSP100400-30Recalls & reported problems
Medos International Sàrl · MICRUSPHERE microcoil MICRUSPHERE 10 CERECYTE Coil 4mm x 7.5cm 190cm Contents: 1 Detachable Coil System · Class 2 · Implant
All models of MICRUSPHERE CERECYTE (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00878528001041), received between Jun 17, 2017 and Dec 17, 2020 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Difficult to Advance” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Difficult to Advance1
- Failure to Advance1
- Separation Failure1
- Stretched1
Patient problems
- Aneurysm2
- Reaction2
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 17, 2020Stretched; Failure to Advance; Difficult to Advance; Stretched; Failure to Advance; Difficult to Advance
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION, A 4MMX7.5CM MSPHERE 10 CERE COIL WAS BEING USED AS A FILLING COIL. HOWEVER, THE COIL WAS NOT ABLE TO ADVANCE NOR RETRIEVE BY THE PHYSICIAN. THE PHYSICIAN PULLED THE COIL AND UNSPECIFIED MICROCATHETER (MC) AT THE SAME TIME. AFTER INSPECTION, IT TURNS OUT TO BE STRETCHED. THE PHYSICIAN USED A SAME LIKE PRODUCT. THE PROCEDURE WAS SUCCESSFULLY FINISHED. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11027382 ↗MalfunctionJun 7, 2019Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE PROCEDURE WITH THE TARGET SITE ON THE INTERNAL CAROTID ARTERY (ICA) WHILE ATTEMPTING TO DETACH THE 4MM X 7.5CM MICRUSPHERE 10 CERECYTE COIL (CSP10040030 / S10465), THE ENPOWER DETACHMENT CONTROL BOX (DCB00000500 / C27718) DID NOT WORK PROPERLY AND THE COIL DID NOT DETACH. THE COIL WAS REMOVED STILL ATTACHED TO THE DELIVERY SYSTEM AND NOT STRETCHED. A PRE-DEPLOYMENT CHECK WAS PERFORMED AND THERE WAS NO ISSUE NOTED DURING THE CHECK. THERE WAS NO EVIDENCE OF LOW BATTERY, THE BOX HAD BEEN USED WITH PREVIOUS COILS; THE BOX HAD BEEN USED OVER THE LAST SEVERAL MONTHS WITHOUT ISSUE. THE COIL WAS REPLACED, AND THE SAME ISSUE OCCURRED. IT WAS REPORTED THAT THE CUSTOMER CONTINUED TO USE THE DCB WITH THREE MORE COILS BECAUSE THEY HAD ASSUMED THE ISSUE WAS RELATED TO THE COILS AND NOT THE BOX. THE CUSTOMER REPLACED THE DCB AND DETERMINED THAT THE ENPOWER DETACHMENT CONTROL BOX DID NOT WORK PROPERLY AND RESULTED IN THE FAILED DETACHMENT OF THE COILS. THIS COIL AND THE OTHER THREE COILS WERE NOT RE-CHALLENGED AFTER THE DETACHMENT BOX WAS REPLACED. IT WAS REPORTED THAT THE COIL WAS DISCARDED ONCE IT WAS DETERMINED THAT THE ISSUE WAS WITH THE DCB
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8680270 ↗InjuryJun 17, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IN THE LITERATURE ARTICLE ¿ABNORMAL WHITE MATTER CHANGES AFTER CEREBRAL ANEURYSM TREATMENT WITH POLYGLYCOLIC-POLYLACTIC ACID COILS¿ BY LESLI E. SKOLARUS, JOSEPH J. GEMMETE, TIFFANY BRALEY, LEWIS B. MORGENSTERN, ADITYA PANDEY, PUBLISHED WORLD NEUROSURGERY, DOI:10.1016/J.WNEU.2010.03.026, IT WAS REPORTED THAT A FEMALE PATIENT HAD ANEURYSM RECANALIZATION AFTER INITIAL COILING WITH MICRUSPHERE COILS (CSP10070030/F43069, CSP10035030/F40423, CSP10030030/F42941, CSP10050030/F44130, CSP10035030/F44304, CSP10030030/F42941, CSP10025030/F42083, CSP10025030/F43822, CSP10025030/F39589, CSP10035030/F44304, CSP10035030/F42085, CSP10030030/F44030 X 2, CSP10040030/F43624, CSP10040030/F42769), THEN HAD COIL COMPACTION AFTER THE SECOND COIL EMBOLIZATION WITH A MICRUSPHERE COIL (CSP10060030/F44693), AND THEN HAD CEREBRAL LESIONS DEVELOP. PER THE ARTICLE, A (B)(6) FEMALE SMOKER (CASE 2) SUFFERED A HUNT-HESS GRADE 3, FISHER GRADE 4 ANEURYSMAL SAH. CEREBRAL ANGIOGRAM REVEALED A 12 X 8-MM BASILAR TIP ANEURYSM. SHE UNDERWENT EMERGENT ENDOVASCULAR COILING ON (B)(6) 2008 WITH IMPLANTATION OF 15 MICRUSPHERE COILS, WHICH RESULTED IN GREATER THAN 90% ANEURYSM OBLITERATION. MRI SHOWED EVIDENCE OF ANEURYSM RECURR
Patient problems: Aneurysm; Reaction
MAUDE report 6649332 ↗InjuryJun 17, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IN THE LITERATURE ARTICLE ¿ABNORMAL WHITE MATTER CHANGES AFTER CEREBRAL ANEURYSM TREATMENT WITH POLYGLYCOLIC-POLYLACTIC ACID COILS¿ BY LESLI E. SKOLARUS, JOSEPH J. GEMMETE, TIFFANY BRALEY, LEWIS B. MORGENSTERN, ADITYA PANDEY, PUBLISHED WORLD NEUROSURGERY, DOI:10.1016/J.WNEU.2010.03.026, IT WAS REPORTED THAT A FEMALE PATIENT HAD ANEURYSM RECANALIZATION AFTER INITIAL COILING WITH MICRUSPHERE COILS (CSP10070030/F43069, CSP10035030/F40423, CSP10030030/F42941, CSP10050030/F44130, CSP10035030/F44304, CSP10030030/F42941, CSP10025030/F42083, CSP10025030/F43822, CSP10025030/F39589, CSP10035030/F44304, CSP10035030/F42085, CSP10030030/F44030 X 2, CSP10040030/F43624, CSP10040030/F42769), THEN HAD COIL COMPACTION AFTER THE SECOND COIL EMBOLIZATION WITH A MICRUSPHERE COIL (CSP10060030/F44693), AND THEN HAD CEREBRAL LESIONS DEVELOP . PER THE ARTICLE, A (B)(6) FEMALE SMOKER (CASE 2) SUFFERED A HUNT-HESS GRADE 3, FISHER GRADE 4 ANEURYSMAL SAH. CEREBRAL ANGIOGRAM REVEALED A 12 X 8-MM BASILAR TIP ANEURYSM. SHE UNDERWENT EMERGENT ENDOVASCULAR COILING ON (B)(6) 2008 WITH IMPLANTATION OF 15 MICRUSPHERE COILS, WHICH RESULTED IN GREATER THAN 90% ANEURYSM OBLITERATION. MRI SHOWED EVIDENCE OF ANEURYSM R
Patient problems: Aneurysm; Reaction
MAUDE report 6649289 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page