CASHMERE CERECYTE CRC140304-30Recalls & reported problems
Medos International Sàrl · CASHMERE microcoil CASHMERE 14 CERECYTE Coil 3mm x 4cm 190cm Contents: 1 Detachable Coil System · Class 2 · Implant
All models of CASHMERE CERECYTE (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00878528004424), received between Sep 21, 2017 and Nov 9, 2021 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K072173); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Separation Failure2
- Device Difficult to Setup or Prepare1
- Material Deformation1
- Material Integrity Problem1
- Stretched1
Patient problems
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 9, 2021Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION OF THE INFERIOR CEREBELLAR ARTERY (ICA), THE PHYSICIAN DECIDED TO PULL OUT A CASH 14 CERE COIL 3MMX4.0CM (CRC14030430, S14375) BECAUSE IT¿S SIZE WAS NOT APPROPRIATE WITH ANEURYSM. DURING RESHEATHING OF THE COIL, IT WAS FAILED TO BE RE-ZIPPED. THERE WAS NO SURGERY DELAYED DUE TO THE REPORTED EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. ADDITIONAL INFORMATION RECEIVED INDICATED THAT AN ADEQUATE CONTINUOUS FLUSH WAS MAINTAINED THROUGH THE UNSPECIFIED MICROCATHETER (MC). THE DELIVERY WIRE WAS NOT PROTRUDING FROM THE INTRODUCER. THERE WAS NO DAMAGE NOTED TO THE SHEATH INTRODUCER. THERE WAS NO SPLIT. THERE WAS NO RESISTANCE AT ANY TIME DURING ADVANCEMENT OF THE COIL THROUGH THE MC. IT WAS NOT NECESSARY TO REMOVE THE MICROCATHETER. NO ADDITIONAL INTERVENTION WAS REQUIRED. THERE WAS NO BLOOD FLOW RESTRICTION/REDUCTION AS A RESULT OF THE EVENT. BASED ON THE ANALYSIS OF THE DEVICE RECEIVED, THE EMBOLIC COIL IS STRETCHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12778139 ↗MalfunctionJan 11, 2019Separation Failure; Material Integrity Problem; Separation Failure; Material Integrity Problem
A REPORT WAS RECEIVED THAT DURING A COIL EMBOLIZATION, THE 2.5MM X 6CM ULTIPAQ 10 (FSR10025620/C27367) AND 3MM X 4CM CASHMERE 14 (CRC14030430/S10171) COILS COULD NOT BE DEPLOYED. THE COILS WERE REMOVED EASILY WITHOUT THE NEED FOR ADDITIONAL INTERVENTION. THE COILS WERE REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE SURGERY WAS NOT DELAYED DUE TO THE EVENT. NO PATIENT COMPLICATIONS OCCURRED AS A RESULT OF THE EVENT. THE PRODUCTS WILL BE RETURNED FOR EVALUATION. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8241656 ↗MalfunctionAug 20, 2018Separation Failure; Material Deformation; Separation Failure; Material Deformation
AS REPORTED BY A HEALTHCARE PROFESSIONAL, THE PHYSICIAN ATTEMPTED TO DETACH THE 3 MM X 4.0 CM CASHMERE 14 (CRC14030430/UNK LOT) THERMO-MECHANICAL COIL BUT THE COIL FAILED TO DETACH. THE COIL WAS REMOVED FROM THE PATIENT WITHOUT DIFFICULTY AS IT WAS THE FIRST COIL USED IN THE CASE. THERE WAS NO REPORT OF CONSEQUENCE OR IMPACT TO THE PATIENT. IT WAS REPORTED THAT THE INFORMATION PROVIDED FOR THIS COMPLAINT WAS ALL OF THE DETAILS THAT WERE KNOWN/AVAILABLE REGARDING THIS EVENT. THE PRODUCT WILL BE RETURNED FOR EVALUATION. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7797304 ↗MalfunctionSep 21, 2017Stretched; Stretched
IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL THAT A CASHMERE COIL ( CRC14030430/C27378) STRETCHED DURING AN ANEURYSM EMBOLIZATION PROCEDURE. ANOTHER COIL WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6883873 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K072173. One submission often covers many models, so check the product and lot details in each record.
- Z-0845-2011 · Oct 21, 2010 · TerminatedMicrus Endovascular Microcoil Delivery System; The Micrus Microcoil Delivery System consists of three components, a Microcoil System, a Connecting Cable, and a Detachment Control B
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page