CASHMERE CERECYTE CRC140820-30Recalls & reported problems
Medos International Sàrl · CASHMERE microcoil CASHMERE 14 CERECYTE Coil 8mm x 20cm 190cm Contents: 1 Detachable Coil System · Class 2 · Implant
All models of CASHMERE CERECYTE (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00878528004646), received between Oct 12, 2016 and Jul 23, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Positioning Failure” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Positioning Failure1
- Separation Failure1
- Stretched1
Patient problems
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 23, 2018Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE PROCEDURE, THE CASHMERE 14 CERECYTE COIL 8 MM X 20 CM (CRC14082030/ C27073) COULD NOT BE DETACHED. THE COIL WAS WITHDRAWN AND ANOTHER COIL WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORT OF ANY PATIENT INJURY OR CONSEQUENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7712771 ↗MalfunctionOct 13, 2017Positioning Failure; Positioning Failure
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING THE PROCEDURE, THE CASHMERE COIL (CRC14082030/C27073) COULD NOT BE DETACHED. THEY USED ANOTHER ONE TO COMPLETE THE PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY. IT WAS REPORTED THAT THE DEVICE WOULD BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6947128 ↗MalfunctionOct 12, 2016Stretched; Stretched
AS REPORTED BY A HEALTH CARE PROFESSIONAL, DURING A COIL EMBOLIZATION PROCEDURE A CASHMERE 14 CERE 8 MM X 20 CM (CRC14082030/C24696) STRETCHED. THE ENTIRE COIL WAS WITHDRAWN FROM THE MICROCATHETER AND NEW COIL WAS USED TO COMPLETE THE PROCEDURE; THE MICROCATHETER WAS NOT REMOVED. THERE WERE NO KINKS NOTED ON THE MICROCATHETER AND THERE WAS NO RESISTANCE BETWEEN THE MICROCATHETER AND GUIDEWIRE WHEN ACCESSING THE TARGET SITE. AN ADEQUATE CONTINUOUS FLUSH WAS MAINTAINED THROUGH THE MICROCATHETER AT ALL TIMES. THERE WAS NO REPORT ON PATIENT INJURY AND NO REPORTS ON INTRA OR POST-PROCEDURAL COMPLICATIONS. PATIENT INFORMATION IS UNKNOWN. IT WAS INITIALLY REPORTED THAT THE COMPLAINT DEVICE IS AVAILABLE FOR RETURN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6021785 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page