DELTAPAQ DFS100210-20Recalls & reported problems
Medos International Sàrl · DELTAPAQ microcoil DELTAPAQ 10 Stretch Resistant Coil 2mm x 10cm 190cm Contents: 1 Detachable Coil System · Class 2 · Implant
All models of DELTAPAQ (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00878528007616), received between Jan 9, 2017 and Oct 7, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Deformation Due to Compressive Stress1
- Positioning Failure1
- Separation Failure1
- Stretched1
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 7, 2021Separation Failure; Separation Failure
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION, A DELTAPAQ 2MMX10CM SR PLAT COIL (DFS10021020, UNKNOWN LOT) WOULD NOT DETACH WHEN IT ARRIVED AT THE SIGHT FOR DEPLOYMENT. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12590165 ↗MalfunctionDec 28, 2017Positioning Failure; Stretched; Positioning Failure; Stretched
AS REPORTED BY A PHYSICIAN, DURING COIL EMBOLIZATION OF THE RIGHT INTERNAL CAROTID ARTERY, AN ENPOWER DETACHMENT CONTROL BOX (DCB00000500/F42808) WAS UNABLE TO DETACH A DELTAPAQ COIL (DFS10021020/ S14661). UPON ATTEMPTING TO DETACH THE COIL, A RED LIGHT WAS VISIBLE ON THE BOX. THEY TURNED THE BOX OFF AND ON AND THE DCB SEEMED TO REINITIALIZE, BUT THE ACCOUNT USED ANOTHER DCB FOR THE REST OF THE COILING PROCEDURE; HOWEVER, THE DELTAPAQ CONTINUED TO NOT BE ABLE TO DETACH WITH THE NEW BOX. THE COIL WAS REMOVED AND SET ASIDE TO BE RETURNED FOR EVALUATION. THE PHYSICIAN CONTINUED TO SUCCESSFULLY COIL THE RIGHT ICA ANEURYSM WITH NO FURTHER ISSUES USING THE REPLACEMENT DCB. BOTH COMPLAINT PRODUCTS HAD BEEN PREPPED AND USED AS PER THE IFU AND ALL CONNECTIONS FIT PROPERLY WITHOUT NEED FOR EXCESSIVE FORCE. NEITHER DEVICE APPEARED DAMAGED. A PRE-DEPLOYMENT ELECTRICAL CHECK HAD BEEN PERFORMED. IT WAS REPORTED THAT THE PRODUCTS WOULD BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7153697 ↗MalfunctionJan 9, 2017Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
DURING COIL EMBOLIZATION, A DELTAPAQ ((B)(4)) WAS UNABLE TO FIT DOWN THE UNSPECIFIED MICROCATHETER, AND DURING PRODUCT ANALYSIS, IT WAS FOUND THAT THE THE PROXIMAL END OF THE COIL HAD COMPRESSION AND BUCKLING, WHICH WAS UNREPORTED DAMAGE. THEY OPENED THE SAME MODEL OF COIL AND IT WENT DOWN THE CATHETER WITH NO ISSUES. THERE WAS NO PATIENT INJURY OR PROCEDURE DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6232283 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page