DELTAPAQ DFS100925-20Recalls & reported problems
Medos International Sàrl · DELTAPAQ microcoil DELTAPAQ 10 Stretch Resistant Coil 9mm x 25cm 190cm Contents: 1 Detachable Coil System · Class 2 · Implant
All models of DELTAPAQ (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00878528007968), received between Nov 16, 2018 and Mar 25, 2019 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Deformation1
- Material Twisted/Bent1
- Stretched1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 25, 2019Stretched; Material Twisted/Bent; Stretched; Material Twisted/Bent
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE COIL EMBOLIZATION OF A SHORT NECK SPLENIC ARTERY ANEURYSM, THE 9MM X 25CM DELTAPAQ 10 STRETCH RESISTANT COIL (DFS10092520 / S10053) WAS CHOSEN AFTER PREVIOUS COILS WERE ADVANCED AND PLACED THROUGH THE PROWLER SELECT PLUS MICROCATHETER. THE DELTAPAQ 10 STRETCH RESISTANCE COIL BECAME IMPEDED IN THE DISTAL THIRD OF THE MICROCATHETER AND COULD NOT BE ADVANCED TO THE TARGET LESION. THE COIL WAS REMOVED TO BE RETURNED FOR EVALUATION AND ANALYSIS. THERE WAS NO REPORT OF ANY PATIENT CONSEQUENCE OR ADVERSE EVENT ASSOCIATED WITH THE EVENT. THE 9MM X 25CM DELTAPAQ 10 STRETCH RESISTANT COIL WAS RETURNED FOR EVALUATION WHICH REVEALED THAT THE EMBOLIC COIL WAS KINKED AND STRETCHED. BASED ON THE PRODUCT ANALYSIS COMPLETED ON 3/19/2019, THIS EVENT HAS BEEN DEEMED MDR REPORTABLE AS A ¿MALFUNCTION.¿
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8450154 ↗MalfunctionNov 16, 2018Material Deformation; Material Deformation
AS REPORTED BY THE (B)(6), DURING USE OF A 9MM X 25CM DELTAPAQ ((B)(4) THERMO-MECHANICAL COIL, COMPRESSION OF THE ¿SPRING¿ WAS NOTED, MAKING IT IMPOSSIBLE TO ¿PROGRESS¿. THERE WAS NO REPORT OF CONSEQUENCE OR IMPACT TO THE PATIENT. IT WAS REPORTED THAT THE INFORMATION PROVIDED FOR THIS COMPLAINT WAS ALL OF THE DETAILS THAT WERE KNOWN/AVAILABLE REGARDING THIS EVENT. THE PRODUCT IS NOT AVAILABLE FOR EVALUATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8080680 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page