DELTAPLUSH DPL100152-20Recalls & reported problems
Medos International Sàrl · DELTAPLUSH microcoil DELTAPLUSH 10 Stretch Resistant Coil 1.5mm x 2cm 190cm Contents: 1 Detachable Coil System · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00878528008019), received between Aug 24, 2016 and Jul 14, 2017 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Inadequacy of Device Shape and/or Size1
- Positioning Failure1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 14, 2017Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL THAT THE DETACH TIP SIDE OF THE DELTAPLUSH COIL ((B)(4)) IS LONGER THAN OTHER COILS AND WILL PUSH OUT OF THE MICROCATHETER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6714981 ↗MalfunctionAug 24, 2016Positioning Failure; Positioning Failure
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING LEFT ANTERIOR COMMUNICATING ARTERY COIL EMBOLIZATION, UPON ATTEMPTING TO DEPLOY A DELTAPLUSH COIL ((B)(4)), THE PHYSICIAN WAS UNABLE TO DETACH THE COIL. THE PHYSICIAN HAD ALREADY SUCCESSFULLY IMPLANTED AND DETACHED SEVERAL COILS USING THE SAME ENPOWER CABLE AND DCB. SEVERAL ATTEMPTS WERE MADE TO DETACH THE DELTAPLUSH WITH NO SUCCESS. THE PHYSICIAN THEN REMOVED THE COIL IN QUESTION WITHOUT DIFFICULTY, AND THERE WAS NO DAMAGE NOTED TO THE COIL (I.E. KINKED, STRETCHED, SEPARATED, DETACHED). A DIFFERENT DELTAPLUSH OF THE SAME SIZE WAS IMPLANTED AND SUCCESSFULLY DETACHED USING A REPLACEMENT CABLE AND THE SAME DCB WITH NO ISSUE. IT WAS CONFIRMED THAT HERE WAS NO ISSUE WITH THE INITIAL CABLE AND DCB. WHEN ATTEMPTING TO DETACH THE COIL, THE LIGHTS DID NOT ILLUMINATE AS EXPECTED. A PRE-DEPLOYMENT ELECTRICAL CHECK HAD NOT BEEN PERFORMED. ALL CONNECTIONS FIT PROPERLY WITHOUT NEED FOR EXCESSIVE FORCE. THERE WAS NO RESISTANCE BETWEEN THE COIL AND MICROCATHETER, AND A CONTINUOUS FLUSH HAD BEEN MAINTAINED THROUGH THE MICROCATHETER. NO FURTHER DETAILS WERE PROVIDED REGARDING THE COMPLAINT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5901050 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page