DELTAPLUSH DPL100203-20Recalls & reported problems
Medos International Sàrl · DELTAPLUSH microcoil DELTAPLUSH 10 Stretch Resistant Coil 2mm x 3cm 190cm Contents: 1 Detachable Coil System · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00878528008057), received between Oct 19, 2016 and Sep 24, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Bent” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Bent1
- Defective Component1
- Deformation Due to Compressive Stress1
- Material Twisted/Bent1
Patient problems
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2018Material Twisted/Bent; Material Twisted/Bent
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE COIL EMBOLIZATION PROCEDURE WITH THE TARGET VESSEL AT THE INTERNAL CAROTID ARTERY (ICA) ON A (B)(6) PATIENT WITH A HISTORY OF HEAD TRAUMA AND WHOSE VESSEL WAS REPORTED TO BE VERY TORTUOUS, THE DELTAPLUSH 10 2MM X 3CM COIL (DPL10020320 / S14053) WAS REPORTED TO HAVE ADVANCING ISSUE. THE COIL COULD NOT ADVANCE THROUGH THE MICROCATHETER (STRYKER, SL-10). THERE WAS NO DAMAGE TO THE COIL OR COIL DETACHMENT SYSTEM PRIOR TO THE PROCEDURE. IT WAS REPORTED THAT THE COIL COULD NOT BE PUSHED INTO THE MICROCATHETER WITH EASE, THAT IT SEEMS STUCK. THE DEVICE WAS PREPPED IN ACCORDANCE WITH THE INSTRUCTION FOR USE (IFU), ADEQUATE CONTINUOUS FLUSH WAS MAINTAINED THROUGH THE CATHETER AT ALL TIMES. THERE WAS NO REPORT OF ANY CLINICAL DELAY TO THE PROCEDURE DUE TO THE REPORTED ISSUE. THERE WAS NO REPORT OF PATIENT CONSEQUENCE OR ADVERSE EVENT RELATED TO THE REPORTED ISSUE. THE DEVICE WAS RETURNED FOR EVALUATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7904207 ↗MalfunctionApr 13, 2017Defective Component; Defective Component
IT WAS REPORTED THAT DURING A COIL EMBOLIZATION PROCEDURE, WHEN THE COIL WAS PULLED OUT OF THE PACKAGE BY THE FELLOW, THE SURGEON THOUGHT THE DELTAPLUSH COIL (DPL10020320/ P11228) WAS LONGER THAN THE SIZE IDENTIFIED ON THE BOX. THE PROCEDURE WAS DELAYED 10 MINUTES DO TO THE EVENT, BUT THERE WERE NO PATIENT ADVERSE EVENTS. THE PROCEDURE WAS COMPLETED WITH A SAME/LIKE PRODUCT. THE DEVICE HAD BEEN STORED ACCORDING TO THE LABELING INSTRUCTIONS. IT WAS REPORTED THAT THE DEVICE WOULD BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6490961 ↗MalfunctionOct 19, 2016Bent; Deformation Due to Compressive Stress; Bent; Deformation Due to Compressive Stress
AS REPORTED BY A HEALTHCARE PROFESSIONAL, A DELTAPLUSH (DPL10020320/C26897) SHEATH TWISTED DURING ANEURYSM COILING. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT. THE PRODUCT WAS RETURNED FOR ANALYSIS AND THE SHEATH DID NOT APPEAR TWISTED; HOWEVER, THE COIL APPEARED COMPRESSED, BUCKLED AND BENT. MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION WERE UNSUCCESSFUL.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6039674 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page