MICRUSPHERE SSR100714-20Recalls & reported problems
Medos International Sàrl · MICRUSPHERE XL microcoil MICRUSPHERE XL 10 Stretch Resistant Coil 7mm x 14cm 190cm Contents: 1 Detachable Coil System · Class 2 · Implant
All models of MICRUSPHERE (22)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00878528009825), received between Aug 14, 2018 and Apr 26, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Stretched” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Stretched2
- Fracture1
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 26, 2024Fracture; Fracture
AS REPORTED BY THE FIELD, THIS WAS A COIL EMBOLIZATION TO A LEFT ANTERIOR COMMUNICATING ARTERY ANEURYSM (ACOMM) WITH HIGH TORTUOSITY. THE PHYSICIAN ATTEMPTED TO COIL ANEURYSM INITIALLY WITH ONLY COILS. FIRST COIL WAS A UNIFORM 7MM X 30 CM (FCX100730, 31127140). THIS SPECIFIC ANEURYSM HAD A LEDGE TOWARDS THE NECK OF THE ANEURYSM THAT CAUSED A HEADWAY DUO 156CM MICROCATHETER (MC) TO BECOME UNSTABLE. BECAUSE OF THE LEDGE, BECAUSE OF THE FORCE NEEDED TO ADVANCE THE COIL INTO THE ANEURYSM, THE COIL FRACTURED AT THE JUNCTION OF THE COIL AND THE PUSHER WIRE. THE 5 FR SOFIA WAS ADVANCED AND THE COIL WAS TAKEN OUT WITH THE MICROCATHETER AS A UNIT. THE ENTIRE COIL WAS REMOVED. NEW MICROCATHETER WAS USED TO RE-ACCESS THE ANEURYSM. THE SECOND COIL, A UNIFORM 7MM X 13.9 CM (FCX100713, 31098464) WAS ADVANCED INTO THE ANEURYSM WITHOUT ANY ISSUE. ATTEMPTED TO DETACH COIL WITH MECHANICAL DETACHMENT HANDLE (MDH1, 102109430) UNSUCCESSFULLY. THE MANUAL BREAK WAS USED CORRECTLY AND THAT WAS UNSUCCESSFUL AS WELL. THAT COIL WAS REMOVED WITHOUT ANY ISSUES. A MICRUSPHERE XL 7MMX14CM COIL (SSR10071420, 30529251) WAS USED AT THIS POINT WITH THE SAME MICROCATHETER IN PLACE. THAT COIL FRACTURED AS WELL DUE TO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19197119 ↗MalfunctionMay 24, 2019Stretched; Stretched
THE HEALTHCARE PROFESSIONAL REPORTED THAT THE 7MM X 14CM MICRUSPHERE XL 10 COIL (SSR10071420 / P11270) COULD NOT BE ADVANCED INTO THE MICROCATHETER. THERE WAS NO REPORT OF ANY PATIENT ADVERSE EVENT OR COMPLICATION. MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION RELATED TO THE PROCEDURE AND THE REPORTED DEVICE ISSUE WERE UNSUCCESSFUL. IF ADDITIONAL INFORMATION IS RECEIVED AT A LATER DATE, THIS FILE WILL BE UPDATED ACCORDINGLY. THE PRODUCT WAS RETURNED FOR EVALUATION WHICH REVEALED THAT THE 7MM X 14CM MICRUSPHERE XL 10 COIL WAS IN A STRETCHED CONDITION. BASED ON THE PRODUCT ANALYSIS COMPLETED ON 5/24/2019, THIS EVENT HAS BEEN DEEMED MDR REPORTABLE AS A ¿MALFUNCTION.¿
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8642749 ↗MalfunctionAug 14, 2018Stretched; Stretched
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A COIL EMBOLIZATION OF A CAROTID-CAVERNOUS FISTULA (CCF), THE 7 MM X 14 CM MICRUSPHERE XL (SSR10071420/S10032) THERMO-MECHANICAL COIL DELIVERY SYSTEM WAS ADVANCED THROUGH THE ECHELON 10 (MEDTRONIC) MICROCATHETER BUT STRONG RESISTANCE WAS ENCOUNTERED AT THE MIDSHAFT OF THE CATHETER. THE COIL DELIVERY SYSTEM BECAME STUCK AND COULD NO LONGER BE ADVANCED. THE COIL WAS REMOVED FROM THE MICROCATHETER. WHEN THE COIL WAS REMOVED, THE EMBOLIC COIL WAS STILL ATTACHED TO THE DELIVERY SYSTEM, BUT THE USER NOTED THAT THE EMBOLIC COIL WAS STRETCHED. THE SAME MICROCATHETER WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS OCCURRED AS A RESULT OF THE EVENT. THE SURGERY WAS NOT DELAYED DUE TO THE EVENT. PRIOR TO THIS COIL, TWO COILS WERE ADVANCED THROUGH THE SAME MICROCATHETER WITHOUT RESISTANCE. THE COIL DELIVERY SYSTEM WAS PREPPED WITHOUT ANY DIFFICULTY. THE USER DID NOT FASTEN THE ROTATING HEMOSTATIC VALVE (RHV) TOO TIGHTLY AROUND THE INTRODUCER. THE USER CONTINUOUSLY VERIFIED THAT THE INTRODUCER TIP AND MICROCATHETER HUB REMAINED ALIGNED. AN ADEQUATE CONTINUOUS FLUSH WAS MAINTAINED THROUGH THE MICROCATHETER. WHEN THE UNUSUAL FRICTION O
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7782105 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page