MICRUSPHERE SSR180927-20Recalls & reported problems
Medos International Sàrl · MICRUSPHERE XL microcoil MICRUSPHERE XL 18 Stretch Resistant Coil 9mm x 27cm 190cm Contents: 1 Detachable Coil System · Class 2 · Implant
All models of MICRUSPHERE (22)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00878528009917), received between Feb 21, 2017 and Mar 14, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Stretched” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Stretched3
- Premature Activation1
- Premature Separation1
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 14, 2023Stretched; Premature Separation; Stretched; Premature Separation
AS REPORTED BY THE FIELD, DURING A DISTAL POSTERIOR CEREBRAL ARTERY ANEURYSM (PCA) MEASURING AT APPROXIMATELY 20 MM IN SIZE, A MICROSPHERE XL 9MMX27CM COIL (SSR18092720, L14899) WAS ADVANCED THROUGH A HEADWAY DUO 156 CM MICROCATHETER (MC) AND STRETCHED TO THE POINT THAT THE COIL DETACHED INSIDE THE MICROCATHETER. THE PHYSICIAN ATTEMPTED TO USE A COIL PUSHER WIRE TO ADVANCE THE COIL THROUGH THE MICROCATHETER INTO THE ANEURYSM UNSUCCESSFULLY. ULTIMATELY, THE DECISION WAS MADE TO TAKE OUT THE MICROCATHETER. FORTUNATELY, THE ENTIRE COIL WAS REMOVED FROM THE BODY. A STRYKER XT17 WAS USED TO REASSESS THE ANEURYSM AND FINISHED COILING WITH 2 TARGET COILS. THIS IS CONSIDERED A VERY COMPLICATED CASE DUE TO THE LOCATION OF THE ANEURYSM. EITHER PRODUCT WAS NOT ALTERED AFTER REMOVAL, FROM THE PATIENT¿S BODY. THIS DID RESULT IN THE DELAY OF PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16536907 ↗MalfunctionJul 17, 2017Premature Activation; Stretched; Premature Activation; Stretched
AS REPORTED BY A HEALTHCARE PROFESSIONAL, A MICRUSPHERE COIL (SSR18092720/P11729) WAS ADVANCED THROUGH AN ECHELON MICROCATHETER HOWEVER THE COIL WAS STUCK INSIDE THE MICROCATHETER AT THE DISTAL END. THE COIL WAS REMOVED ALONG WITH THE MICROCATHETER. PATIENT WAS NOT EFFECTED BUT THE SURGERY WAS DELAYED BY 10 MINUTES AS THEY WERE REQUIRED TO OPEN ANOTHER COIL AND MICROCATHETER. THERE WERE NO DAMAGES NOTED ON THE COMPLAINT COIL OR THE MICROCATHETER PRIOR TO USE OR UPON REMOVAL. THE COMPLAINT COIL AND THE MICROCATHETER ARE AVAILABLE FOR INVESTIGATION. UPON RECEIVED THE DEVICE FOR INVESTIGATION, ADDITIONAL UNREPORTED DAMAGE WAS FOUND. THE CORE WIRE WAS KINKED, THE EMBOLIC COIL WAS MECHANICALLY DETACHED AND THE PROXIMAL END OF THE COIL WAS STRETCHED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6716828 ↗MalfunctionFeb 21, 2017Stretched; Stretched
IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL THAT DURING AN ANEURYSM COILING PROCEDURE OF AN ANTERIOR COMMUNICATING ARTERY ANEURYSM, A MICROSPHERE COIL (SSR18092720/C10005) BECAME STUCK IN THE MID PORTION OF A COMPETITOR'S MICROCATHETER DURING INSERTION AND WAS STRETCHED WHEN IT WAS REMOVED. THE COIL WAS REMOVED FROM THE CATHETER AND WAS STILL ATTACHED TO THE DELIVERY SYSTEM. THIS WAS THE FIRST COIL ATTEMPTED TO BE DEPLOYED THROUGH THE MICROCATHETER. THERE WERE NO KINKS FOUND ON THE MICROCATHETER. THERE WERE NO DAMAGES NOTED TO THE COIL PRIOR TO USE. IT WAS REPORTED THAT THE COMPLAINT PRODUCT WILL BE RETURNED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6345774 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.