Skip to content

AMS 800 Artificial Urinary Sphincter 72404144Recalls & reported problems

Boston Scientific Corporation · Occlusive Cuff with InhibiZone · Class 3 · Implant

All models of AMS 800 Artificial Urinary Sphincter (31)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00878953003191), received between May 25, 2022 and Sep 21, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

6 other recalls cover devices cleared under the same FDA submission (P000053); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1

Patient problems

  • Erosion1
  • Unspecified Infection1
  • Urinary Incontinence1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 21, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this patient with an Artificial Urinary Sphincter (AUS) experienced pelvic trauma. During a clinic visit a cystoscopy procedure was performed during which it was discovered that there was erosion at the bulbous urethra, and a surgical procedure was performed in (b)(6) 2026, during which the cuff was explanted. An infection was confirmed later on which resulted in another surgical procedure in (b)(6) 2026, during which the pump and Pressure Regulating Balloon (PRB) were explanted. The physician was unable to identify the type of infection the patient had. The infection was treated with antibiotics. There were no device issues, the patient fully recovered, and there was no additional information was provided.

    Patient problems: Erosion; Unspecified Infection

    MAUDE report 50843404 ↗
  • InjuryMay 25, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A SURGICAL PROCEDURE TO HAVE AN ARTIFICIAL URINARY SPHINCTER (AUS) CUFF EXPLANTED AND A NEW SMALLER AUS CUFF IMPLANTED DUE TO RECURRING INCONTINENCE. THE PHYSICIAN BELIEVED THAT A SMALLER CUFF SIZE WOULD RESOLVE THE ISSUE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Urinary Incontinence

    MAUDE report 14502217 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000053. One submission often covers many models, so check the product and lot details in each record.

  • Z-0531-2025 · Oct 28, 2024 · Open, Classified
    AMS Artificial Urinary Sphincter Pressure Regulating Balloon 61-70 cm H2O
  • Z-0530-2025 · Oct 28, 2024 · Open, Classified
    AMS Artificial Urinary Sphincter Control Pump
  • Z-2442-2019 · Jul 19, 2019 · Terminated
    AMS 800 Artificial Urinary Sphincter AMS 800 Belt Cuff, 4.5 cm Part Number: 72400161
  • Z-1280-2013 · Mar 1, 2013 · Terminated
    AMS 800 Urinary Control System. The AMS 800 Urinary Control System is used to treat urinary incontinence due to reduced outlet resistance (intrinsic sphincter deficiency) followi
  • Z-1279-2013 · Mar 1, 2013 · Terminated
    AMS 800 Urinary Control System. The AMS 800 Urinary Control System is used to treat urinary incontinence due to reduced outlet resistance (intrinsic sphincter deficiency) follow
  • Z-2669-2011 · May 9, 2011 · Terminated
    AMS 800 Urinary Control System Control Pump, REF 72400098, (720172-01 Japan only ), Sterilization Method Steam, American Medical systems Inc., 10700 Bren Road West, Minnetonka, M

Other Device, Incontinence, Mechanical/Hydraulic devices

All Device, Incontinence, Mechanical/Hydraulic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page