AMS 700 LGX MS Pump 72404250Recalls & reported problems
Boston Scientific Corporation · Inflatable Penile Prosthesis, Preconnect Penoscrotal, 1 Pump to 2 Cylinders, InhibiZone treated · Class 3 · Implant
All models of AMS 700 LGX MS Pump (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific AMS 700 LGX inflatable penile prosthesis is a surgically implanted pump-and-cylinder implant. The 48 reports mostly describe parts wearing through or breaking down — cylinders that bulge or leak fluid, a pump or reservoir that shifts or herniates, and erosion of tissue at the pump site — often leading to pain, discomfort, infection, or surgery to remove and replace the device. 46 reports describe injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 49
- Deaths reported
- 0
- Injuries reported
- 47
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 00878953003580), received between Nov 28, 2018 and Sep 28, 2026 — classified as 47 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (17 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Problem17
- Adverse Event Without Identified Device or Use Problem12
- Inflation Problem11
- Fluid/Blood Leak9
- Material Puncture/Hole6
- Inadequacy of Device Shape and/or Size5
- Insufficient Device Problem Information5
- Defective Device4
Patient problems
- No Clinical Signs, Symptoms or Conditions24
- Discomfort6
- Erosion6
- Unspecified Infection6
- Pain5
- No Consequences Or Impact To Patient4
- No Known Impact Or Consequence To Patient2
- Blood Loss1
Recent adverse event reports
The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.
InjurySep 28, 2026Fluid/Blood Leak; Inflation Problem; Material Puncture/Hole; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50926217 ↗InjurySep 23, 2026Mechanical Problem
It was reported that a patient with an Inflatable Penile Prosthesis (IPP) had a device no longer functioning. A revision procedure was done to explant and replace the device. No further patient complications were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50867421 ↗InjuryAug 24, 2026Insufficient Device Problem Information
It was reported that a patient with an Inflatable Penile Prosthesis (IPP) experienced scrotal erosion at the pump site, as well as erosion of the left cylinder. A surgical procedure was performed to remove the IPP and replace it with a Tactra. No further patient complications were reported.
Patient problems: Erosion
MAUDE report 50415924 ↗InjuryAug 5, 2026Migration
It was reported that the pump of this Inflatable Penile Prosthesis (IPP) crossed over and the patient experienced discomfort. As a result, the cylinder and pump were explanted and replaced. The reservoir was left in the patient. No further patient complications were reported.
Patient problems: Discomfort
MAUDE report 50150279 ↗InjuryApr 6, 2026Degraded
IT WAS REPORTED THAT THE CYLINDER OF THIS INFLATABLE PENILE PROSTHESIS (IPP) PRESENTED A BULGE. THE CYLINDER WAS EXPLANTED AND REPLACED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24792325 ↗InjuryJan 28, 2026Degraded; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT WITH AN INFLATABLE PENILE PROSTHESIS (IPP) EXPERIENCED CYLINDER EROSION, DISCOMFORT AND THE DEVICE WAS NOT WORKING AS EXPECTED AS THE CYLINDER HAD AN ANEURYSM. A SURGICAL PROCEDURE WAS PERFORMED TO REMOVE AND REPLACE THE IPP. NO ADDITIONAL PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Discomfort; Erosion
MAUDE report 24201525 ↗InjuryJul 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS INFLATABLE PENILE PROSTHESIS (IPP) WAS EXPERIENCING A CHRONIC INFECTION. A SURGICAL PROCEDURE WAS PERFORMED IN WHICH THE DEVICE WAS REMOVED, A WASHOUT WAS PERFORMED, AND A MALLEABLE IMPLANT WAS PLACED. THERE WERE NO FURTHER PATIENT COMPLICATIONS REPORTED.
Patient problems: Unspecified Infection
MAUDE report 22571590 ↗InjuryJun 5, 2025Inadequacy of Device Shape and/or Size; Migration; Inadequacy of Device Shape and/or Size; Migration
IT WAS REPORTED THAT THE RESERVOIR OF THIS INFLATABLE PENILE PROSTHESIS (IPP) HERNIATED, AND THE CYLINDERS WERE TOO SHORT. A SURGICAL PROCEDURE WAS PERFORMED TO REMOVE AND REPLACE THE IPP. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22149719 ↗InjuryMay 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT WITH THIS INFLATABLE PENILE PROSTHESIS (IPP) WAS NOT SATISFIED WITH THE PUMP OF THE DEVICE. IT WAS NOTED THAT THE PATIENT EXPERIENCED DISCOMFORT AND PAIN EVEN AFTER MULTIPLE ADJUSTMENTS WERE ATTEMPTED. A SURGICAL PROCEDURE WAS PERFORMED IN WHICH THE EXISTING DEVICE WAS EXPLANTED. THERE WERE NO FURTHER PATIENT COMPLICATIONS REPORTED.
Patient problems: Discomfort; Pain
MAUDE report 22077969 ↗InjuryFeb 13, 2025Inflation Problem; Mechanical Problem; Inflation Problem; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS INFLATABLE PENILE PROSTHESIS WAS NOT ABLE TO FILL THE PROSTHESIS COMPLETELY FOR 6 MONTHS, THE CYLINDERS WERE NOT COMPLETELY FILLED, AND HE WAS NOT ABLE TO ACHIEVE FULL ERECTION. THE DEVICE IS STILL IMPLANTED. NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21375469 ↗InjuryFeb 5, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT THE PATIENT WITH THIS INFLATABLE PENILE PROSTHESIS (IPP) WAS DISSATISFIED WITH THE DEVICE DUE TO IMPROPER SIZING AND MEASUREMENT BY THE PREVIOUS DOCTOR. A SURGICAL PROCEDURE WAS PERFORMED TO REMOVE THE EXISTING IPP AND REPLACE IT WITH ONE OF APPROPRIATE SIZE, KEEPING THE ORIGINAL RESERVOIR AND PLACING 18 CM CYLINDERS WITH 3 CM RTE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21312692 ↗InjuryNov 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS INFLATABLE PENILE PROSTHESIS (IPP) PRESENTED AT THE FOUR WEEK FOLLOW UP WITH EROSION OF THE PUMP AND A HOLE IN THEIR SCROTUM. THE PHYSICIAN TESTED FOR INFECTION AND CONFIRMED ABNORMAL FINDINGS. THE PATIENT WAS PLACED ON ANTIBIOTICS FOR TWO WEEKS, AND THEN PERFORMED A WASH OUT AND REPLACED THE PUMP. ANOTHER SWAB FOR INFECTION WAS PERFORMED AND RETURNED AS NORMAL. THERE WERE NO ADDITIONAL PATIENT COMPLICATIONS REPORTED.
Patient problems: Erosion; Unspecified Infection
MAUDE report 20599079 ↗InjuryApr 22, 2024Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT THE LEFT CYLINDER OF THE INFLATABLE PENILE PROSTHESIS RUPTURED AND THE DEVICE EXHIBITED FLUID LOSS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19161080 ↗InjuryMar 23, 2024Inadequacy of Device Shape and/or Size; Defective Device; Malposition of Device; Inadequacy of Device Shape and/or Size; Defective Device; Malposition of Device
IT WAS REPORTED THAT THE PATIENT STATED THIS INFLATABLE PENILE PROSTHESIS WAS TOO SHORT AND THE PUMP WAS DIFFICULT TO REACH. IN ADDITION, THE PATIENT FELT LIKE THE DISTAL TIPS WERE NOT PLACED PROPERLY. A SURGICAL PROCEDURE WAS PERFORMED TO REMOVE AND REPLACE CYLINDERS AND PUMP. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18966298 ↗InjuryMar 23, 2024Fluid/Blood Leak; Mechanical Problem; Material Puncture/Hole; Inadequacy of Device Shape and/or Size; Defective Device; Fluid/Blood Leak; Mechanical Problem; Material Puncture/Hole; Inadequacy of Device Shape and/or Size; Defective Device
IT WAS REPORTED THAT THIS INFLATABLE PENILE PROSTHESIS STOPPED WORKING AS THE PATIENT COULD NOT PUMP THE DEVICE. A SURGICAL PROCEDURE WAS PERFORMED, DURING WHICH FLUID LOSS CAUSED BY A HOLE IN THE CYLINDER TUBING WAS IDENTIFIED. IN ADDITION, IT WAS NOTED THAT THE IMPLANT WAS SIGNIFICANTLY SMALLER THAN WHAT WAS MEASURED. ALL COMPONENTS WERE REMOVED AND REPLACED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18966234 ↗InjuryMar 11, 2024Inflation Problem; Mechanical Problem; Inflation Problem; Mechanical Problem
IT WAS REPORTED THAT THE PUMP OF THIS INFLATABLE PENILE PROSTHESIS WAS NOT WORKING PROPERLY, AND THE CYLINDERS WERE NOT REMAINING INFLATED. A SURGICAL PROCEDURE WAS PERFORMED TO REMOVE AND REPLACE ALL THE COMPONENTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18882930 ↗InjuryJan 22, 2024Fluid/Blood Leak; Mechanical Problem; Material Puncture/Hole; Fluid/Blood Leak; Mechanical Problem; Material Puncture/Hole
IT WAS REPORTED THAT THIS INFLATABLE PENILE PROSTHESIS WAS NOT WORKING. THE DEVICE HAD FLUID LOSS. IT WAS IDENTIFIED THAT THE LEFT TUBING WHICH WAS CONNECTED TO THE PUMP WAS BROKEN. THE ENTIRE DEVICE WAS REPLACED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18554975 ↗InjuryFeb 14, 2023Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT THE PATIENT UNDERWENT SURGICAL INTERVENTION TO EXPLANT THE INFLATABLE PENILE PROSTHESIS (IPP) DUE TO THE DEVICE NO LONGER FUNCTIONING. UPON EXPLANT, THE TUBING BETWEEN THE PUMP AND CYLINDERS WERE BROKEN AND THE DEVICE HAD LOST FLUID. A NEW IPP WAS IMPLANTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16373554 ↗InjuryOct 28, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT UNDERWENT A SURGICAL PROCEDURE TO EXPLANT A PENILE PROSTHESIS FOR UNSPECIFIED REASONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15690975 ↗InjuryOct 25, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED BY THE PATIENT THAT THIS INFLATABLE PENILE PROSTHESIS (IPP) HAD NOT BEEN FUNCTIONING PROPERLY SINCE IMPLANT. TROUBLESHOOTING TECHNIQUES WERE PROVIDED BY PATIENT SERVICES (PS); HOWEVER, THE PATIENT STATED THAT HE HAD TRIED THEM ALL WITHOUT SUCCESS. HE ALSO REPORTED PAIN IN THE PERINEUM AND RECTUM AND MENTIONED THAT HE HAS BEEN PASSING STONES. THE BOSTON SCIENTIFIC PS REPRESENTATIVE ENCOURAGED HIM TO GET CLINICAL EVALUATION. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Pain
MAUDE report 15671555 ↗InjurySep 13, 2022Deflation Problem; Inflation Problem; Deflation Problem; Inflation Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INFLATABLE PENILE PROSTHESIS (IPP) REPLACEMENT SURGERY DUE TO INFLATION AND DEFLATION ISSUES. ALL COMPONENTS WERE REMOVED AND REPLACED WITH AN AMBICOR PENILE PROSTHESIS (APP). THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15412912 ↗InjuryAug 25, 2022Fluid/Blood Leak; Mechanical Problem; Fluid/Blood Leak; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT HAD THE INFLATABLE PENILE PROSTHESIS REMOVED AS IT STOPPED WORKING DUE TO A LEAK. THE PUMP WAS EMPTY UPON REMOVAL. A NEW INFLATABLE PENILE PROSTHESIS WAS IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15297039 ↗InjuryAug 22, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INFLATABLE PENILE PROSTHESIS (IPP) REPLACEMENT SURGERY BECAUSE THE DEVICE WAS NOT WORKING ANYMORE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15271811 ↗InjuryFeb 21, 2022Mechanical Problem; Material Puncture/Hole; Adverse Event Without Identified Device or Use Problem; Mechanical Problem; Material Puncture/Hole; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INFLATABLE PENILE PROSTHESIS (IPP) SURGERY DUE TO THE DEVICE NOT WORKING. IT WAS SAID THAT THE RIGHT CYLINDER WAS DISRUPTED. BOTH CYLINDERS OUTER LAYER DISLODGED FROM INNER FABRIC LAYER AND THE DACRON LAYER WAS TORN. THE DEVICE WAS REMOVED AND THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13573760 ↗InjuryFeb 15, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT A SURGICAL PROCEDURE TO REMOVE THIS INFLATABLE PENILE PROSTHESIS (IPP) BECAUSE IT WAS NON-FUNCTIONING. THE EXISTING DEVICE WAS REMOVED AND REPLACED WITH AN AMBICOR PENILE PROSTHESIS (APP). NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13536052 ↗
Other Prosthesis, Penis, Inflatable devices
- AMS 700 MS Pump 72404310Boston Scientific Corporation · Prosthesis, Penis, Inflatable992 reports
- AMS 700 LGX MS Pump 72404252Boston Scientific Corporation · Prosthesis, Penis, Inflatable1 death757 reports
- AMS 700 CX MS Pump 72404233-12Boston Scientific Corporation · Prosthesis, Penis, Inflatable656 reports
- AMS 700 CX MS Pump 72404232Boston Scientific Corporation · Prosthesis, Penis, Inflatable639 reports
- AMS 700 LGX MS Pump 72404251Boston Scientific Corporation · Prosthesis, Penis, Inflatable611 reports
- Conceal 720185-01Boston Scientific Corporation · Prosthesis, Penis, Inflatable586 reports
- AMS 700 LGX MS Pump 72404252-10Boston Scientific Corporation · Prosthesis, Penis, Inflatable561 reports
- AMS 700 CX MS Pump 72404232-10Boston Scientific Corporation · Prosthesis, Penis, Inflatable533 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗