Skip to content

AMS 700 CXR MS Pump 72404264Recalls & reported problems

Boston Scientific Corporation · Inflatable Penile Prosthesis, Preconnect Penoscrotal, 1 Pump to 2 Cylinders · Class 3 · Implant

All models of AMS 700 CXR MS Pump (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
11
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00878953003726), received between Oct 12, 2018 and Jul 30, 2026 — classified as 11 injuries.

The most frequently reported specific device problem is “Migration” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Migration3
  • Deflation Problem1
  • Difficult or Delayed Activation1
  • Fluid/Blood Leak1
  • Human-Device Interface Problem1
  • Inadequacy of Device Shape and/or Size1
  • Insufficient Device Problem Information1

Patient problems

  • Erosion3
  • No Clinical Signs, Symptoms or Conditions3
  • Pain2
  • Unspecified Tissue Injury2
  • Capsular Contracture1
  • No Known Impact Or Consequence To Patient1
  • Unspecified Infection1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 30, 2026Human-Device Interface Problem

    It was reported that the patient implanted with an Inflatable Penile Prosthesis (IPP) experienced scrotal pain when operating the Momentary Squeeze (MS) pump. A surgical procedure was performed, during which no device issue was identified. Although the pump continued to inflate the IPP as intended, its use caused significant pain to the patient. The physician noted difficulty when manipulating the MS pump and elected to replace it with a Tenacio pump. The new pump was implanted in a different scrotal pocket. In addition, a capsule believed to be contributing to the patient's pain was excised. No additional patient complications were reported.

    Patient problems: Capsular Contracture; Pain

    MAUDE report 50087091 ↗
  • InjuryJul 14, 2026Nonstandard Device

    THIS PRODUCT WAS RETURNED TO THE MANUFACTURER WITHOUT ANY REPORT OF PERFORMANCE ISSUES OR ADVERSE PATIENT EFFECTS. THE RETURNED DEVICE WAS ANALYZED AND A NON-CONFORMANCE WAS IDENTIFIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25820434 ↗
  • InjuryJan 7, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WITH AN INFLATABLE PENILE PROSTHESIS (IPP) EXPERIENCED CYLINDER PROTRUSION AGAINST THE SKIN WITH IMPENDING EROSION; THEREFORE, A SURGICAL PROCEDURE WAS PERFORMED TO REMOVE AND REPLACE THE CYLINDERS, PUMP, AND REAR-TIP EXTENDERS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Erosion; Unspecified Tissue Injury

    MAUDE report 23989545 ↗
  • InjuryOct 23, 2025Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT THIS INFLATABLE PENILE PROSTHESIS (IPP) PRESENTED DEFLATION ISSUES AFTER KEEPING INFLATED FOR AN EXTENDED PERIOD. A SURGICAL PROCEDURE WAS PERFORMED TO REMOVE AND REPLACE THE IPP. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23376254 ↗
  • InjuryMar 31, 2025Migration; Migration

    IT WAS REPORTED THAT THE PATIENT WITH THIS INFLATABLE PENILE PROSTHESIS (IPP) EXPERIENCE ERECTILE DYSFUNCTION. IT WAS NOTICED THAT THE DEVICE HAD PROXIMAL CROSSOVER. THERE WERE NO DEVICE ISSUES. THE PUMP AND CYLINDER WERE EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS REPORTED.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 21730226 ↗
  • InjurySep 20, 2024Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT THE PATIENT HAD A SURGICAL PROCEDURE TO REVISE AND REPLACE AN INFLATABLE PENILE PROSTHESIS (IPP) DUE TO PATIENT DISSATISFACTION. THE PATIENT FELT THAT THE CURRENT IPP WAS TOO NARROW AND WANTED TO UPSIZE THE DEVICE TO LARGER CYLINDERS. THE PATIENT IS EXPECTED TO FULLY RECOVER FOLLOWING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20268526 ↗
  • InjuryNov 29, 2023Migration; Migration

    IT WAS REPORTED THAT THE PATIENT UNDERWENT SURGICAL INTERVENTION TO EXPLANT THE INFLATABLE PENILE PROSTHESIS (IPP) DUE TO THE PUMP ERODING THROUGH THE SCROTUM. THERE WERE NO ADDITIONAL PATIENT COMPLICATIONS REPORTED.

    Patient problems: Erosion

    MAUDE report 18229572 ↗
  • InjuryOct 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INFLATABLE PENILE PROSTHESIS (IPP) REPLACEMENT SURGERY DUE TO INFECTION. ALL COMPONENTS WERE REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 17993230 ↗
  • InjuryMar 8, 2023Difficult or Delayed Activation; Adverse Event Without Identified Device or Use Problem; Difficult or Delayed Activation; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AROUND THE SCROTUM WHEN PUMPING HIS INFLATABLE PENILE PROSTHESIS (IPP) PUMP MULTIPLE TIMES. THE PUMP EXHIBITED ACTIVATION ISSUES. THE PATIENT UNDERWENT A REVISION SURGERY TO REPOSITION THE PUMP. THERE WERE NO FURTHER PATIENT COMPLICATIONS.

    Patient problems: Pain

    MAUDE report 16506858 ↗
  • InjuryNov 1, 2021Fluid/Blood Leak; Material Puncture/Hole; Adverse Event Without Identified Device or Use Problem; Migration; Fluid/Blood Leak; Material Puncture/Hole; Adverse Event Without Identified Device or Use Problem; Migration

    IT WAS REPORTED THAT THE PATIENT THE PATIENT HAD THE INFLATABLE PENILE PROSTHESIS PUMP AND CYLINDERS REMOVED AS THE CYLINDER HERNIATED OUT OF THE LEFT CORPORA. THE CYLINDERS RUBBED AGAINST EACH OTHER AND FORMED A SMALL HOLE THAT LED TO A FLUID LEAK. NEW INFLATABLE PENILE PROSTHESIS PUMP AND CYLINDERS COMPONENTS WERE IMPLANTED. THE PATIENT WAS EXPECTED TO FULLY RECOVER.

    Patient problems: Erosion

    MAUDE report 12729704 ↗
  • InjuryOct 12, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT HAD AN INFLATABLE PENILE PROSTHESIS EXPLANTED AND REPLACED WITH A 18CM X 9.5 MM IPP DUE TO THE DEVICE NOT WORKING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED IN RELATION TO THIS EVENT. FURTHER INFORMATION WAS REQUESTED BUT NOT YET RECEIVED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7960948 ↗

Other Prosthesis, Penis, Inflatable devices

All Prosthesis, Penis, Inflatable devices →

Sources and updates