Skip to content

AMS 700 LGX MS Pump 72404300Recalls & reported problems

Boston Scientific Corporation · Inflatable Penile Prosthesis, Preconnected 1 Pump to 2 Cylinders, Penoscrotal · Class 3 · Implant

All models of AMS 700 LGX MS Pump (18)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00878953005713), received between Jun 3, 2019 and Feb 11, 2025 — classified as 4 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Collapse” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Collapse2
  • Fracture2
  • Inflation Problem2
  • Material Puncture/Hole2
  • Mechanical Problem2
  • Defective Device1
  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient2
  • Pain1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 11, 2025Collapse; Inflation Problem; Collapse; Inflation Problem

    IT WAS REPORTED THAT THIS INFLATABLE PENILE PROSTHESIS (IPP) HAD INFLATION ISSUES. THE MAGNETIC RESONANCE IMAGING SHOWS SIGNS OF MALFUNCTION OF THE LEFT SHAFT OF THE INFLATABLE PENILE PROSTHESIS. ALSO, IT WAS NOTICED THAT THE PUMP STAYED FLAT. THE DEVICE IS STILL IMPLANTED, AND A SURGERY WILL BE BOOKED. NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21360252 ↗
  • MalfunctionSep 11, 2024Fracture; Material Puncture/Hole; Defective Device; Fracture; Material Puncture/Hole; Defective Device

    IT WAS REPORTED THAT THIS INFLATABLE PENILE PROSTHESIS (IPP) WAS NOT PERFORMING AS EXPECTED. THE PATIENT STATED THE DEVICE WAS LOSING RIGIDITY AS THE PATIENT WAS UNABLE TO PERFORM SEXUAL INTERCOURSE. AN MRI SCAN IDENTIFIED A POSSIBLE FRACTURE IN THE PROXIMAL SEGMENT OF THE LEFT SHAFT OF THE DEVICE. THE DEVICE REMAINS IMPLANTED, AND THE EXPLANT SURGERY IS ALREADY BOOKED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20195949 ↗
  • InjuryOct 18, 2021Mechanical Problem; Material Puncture/Hole; Adverse Event Without Identified Device or Use Problem; Mechanical Problem; Material Puncture/Hole; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD THE INFLATABLE PENILE PROSTHESIS CYLINDERS AND PUMP REMOVED DUE TO A CYLINDER MICRO PUNCTURE AND PUMP MALFUNCTION. NEW INFLATABLE PENILE PROSTHESIS CYLINDERS AND PUMP WERE IMPLANTED. FOLLOWING THE SURGERY THE EVENT RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12647467 ↗
  • InjurySep 14, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A SURGICAL PROCEDURE TO REVISE THIS INFLATABLE PENILE PROSTHESIS (IPP) DUE TO PAIN AT THE SURGICAL SITE. THE IPP WAS EXPLANTED AND NOT REPLACED WITH NO CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: Pain

    MAUDE report 12470981 ↗
  • InjuryJun 28, 2021Collapse; Mechanical Problem; Collapse; Mechanical Problem

    IT WAS REPORTED THAT THE PATIENT INDICATED THE INFLATABLE PENILE PROSTHESIS (IPP) PUMP FELT VERY HARD AFTER WORKING NORMALLY FOR THE FIRST THREE MONTHS. DURING AN APPOINTMENT WITH THE PHYSICIAN, THE PUMP INFLATED THE FIRST TIME IT WAS TESTED; HOWEVER, ON SUBSEQUENT ATTEMPTS THE PUMP WOULD REMAIN COLLAPSED. TROUBLESHOOTING WAS ATTEMPTED TO NO AVAIL BY FOLLOWING THE INSTRUCTIONS ON THE OPERATING ROOM MANUAL. A REPLACEMENT SURGERY WAS PERFORMED IN WHICH THE EXISTING DEVICE WAS REMOVED AND A NEW IPP WAS IMPLANTED. THERE WERE NO PATIENT COMPLICATIONS ASSOCIATED TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12075273 ↗
  • InjurySep 23, 2020Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture

    IT WAS REPORTED THAT DUE TO A TUBE CRACK AND FLUID LOSS THE PATIENT HAD HIS INFLATABLE PENILE PROSTHESIS (IPP) EXPLANTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10577616 ↗
  • MalfunctionJun 3, 2019Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED WITH THINNING OF THE PENIS AND AN INFLATED IMPLANT WITHOUT AXIAL STIFFNESS WITH A FOLD IN THE MIDDLE RESULTING IN A NONFUNCTIONAL PENIS NEW SCHEDULED SURGERY AND THE PROCEDURE WILL BE THE RECONSTRUCTION OF THE CAVERNOUS BODIES AND EXCHANGE OF PENILE PROSTHESIS. FURTHER INFORMATION WAS REQUESTED AND NOT YET RECEIVED. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8663239 ↗

Other Prosthesis, Penis, Inflatable devices

All Prosthesis, Penis, Inflatable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page