Modular Arthrodesis Nail CP260600Recalls & reported problems
Biomet Orthopedics, LLC · Class 2 · Implant
All models of Modular Arthrodesis Nail (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00880304082977), received between Oct 28, 2021 and Jun 12, 2024 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
2 other recalls cover devices cleared under the same FDA submission (K042409); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
- Loosening of Implant Not Related to Bone-Ingrowth1
- Malposition of Device1
- Migration1
- Off-Label Use1
- Unstable1
Patient problems
- Failure of Implant3
- No Clinical Signs, Symptoms or Conditions2
- Ambulation or Postural Difficulties1
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Joint Laxity1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryJun 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Ambulation or Postural Difficulties; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 19518876 ↗MalfunctionJun 12, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19518865 ↗InjuryNov 16, 2022Fracture; Unstable; Detachment of Device or Device Component; Migration; Fracture; Unstable; Detachment of Device or Device Component; Migration
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY. APPROXIMATELY THREE YEARS POST-IMPLANTATION, THE PATIENT BEGAN TO EXPERIENCE INSTABILITY AND IT WAS DETERMINED THAT THE LOCKING COLLAR DISASSOCIATED. SUBSEQUENTLY, THE PATIENT UNDERWENT REVISION SURGERY WHERE IT WAS REVEALED THAT THE SMALL LOCKING SCREWS OF THE COLLAR HAD BACKED OUT AND ONE LOCKING SCREW HAD FRACTURED. THE REVISION SURGERY WAS COMPLETED WITHOUT COMPLICATION. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant; Joint Laxity
MAUDE report 15803961 ↗InjuryFeb 25, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THE PATIENT WAS REVISED 3 MONTHS POST REVISION DUE TO LOOSENING. NO ADDITIONAL INFORMATION AVAILABLE. PREVIOUSLY, THE PATIENT UNDERWENT A KNEE REPLACEMENT PROCEDURE. THE PATIENT SPUN INTO EXTERNAL ROTATION AND WAS REVISED APPROXIMATELY 5 MONTHS POST IMPLANTATION. IT WAS ALSO NOTED- HCP USED A DUAL COMPRESS MOD A INTERCALARY SEGMENT AS A MID SHAFT FEMUR INTERCALARY. IT WAS KNOWN TO BE OFF LABEL.
Patient problems: Failure of Implant
MAUDE report 13601094 ↗InjuryFeb 11, 2022Off-Label Use; Malposition of Device; Off-Label Use; Malposition of Device
IT WAS REPORTED THE PATIENT UNDERWENT A KNEE PROCEDURE. SUBSEQUENTLY, THE PATIENT SPUN INTO EXTERNAL ROTATION AND WAS REVISED APPROXIMATELY 5 MONTHS POST IMPLANTATION. IT WAS ALSO NOTED- HCP USED A DUAL COMPRESS MOD A INTERCALARY SEGMENT AS A MID SHAFT FEMUR INTERCALARY. IT WAS KNOWN TO BE OFF LABEL. NO ADDITIONAL INFORMATION AT THIS TIME.
Patient problems: Failure of Implant
MAUDE report 13515208 ↗MalfunctionOct 28, 2021Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12712076 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K042409. One submission often covers many models, so check the product and lot details in each record.
- Z-1036-2023 · Nov 29, 2022 · Open, ClassifiedOSS Modular Arthrodesis Nail, 7 Degree Collar Assembly with Locking Bolts-For use in the replacement of segmental defects of long bone, including midshaft replacement, and arthrode
- Z-1522-2018 · Apr 3, 2018 · TerminatedOrthopedic Salvage System (OSS) Modular Arthrodesis 0 Degree Locking Collar, Model Number CP260600 Product Usage: The replacement of segmental defects of long bone, including mids
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page