EXPLOR 11-210063Recalls & reported problems
Biomet Orthopedics, LLC · Class 2 · Implant
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00880304210462), received between Mar 10, 2020 and Mar 30, 2026 — classified as 9 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
2 other recalls cover devices cleared under the same FDA submission (K040611, K051385); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Detachment of Device or Device Component3
- Corroded2
- Loosening of Implant Not Related to Bone-Ingrowth2
- Inadequacy of Device Shape and/or Size1
- Noise, Audible1
- Patient Device Interaction Problem1
- Unstable1
Patient problems
- Pain6
- Loss of Range of Motion5
- Failure of Implant3
- Metal Related Pathology3
- Aseptic loosening2
- Extraskeletal ossification2
- Joint Contracture2
- Scar Tissue2
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryMar 30, 2026Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT A PATIENT UNDERWENT AN INITIAL LEFT ELBOW ARTHROPLASTY. SUBSEQUENTLY THE PATIENT WAS REVISED APPROXIMATELY 4 MONTHS AND 16 DAYS POST IMPLANTATION DUE TO LOOSENING, ECTOPIC BONE FORMATION, ARTHROLYSIS, AND PAIN. DURING THE REVISION NOTED SCAR TISSUE, IMPLANT ¿CATCHING¿ AND WEAR. THE HEAD WAS EXCHANGED FOR A SMALLER SIZE WITHOUT COMPLICATIONS.ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Aseptic loosening; Extraskeletal ossification; Joint Contracture; Loss of Range of Motion; Pain; Scar Tissue
MAUDE report 24716380 ↗InjuryAug 1, 2025Corroded; Corroded
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL ELBOW ARTHROPLASTY APPROXIMATELY FOUR (4) YEARS AND SIX (6) MONTHS AGO. SUBSEQUENTLY, THE PATIENT UNDERWENT A REVISION SURGERY DUE TO SOFT TISSUE METALLOSIS AND DESTRUCTION OF THE CAPITULUM HUMERI SURFACE.
Patient problems: Metal Related Pathology
MAUDE report 22689111 ↗InjuryAug 1, 2025Corroded; Corroded
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL ELBOW ARTHROPLASTY APPROXIMATELY FOUR (4) YEARS AGO. SUBSEQUENTLY, THE PATIENT THE PATIENT UNDERWENT A REVISION SURGERY ON AN UNKNOWN DATE DUE TO SOFT TISSUE METALLOSIS AND DESTRUCTION OF THE CAPITULUM HUMERI SURFACE.
Patient problems: Metal Related Pathology
MAUDE report 22687256 ↗InjuryJul 17, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Aseptic loosening; Loss of Range of Motion; Pain
MAUDE report 22538414 ↗InjuryJul 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT EXPERIENCED PAIN AND STIFFNESS WHICH LED TO THE REMOVAL OF THE IMPLANTS. ATTEMPTS HAVE BEEN MADE AND THERE IS NO FURTHER INFORMATION AT THIS TIME.
Patient problems: Pain
MAUDE report 19651278 ↗InjuryMar 4, 2024Detachment of Device or Device Component; Patient Device Interaction Problem; Detachment of Device or Device Component; Patient Device Interaction Problem
IT WAS REPORTED THAT A EXPLORE RADIAL HEAD IMPLANT DISASSOCIATED FROM STEM AFTER 3 YEARS OF BEING IMPLANTED. THE SURGEON REVISED, TAKING THE DISSOCIATED HEAD OUT WHICH HAD THE SCREW IN IT. THE STEM WAS STABLE AND STAYED IN. HE CLEANED THE FEMALE THREADS CHECKED TO MAKE SURE THE THREADS WEREN'T STRIPPED BY USING A SCREW AND TORQUING ON IT. NOTHING WAS CROSS THREADED OR STRIPPED. HE THEN RETRIED AND IMPLANTED A NEW HEAD WITH A NEW SCREW. ATTEMPTS HAVE BEEN MADE AND THERE IS NO FURTHER INFORMATION AT THIS TIME.
Patient problems: Failure of Implant; Metal Related Pathology
MAUDE report 18826839 ↗InjurySep 21, 2023Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Noise, Audible; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Noise, Audible
IT WAS REPORTED THAT THE PATIENT UNDERWENT A TRAUMA PROCEDURE APPROXIMATELY SEVEN (7) YEARS AGO FOLLOWING A FALL THAT FRACTURED THE LEFT ULNAR AND RADIAL HEAD OF THE HUMERUS. SUBSEQUENTLY, THE PATIENT BEGAN EXPERIENCING PAIN, DECREASED ROM, SWELLING AND NOISE APPROXIMATELY FIVE (5) YEARS AND SIX (6) MONTHS POST-IMPLANTATION. X-RAY IMAGES WERE TAKEN AND REVEALED THE RADIAL HEAD HAD DISASSOCIATED FROM STEM AND PROGRESSION OF ARTHRITIS WAS NOTED. THE PATIENT LATER UNDERWENT A REVISION SURGERY APPROXIMATELY FOUR (4) MONTHS AGO. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant; Joint effusion; Loss of Range of Motion; Pain; Swelling/ Edema
MAUDE report 17795189 ↗InjuryMay 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Discomfort; Extraskeletal ossification; Joint Contracture; Loss of Range of Motion; Pain; Scar Tissue
MAUDE report 16947510 ↗InjuryMar 10, 2020Unstable; Detachment of Device or Device Component; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Detachment of Device or Device Component; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant; Loss of Range of Motion; Pain
MAUDE report 9810964 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K040611, K051385. One submission often covers many models, so check the product and lot details in each record.
- Z-0269-2018 · Jun 16, 2017 · TerminatedExploR Radial Head Replacement System. PN: 11-210061 LN: 243610. The stem is placed inside a polyurethane bag. The bagged component will then be placed in an inner PETG tray. The
- Z-1615-2014 · Apr 22, 2014 · TerminatedExploR 7x26mm Modular Radial Stem, Catalog number 11-210062. Elbow prosthesis orthopedic implant.
Other PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER devices
- EVOLVE 496S275Wright Medical Technology, Inc. · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall6 reports
- NA 09.405.966SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall2 reports
- NA 03.405.000SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall2 reports
- NA 09.405.250SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall1 report
- NA 09.405.263SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall1 report
- NA 09.405.570SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall1 report
- EXPLOR 11-210061Biomet Orthopedics, LLC · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall1 report
- NA 09.405.583SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.