Skip to content

COMPREHENSIVE SHOULDER SYSTEM 113631Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

All models of Comprehensive Shoulder System (68)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00880304462687), received between Nov 4, 2020 and Aug 29, 2023 — classified as 2 injuries.

1 FDA recall record name this device's identifier.

6 other recalls cover devices cleared under the same FDA submission (K060692); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Jan 7, 2022Terminated

    The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

    Z-0674-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Pain2
  • Hypersensitivity/Allergic reaction1
  • Metal Related Pathology1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 29, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PATIENT UNDERWENT A PROCEDURE APPROXIMATELY 4 YEARS AGO. SUBSEQUENTLY, THE PATIENT WAS REVISED ABOUT A MONTH AGO DUE TO PAIN, REACTIVE TISSUE, FLUID IN THE CAPSULE, AND TISSUE BETWEEN THE BASEPLATE AND GLENOSPHERE. THERE IS POSSIBLE METALLOSIS OR ALLERGY. ATTEMPTS HAVE BEEN MADE AND THERE IS NO FURTHER INFORMATION AT THIS TIME.

    Patient problems: Hypersensitivity/Allergic reaction; Metal Related Pathology; Pain

    MAUDE report 17643911 ↗
  • InjuryNov 4, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A HEMI-ARTHROPLASTY AROUND 2-3 YEARS AGO. SUBSEQUENTLY, PATIENT WAS REVISED DUE TO SHOULDER PAIN. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: Pain

    MAUDE report 10786546 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K060692. One submission often covers many models, so check the product and lot details in each record.

  • Z-1543-2020 · Feb 17, 2020 · Terminated
    Comprehensive Shoulder System Mini Humeral Stem 9mm,, Model No. 113629 - Product Usage: The intended use of the comprehensive humeral stem mini is to act as the humeral component o
  • Z-0591-2019 · Oct 10, 2018 · Terminated
    Comprehensive Mini Stem, Item No. 113631:
  • Z-0979-2018 · Dec 8, 2017 · Terminated
    Humeral Stem 83mm(Size 17) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
  • Z-0980-2018 · Dec 8, 2017 · Terminated
    Humeral Stem 55mm(Size 13) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
  • Z-0299-2018 · Jun 5, 2017 · Terminated
    COMP PRIMARY STEM 8MM MINI, model # 113628
  • Z-1755-2015 · Apr 16, 2015 · Terminated
    Comprehensive Shoulder System, 4 mm, 122 mm long, Primary Shoulder Stem, Ti-6Al-4V, Standard Length Porous Plasma. Shoulder joint replacement components include humeral stems, hum

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page