Comprehensive Reverse Shoulder 115370Recalls & reported problems
Biomet Orthopedics, LLC · Class 2 · Implant
Quick summary
This Biomet Comprehensive Reverse Shoulder implant has 42 adverse event reports, with 31 injuries, 11 malfunctions and no deaths. Reports most often describe parts of the implant fracturing — especially the humeral tray or its stem — along with pain, dislocation and components that were hard to position, frequently leading to revision surgery years after implantation. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 42
- Deaths reported
- 0
- Injuries reported
- 31
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 00880304543201), received between Oct 17, 2019 and Mar 25, 2026 — classified as 31 injuries, 11 malfunctions.
The most frequently reported specific device problem is “Fracture” (9 reports).
19 other recalls cover devices cleared under the same FDA submission (K080642, K113069, K193373); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem13
- Fracture9
- Difficult to Insert5
- Positioning Failure5
- Detachment of Device or Device Component2
- Device Dislodged or Dislocated2
- Naturally Worn2
- Noise, Audible2
Patient problems
- Pain17
- Failure of Implant8
- No Clinical Signs, Symptoms or Conditions6
- Joint Dislocation4
- No Known Impact Or Consequence To Patient4
- Erosion3
- Insufficient Information3
- Loss of Range of Motion2
Recent adverse event reports
The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.
InjuryMar 25, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT WITH A PRIOR RIGHT REVERSE TOTAL SHOULDER ARTHROPLASTY SUBSEQUENTLY UNDERWENT A RIGHT SHOULDER BIOPSY WITH POSSIBLE POLYETHYLENE LINER EXCHANGE. AS IT IS UNKNOWN WHAT THE REASON FOR REVISION WAS, ALL IMPLANTS ARE INCLUDED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Insufficient Information
MAUDE report 24688463 ↗InjuryNov 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Pain
MAUDE report 23660836 ↗InjuryNov 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Discomfort; Pain
MAUDE report 23575257 ↗InjuryOct 29, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Pain
MAUDE report 23412218 ↗InjurySep 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Extraskeletal ossification; Pain; Tendinopathy
MAUDE report 23100067 ↗InjuryAug 8, 2025Fracture; Fracture
IT WAS REPORTED A PATIENT HAD AN INITIAL RIGHT TOTAL SHOULDER ARTHROPLASTY FOLLOWED BY A REVISION ON APPROXIMATELY 6 YEARS POST-IMPLANTATION DUE TO PAIN. SUBSEQUENTLY, REPORTED AN UNKNOWN IMPLANT BROKE AND UNDERWENT A SECOND REVISION APPROXIMATELY 7 YEARS POSTERIOR. THE BEARING AND HUMERAL TRAY WERE REVISED AND NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Insufficient Information
MAUDE report 22748410 ↗InjuryMay 19, 2025Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT WAS IMPLANTED WITH A TOTAL SHOULDER ARTHOPLASTY. SUBSEQUENTLY, THE PATIENT WAS REVISED DUE TO FRACTURE OF THE HUMERAL TRAY. THERE WAS HARM/INJURY TO THE PATIENT AS THE PATIENT HAD TO UNDERWENT SURGICAL/MEDICAL INTERVENTION. DUE DILIGENCE IS COMPLETE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 22044501 ↗InjuryMay 5, 2025Fracture; Noise, Audible; Fracture; Noise, Audible
IT WAS REPORTED THAT THE PATIENT UNDERWENT A SHOULDER REVISION APPROXIMATELY NINE (9) YEARS AND FIVE (5) MONTHS POST-IMPLANTATION DUE TO EXPERIENCING A FRACTURE OF THE HUMERAL TRAY'S STEM/KNOB. IT WAS NOTED THE PATIENT WAS EXERCISING WHEN THEY HEARD A POP, FOLLOWED BY EXPERIENCING PAIN. IT WAS NOTED BECAUSE THE HUMERAL TRAY'S KNOB WAS FRACTURED OFF INTO THE HUMERAL STEM, THE STEM ALSO HAD TO BE REMOVED, CAUSING THE REVISION TO BE PROLONGED 2 HOURS LONGER THAN ANTICIPATED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant; Pain
MAUDE report 21960021 ↗InjuryMar 14, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Insufficient Information
MAUDE report 21611279 ↗InjurySep 18, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
NO ADDITIONAL INFORMATION IS AVAILABLE REGARDING THE INCIDENT.
Patient problems: Failure of Implant
MAUDE report 20249435 ↗InjuryJun 11, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Loss of Range of Motion; Pain
MAUDE report 19509351 ↗InjuryApr 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED A PATIENT WAS PRESCRIBED PAIN MEDICATION APPROXIMATELY 2.5 YEARS POST IMPLANTATION OF AN RTSA DUE TO CHRONIC SHOULDER PAIN. ALL DEVICES REMAIN IMPLANTED AND NO FURTHER INTERVENTION HAS OCCURRED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Pain
MAUDE report 19035632 ↗MalfunctionMar 4, 2024Positioning Failure; Fracture; Positioning Failure; Fracture
IT WAS REPORTED DURING A TOTAL SHOULDER ARTHROPLASTY, THE SURGEON WAS ATTEMPTING TO SEAT THE POLYETHYLENE BEARING WITHIN THE HUMERAL TRAY USING MANUAL PRESSURE ON THE BACK TABLE. AFTER MULTIPLE ATTEMPTS, THE IMPLANTS ASSEMBLED. THE SURGEON NOTED THAT THE RING APPEARED TO BE BENT INITIALLY, AND AFTER ASSEMBLY, THE METAL RING WAS OBSERVED TO BE FRACTURED AT THE EDGE OF THE RING, RESULTING IN THE TRAY AND POLYETHYLENE BEARING COMING APART WITH LITTLE EFFORT. THE SURGEON USED A BACKUP TRAY AND THE ORIGINAL BEARING TO COMPLETE THE PROCEDURE WITHOUT FURTHER ISSUES. THERE WAS NO HARM TO THE PATIENT AS A RESULT OF THE MALFUNCTION. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18825677 ↗InjuryAug 29, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THE PATIENT UNDERWENT A PROCEDURE APPROXIMATELY 4 YEARS AGO. SUBSEQUENTLY, THE PATIENT WAS REVISED ABOUT A MONTH AGO DUE TO PAIN, REACTIVE TISSUE, FLUID IN THE CAPSULE, AND TISSUE BETWEEN THE BASEPLATE AND GLENOSPHERE. THERE IS POSSIBLE METALLOSIS OR ALLERGY. ATTEMPTS HAVE BEEN MADE AND THERE IS NO FURTHER INFORMATION AT THIS TIME.
Patient problems: Hypersensitivity/Allergic reaction; Metal Related Pathology; Pain
MAUDE report 17643718 ↗InjuryJun 28, 2023Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant; Joint Dislocation; Pain
MAUDE report 17219154 ↗InjuryJun 6, 2023Material Erosion; Naturally Worn; Material Erosion; Naturally Worn
IT WAS REPORTED THAT THE PATIENT UNDERWENT A RIGHT SHOULDER REVERSE TOTAL SHOULDER ARTHROPLASTY APPROXIMATELY SEVEN (7) YEARS AND (B)(6) AGO, SUBSEQUENTLY, THE PATIENT UNDERWENT A REVISION SURGERY DUE TO METALOSIS OF THE IMPLANTS. THE PATIENT IS SCHEDULED FOR A SECOND REVISION SURGERY IN APPROXIMATELY THREE (3) WEEKS TO IMPLANT CUSTOM IMPLANTS AND COMPLETE THE REVISION.
Patient problems: Erosion; Failure of Implant; Joint Dislocation; Metal Related Pathology; Osteolysis; Osteopenia/ Osteoporosis; Pain
MAUDE report 17072920 ↗MalfunctionMay 9, 2023Positioning Failure; Positioning Failure
NO FURTHER EVENT INFORMATION AT TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16894997 ↗InjuryMar 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Pain; Peripheral Nervous Injury
MAUDE report 16629971 ↗InjuryMar 27, 2023Fracture; Fracture
IT WAS REPORTED A PATIENT UNDERWENT A LEFT REVERSE SHOULDER ARTHROPLASTY APPROXIMATELY FIVE YEARS AGO. WHILE PUSHING ON A TIRE, THE PATIENT FELT THE SHOULDER POP. THE PATIENT WAS REVISED WHERE THE HUMERAL BASEPLATE WAS BROKEN. DURING THE REVISION THE BASEPLATE WAS FOUND TO BE DISLOCATED POSTERIORLY AND NOTED TO HAVE SHEARED AT THE JUNCTION OF THE PEG AND PLATE, WITH THE PEG STILL ATTACHED WITHIN THE HUMERAL STEM. MEDICAL RECORDS INDICATE THE STEM WAS WELL FIXED TO THE HUMERUS. IN AN ATTEMPT TO REMOVE THE BROKEN PIECES FROM THE STEM, THE STEM DISLODGED FROM THE HUMERUS. THE STEM, TIBIAL TRAY, AND BEARING WERE ALL REVISED WITHOUT COMPLICATION. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 16619079 ↗MalfunctionMar 20, 2023Positioning Failure; Material Deformation; Positioning Failure; Material Deformation
NO FURTHER EVENT INFORMATION AT TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16573758 ↗InjuryFeb 15, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED A PATIENT HAD A CONVERSION FROM A HEMIARTHROPLASTY TO A REVERSE SHOULDER ARTHROPLASTY APPROXIMATELY THREE (3) YEARS AND EIGHT (8) MONTHS AGO. SUBSEQUENTLY, ONE (1) YEAR AND SEVEN (7) MONTHS LATER, PATIENT REPORTS UNUSUAL PAIN FOLLOWING A FALL ONTO OUTSTRETCHED HANDS. THE PATIENT UNDERWENT A COURSE OF PHYSICAL THERAPY AND TAKES OTC MEDICATIONS AS NEEDED FOR PAIN. APPROXIMATELY TWO (2) MONTHS AGO, RADIOGRAPHIC IMAGING DISPLAYED STRESS SHIELDING IN LATERAL CORTEX AND THE PATIENT IS EXPERIENCING SOME DIFFICULTY PERFORMING DAILY ACTIVITIES. NO PLAN FOR REVISION HAS BEEN REPORTED AND THE PATIENT HAS REMAINED EXTREMELY SATISFIED THROUGHOUT THE STUDY. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Erosion; Fall; Pain
MAUDE report 16382709 ↗MalfunctionNov 30, 2022Positioning Failure; Detachment of Device or Device Component; Positioning Failure; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AT TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15887800 ↗InjuryAug 11, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED PATIENT IS EXPERIENCING SHOULDER PAIN, INSTABILITY, AND ROM ISSUES APPROXIMATELY SEVEN YEARS POST IMPLANTATION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Pain
MAUDE report 15212152 ↗InjuryMar 2, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT WAS REVISED APPROXIMATELY 1 YEAR AND 2 DAYS POST IMPLANTATION DUE TO A FOREIGN OBJECT IDENTIFIED ON POST OP RADIOGRAPHS. THE OBJECT WAS REMOVED AND ALL COMPONENTS WERE REMOVED AND REPLACED. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Foreign Body In Patient
MAUDE report 13653259 ↗InjuryMar 1, 2022Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVERSE SHOULDER ARTHROPLASTY ON AN UNKNOWN DATE. SUBSEQUENTLY, IN AN UNKNOWN TIMEFRAME POST-IMPLANTATION, THE PATIENT WAS REVISED DUE TO ASEPTIC LOOSENING WITH PERIPROSTHETIC OSTEOLYSIS, WEAR OF THE POLY COMPONENT, AND METALLOSIS. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant; Osteolysis
MAUDE report 13634059 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K080642, K113069, K193373. One submission often covers many models, so check the product and lot details in each record.
- Z-0619-2021 · Nov 25, 2020 · TerminatedComprehensive Reverse Shoulder System, Item numbers 115394, 115395, 115396, 115398, 180550, 180551, 180552, 180553, 180554, 180555, 180556, 180557, 180558, 180559, 180560, 1805
- Z-0093-2021 · Sep 2, 2020 · TerminatedBiomet Comprehensive Reverse Shoulder Glenosphere Mini Baseplate with Taper Adapter-Shoulder Prosthesis item: 010000589 - Product Usage: intended to serve as the foundation for t
- Z-2144-2020 · Apr 17, 2020 · TerminatedExtremities Item Number: 1)113954 Hybrid Glenoid Glenoid Base, 4 MM 2) XL-115366 Comprehensive Reverse Shoulder Humeral Bearing, ArComXL, 44 MM X 41 MM, Standard 3) XL-115364 Comp
- Z-1750-2020 · Mar 13, 2020 · TerminatedComprehensive Humeral Tray Taper Extraction Pliers - Replacement Tips, Model Number 110028522 - Product Usage: Intended to be used in surgery to relieve pain and to replace the gle
- Z-1859-2017 · Feb 17, 2017 · Terminatedvarious polyethylene implants Extremities, Shoulders and Elbows BioModular Shoulder System; Comprehensive Shoulder System; Comprehensive Reverse Shoulder System Product Usage: Sh
- Z-1103-2017 · Dec 20, 2016 · TerminatedComprehensive Reverse Shoulder System Humeral Tray Model 115340
- Z-1548-2013 · May 15, 2013 · TerminatedSuperior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain.
- Z-2550-2011 · Apr 22, 2011 · TerminatedCustom comprehensive reverse shoulder bio-modular base 44 mm humeral tray standard, REF CP561510, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder product
- Z-2547-2011 · Apr 22, 2011 · TerminatedCustom solar comprehensive reverse humeral tray with lock ring 44 mm +5mm, sterile, REF CP561659, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder product
- Z-2546-2011 · Apr 22, 2011 · TerminatedCustom bi-angular to reverse shoulder humeral tray 44 mm sterile, REF CP561548, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder products are indicated fo
- Z-2544-2011 · Apr 22, 2011 · TerminatedComprehensive reverse shoulder humeral tray with locking ring, 44 mm +5 mm, sterile, single use, REF 115345, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulde
- Z-2545-2011 · Apr 22, 2011 · TerminatedComprehensive reverse shoulder humeral tray with locking ring, 44 mm standard, sterile, single use, REF 115340, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Sho
- Z-2541-2011 · Apr 22, 2011 · TerminatedBio-modular reverse shoulder humeral tray with lock ring, 44 mm 10 degree angle, sterile, single use, REF 11-113685, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse S
- Z-2543-2011 · Apr 22, 2011 · TerminatedComprehensive reverse shoulder humeral tray with locking ring, 44 mm +10 mm, sterile, single use, REF 115348, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Should
- Z-2542-2011 · Apr 22, 2011 · TerminatedBio-modular reverse shoulder humeral tray with lock ring, 44 mm standard, sterile, single use, REF 11-113680, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder
…and 4 more.
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.