COMPREHENSIVE SHOULDER SYSTEM 113618Recalls & reported problems
Biomet Orthopedics, LLC · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00880304554191), received between Dec 7, 2021 and Oct 30, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Markings/Labelling Problem” (1 report).
1 FDA recall record name this device's identifier.
6 other recalls cover devices cleared under the same FDA submission (K060692); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jan 7, 2022Terminated
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.
Z-0673-2022 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Device Markings/Labelling Problem1
- Malposition of Device1
Patient problems
- Bone Fracture(s)1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 30, 2023Malposition of Device; Malposition of Device
IT WAS REPORTED A PATIENT UNDERWENT A SHOULDER ARTHROPLASTY APPROXIMATELY 2 WEEKS AGO. SUBSEQUENTLY, THE PATIENT WAS REVISED DUE TO INCORRECT ALIGNMENT OF IMPLANTS. THE BASEPLATE WAS LEFT IN, BUT THERE WAS A SIZE EXCHANGE OF THE OTHER COMPONENTS. ATTEMPTS HAVE BEEN MADE AND THERE IS NO FURTHER INFORMATION AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18029635 ↗InjuryDec 7, 2021Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT SHOULDER REPLACEMENT SURGERY. DURING THE PROCEDURE THE SURGEON ASKED FOR A 18X55 STEM. AFTER VERIFYING ON THE BOX THE CORRECT SIZE, THE STEM WAS HANDED TO THE CIRCULATOR TO OPEN ONTO THE STERILE FIELD. AS THE SURGEON WAS HANDED THE IMPLANT, HE NOTICED THAT IT WAS WAY TOO BIG COMPARED TO WHAT WE NORMALLY IMPLANT. PACKAGE WAS CHECKED AND CONFIRMED EVERYTHING SAID IT WAS SUPPOSED TO BE A 18X55 MICRO STEM BUT INSTEAD THE IMPLANT WAS A 19X122 STEM. KNOWING ANOTHER SET OF MICRO STEMS WERE NOT AVAILABLE, THE SURGEON ELECTED TO USE A MINI STEM WHICH RESULTED IN A FRACTURE OF THE HUMOROUS. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Bone Fracture(s)
MAUDE report 12952279 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K060692. One submission often covers many models, so check the product and lot details in each record.
- Z-1543-2020 · Feb 17, 2020 · TerminatedComprehensive Shoulder System Mini Humeral Stem 9mm,, Model No. 113629 - Product Usage: The intended use of the comprehensive humeral stem mini is to act as the humeral component o
- Z-0591-2019 · Oct 10, 2018 · TerminatedComprehensive Mini Stem, Item No. 113631:
- Z-0980-2018 · Dec 8, 2017 · TerminatedHumeral Stem 55mm(Size 13) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
- Z-0979-2018 · Dec 8, 2017 · TerminatedHumeral Stem 83mm(Size 17) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
- Z-0299-2018 · Jun 5, 2017 · TerminatedCOMP PRIMARY STEM 8MM MINI, model # 113628
- Z-1755-2015 · Apr 16, 2015 · TerminatedComprehensive Shoulder System, 4 mm, 122 mm long, Primary Shoulder Stem, Ti-6Al-4V, Standard Length Porous Plasma. Shoulder joint replacement components include humeral stems, hum
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page