Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00880304677180), received between Dec 1, 2020 and May 4, 2026 — classified as 13 injuries.
The most frequently reported specific device problem is “Fracture” (10 reports).
19 other recalls cover devices cleared under the same FDA submission (K080642, K130390, K132239); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture10
- Loosening of Implant Not Related to Bone-Ingrowth3
- Difficult to Insert1
Patient problems
- Failure of Implant7
- Pain2
- Ambulation or Postural Difficulties1
- Aseptic loosening1
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- Inadequate Osseointegration1
- Insufficient Information1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
InjuryMay 4, 2026Fracture
IT WAS REPORTED BY THE PMI GROUP THAT A PATIENT UNDERWENT A SHOULDER ARTHROPLASTY THREE (3) YEARS AND SIX (6) MONTHS AGO TO RECEIVE A VRS SHOULDER. SUBSEQUENTLY, THE PATIENT WAS REVISED ON SIX (6) MONTHS POST-IMPLANTATION DUE TO SCREW FRACTURE AND RECEIVED A COMPETITOR BASEPLATE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Insufficient Information
MAUDE report 25068503 ↗InjuryOct 17, 2025Fracture; Fracture
INITIAL RIGHT REVERSE TOTAL SHOULDER PERFORMED ON AN UNKNOWN DATE. SUBSEQUENTLY, IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO BASEPLATE SCREWS BEING BROKEN, ALL IMPLANTS WERE EXPLANTED, AND A GLENOID BONE GRAFT APPLIED. SPACER WAS IMPLANTED DURING THE REVISION, SUSPECTED POSSIBLE INFECTION. PATIENT WILL BE REIMPLANTED AT A LATER DATE. GLENOSPHERE WAS ALSO NOTED TO BE SUPERIORLY INCLINED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Medical device site infection; No Clinical Signs, Symptoms or Conditions
MAUDE report 23321522 ↗InjuryMay 28, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT THE PATIENT UNDERWENT A LEFT SHOULDER ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT WAS CONSIDERED FOR A PMI PRODUCT AND REVISED DUE TO LOOSENING AND INSTABILITY OF THE IMPLANT. DURING THE REVISION, IT WAS NOTED THAT THE CENTRAL SCREW WAS FRACTURED. ALL COMPONENTS WERE REMOVED AND REPLACED; HOWEVER, A PIECE OF THE FRACTURED CENTRAL SCREW COULD NOT BE REMOVED AS IT WAS RETAINED IN THE GLENOID.
Patient problems: Aseptic loosening
MAUDE report 22101929 ↗InjuryMay 16, 2025Fracture; Fracture
IT WAS REPORTED THE PATIENT WAS UNDERGOING A RIGHT SHOULDER IMPLANT AND WHILE IMPLANTING A PERIPHERAL SCREW, THE PHYSICIAN COULDN¿T SCREW CLOCKWISE OR COUNTERCLOCKWISE, AND THE HEAD OF THE SCREW FRACTURED. THE REST OF THE SCREW WAS FAR ENOUGH IN AND IT REMAINED IMPLANTED ON THE PATIENT. THERE IS NO PLAN TO REMOVE THE FRACTURED SCREW AT THIS TIME. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Foreign Body In Patient
MAUDE report 22040031 ↗InjuryNov 18, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS FURTHER REPORTED THAT A PATIENT UNDERWENT A REVISION PROCEDURE APPROXIMATELY 13 MONTHS POST IMPLANTATION DUE TO GLENOID LOOSENING AND PAIN IN THE SHOULDER. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Failure of Implant; Pain
MAUDE report 20711762 ↗InjuryOct 1, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Inadequate Osseointegration
MAUDE report 20346750 ↗InjuryJun 25, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Ambulation or Postural Difficulties; Failure of Implant; Pain
MAUDE report 19606507 ↗InjuryJul 14, 2023Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT A SECOND STAGE REVISION PROCEDURE. SUBSEQUENTLY, THE PATIENT UNDERWENT A REVISION PROCEDURE DUE TO FRACTURED SCREWS APPROXIMATELY FOUR (4) MONTHS POST-IMPLANTATION.
Patient problems: Failure of Implant
MAUDE report 17324312 ↗InjuryJun 16, 2023Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant
MAUDE report 17143672 ↗InjuryMay 31, 2023Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT A SHOULDER ARTHROPLASTY ON AN UNKNOWN DATE. THE PATIENT UNDERWENT A REVISION SURGERY OVER A YEAR AND A HALF AGO DUE TO FRACTURED SCREWS. SUBSEQUENTLY, THE PATIENT UNDERWENT A SECOND REVISION SURGERY DUE TO FRACTURED VRS SCREWS. PATIENT WAS IMPLANTED WITH A SHOULDER HEMI. THIS IS THE THIRD REVISION SURGERY AND SECOND SURGERY DUE TO FRACTURING OF IMPLANTS.
Patient problems: Failure of Implant
MAUDE report 17031056 ↗InjuryAug 3, 2021Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant
MAUDE report 12267196 ↗InjuryAug 3, 2021Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant
MAUDE report 12267190 ↗InjuryDec 1, 2020Fracture; Difficult to Insert; Fracture; Difficult to Insert
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10927122 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K080642, K130390, K132239. One submission often covers many models, so check the product and lot details in each record.
- Z-1131-2021 · Jan 15, 2021 · TerminatedCentral Screw 6.5x25mm - Product Usage: used as a sterile screw that is intended for implantable shoulder prosthesis assembly and/or fixation.
- Z-0619-2021 · Nov 25, 2020 · TerminatedComprehensive Reverse Shoulder System, Item numbers 115394, 115395, 115396, 115398, 180550, 180551, 180552, 180553, 180554, 180555, 180556, 180557, 180558, 180559, 180560, 1805
- Z-2144-2020 · Apr 17, 2020 · TerminatedExtremities Item Number: 1)113954 Hybrid Glenoid Glenoid Base, 4 MM 2) XL-115366 Comprehensive Reverse Shoulder Humeral Bearing, ArComXL, 44 MM X 41 MM, Standard 3) XL-115364 Comp
- Z-1750-2020 · Mar 13, 2020 · TerminatedComprehensive Humeral Tray Taper Extraction Pliers - Replacement Tips, Model Number 110028522 - Product Usage: Intended to be used in surgery to relieve pain and to replace the gle
- Z-1859-2017 · Feb 17, 2017 · Terminatedvarious polyethylene implants Extremities, Shoulders and Elbows BioModular Shoulder System; Comprehensive Shoulder System; Comprehensive Reverse Shoulder System Product Usage: Sh
- Z-1103-2017 · Dec 20, 2016 · TerminatedComprehensive Reverse Shoulder System Humeral Tray Model 115340
- Z-1548-2013 · May 15, 2013 · TerminatedSuperior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain.
- Z-2545-2011 · Apr 22, 2011 · TerminatedComprehensive reverse shoulder humeral tray with locking ring, 44 mm standard, sterile, single use, REF 115340, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Sho
- Z-2546-2011 · Apr 22, 2011 · TerminatedCustom bi-angular to reverse shoulder humeral tray 44 mm sterile, REF CP561548, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder products are indicated fo
- Z-2548-2011 · Apr 22, 2011 · TerminatedCustom biomodular humeral tray 44 mm sterile, REF CP561686, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder products are indicated for use in patients who
- Z-2544-2011 · Apr 22, 2011 · TerminatedComprehensive reverse shoulder humeral tray with locking ring, 44 mm +5 mm, sterile, single use, REF 115345, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulde
- Z-2543-2011 · Apr 22, 2011 · TerminatedComprehensive reverse shoulder humeral tray with locking ring, 44 mm +10 mm, sterile, single use, REF 115348, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Should
- Z-2547-2011 · Apr 22, 2011 · TerminatedCustom solar comprehensive reverse humeral tray with lock ring 44 mm +5mm, sterile, REF CP561659, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder product
- Z-2541-2011 · Apr 22, 2011 · TerminatedBio-modular reverse shoulder humeral tray with lock ring, 44 mm 10 degree angle, sterile, single use, REF 11-113685, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse S
- Z-2549-2011 · Apr 22, 2011 · TerminatedCustom comprehensive reverse humeral tray set w/ lock rings 44 mm + 15 mm & + 18 mm sterile, REF CP561705, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder
…and 4 more.
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page